  • Page 1 MC5 CONTROLLER INSTRUCTION BOOK Page 1 of 24 M7044 iss 3.3...
  • Page 2 In the interest of continued development Bibby Scientific Limited reserve the right to alter or modify the design and /or assembly process of their products without prior notification. This product is manufactured in Great Britian by Electrothermal, part of the Bibby Scientific Group of companies.
  • Page 3: Section 1 Introduction

    1.2. A rod support clamp is provided at the rear of the controller to take a standard 12.5mm (I/2”) diameter rod. 1.3. The MC5 has a short mains output lead with an IEC socket to connect it to the resistive load. An accessory extension mains lead is available where remote operation is required (e.g. in a fume extraction unit).
  • Page 4: Section 2 Symbols And Using This Instruction Book

    2. SYMBOLS AND USING THIS INSTRUCTION BOOK. 2.1. Throughout this Instruction book the following symbols are shown to identify conditions which pose a hazard to the user, or to identify actions that should be observed. These symbols are also shown on the product, or its packaging. When a symbol is shown next to a paragraph or statement it is recommended the user takes particular note of that instruction in order to prevent damage to the equipment or to prevent injury to one’s self or other people.
  • Page 5: Section 3 Safety Information

    3. SAFETY INFORMATION This product has been designed for safe operation when used as detailed in accordance with the manufactures instructions. NOTE: Failure to use this equipment in accordance with the manufactures operating instructions may compromise your basic safety protection afforded by the equipment and may invalidate the warranty / guarantee.
  • Page 6 To prevent electronic overheat and potential fire Do not cover this product when connected to the mains power supply. Do not obstruct the ventilation slots in the controller base. It is not recommended to leave any heating apparatus unattended during operation. Only use Original Equipment manufacture’s spares and accessories.
  • Page 7 4. UNPACKING AND CONTENTS. 4.1. Please check the contents of your carton against the diagram. (Mains plug and cord set may differ from illustration). Contents. Item Description Control unit Moulded mains cord and plug set. (May be different to illustration). Instructions book (this document).
  • Page 8: Section 5 Installation

    5. INSTALLATION 5.1. Electrical safety and installation. 5.2. This equipment is designed to safe under the following conditions:- Indoor use. • Altitude up to 2000 meters. • Temperatures between 5°C and 40°C. • Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to •...
  • Page 9: Section 6 Environmental Protection

    ‘Appendix A’ of this instruction book. Bibby Scientific’s Electrothermal branded product range is registered with the Environment Agency under the name of as Electrothermal Engineering Limited as being a producer of WEEE (Waste Electronic and Electrical Equipment) through b2b Compliance, an authorised waste collection compliance scheme.
  • Page 10 7. EQUIPMENT OPERATION. 7.1. The MC5 controller has been designed for easy operation. The illustrations below show detailed layouts of this control unit. IEC mains input socket Earth continuity test point Data plate label. (See for correct voltage type). Mains Output IEC plug.
  • Page 11 7.3. Connect to the mains electricity supply. Observe the white neon lamp is illuminated showing the presence of mains electricity. 7.4. Switch the controller on by turning the regulator control knob clockwise to increase power. The amber neon lamp on the controller will pulsate to show that power is being supplied to the mantel, heating tape or cord etc.
  • Page 12: Section 8 Technical Specification

    8. TECHNICAL SPECIFICATION. 8.1. Specifications. Mains supply voltage 230V AC 50/60 Hz Mains supply voltage MC5X1 115V AC 50/60 Hz Maximum coupled power 800 Watts. consumption 230V Maximum coupled power 460 Watts consumption 115V Controller power consumption <1 Watt. Fuse rating 5A 5x20mm Glass Quick blow.
  • Page 13: Section 9 Maintenance

    Unplug the unit from the mains voltage supply and allow it to cool before undertaking any maintenance tasks. Maintenance should only be carried out under the direction of the Responsible Body, by a competent electrician. Failure to do so may result in damage to the product and in extreme cases be a danger to the end user.
  • Page 14 9.3.2. Remove and retain the regulator spindle locknut and shakeproof washer. 9.3.3. Turn the controller up-side down and remove the two fixing screws retaining the base. Lift the base off. 9.3.4. Disconnect the tagged wires one at a time from old and connect to the new energy regulator.
  • Page 15 9.3.6. Replace the energy regulator control knob ensuring the flat on surface on the energy regulator spindle is aligned. 9.3.7. Replace the base and refit the two fixing screws. The responsible body shall check the electrical safety of the unit before further use. 9.4.
  • Page 16 9.5. Replacing the Controller Output lead. 9.5.1. Turn the controller off and remove it from the mains supply. Disconnect it from any heating resistive loads such as Heating Mantles, Heating Tapes and Cords. 9.5.2. Turn the controller up-side down and remove the two fixing screws retaining the base. Lift the base off.
  • Page 17: Section 10 Customer Support

    9.5.11. Reconnect the mains and neutral wires from the cable to the PCB. (See wire diagram). 9.5.12. Replace the base and refit the two securing screws. The responsible body shall check the electrical safety of the unit before further use. 9.6.
  • Page 18 Toll free:800-225-9243Tel: 609-589-2560 Fax: 609-589-2571 Tel: +44(0)1785 812121 Email: labproducts@techneusa.com Fax: +44(0)1785 810405 Http www.techneusa.com General enquiries : • info@bibby-scientific.com Order enquiries : • sales@bibby-scientific.com Technical support : • electrothermalhelp@bibby- scientific.com www.electrothermal.com • Page 18 of 24 M7044 iss 3.3...
  • Page 19 APPENDIX ‘A’. DECONTAMINATION CERTIFICATE. Bibby Scientific Limited Beacon Road, Stone, Staffordshire ST15 0SA. Great Britain E-mail: electrothermalhelp@bibby-scientific.com Tel: +44(0)1785 812121. Fax: +44(0)1785 810405 DECONTAMINATION CLEARANCE CERTIFICATE For the Inspection, Repair or Return of Medical, Laboratory or Industrial Equipment. Prior to a Service Engineer working on equipment that has been in an environment where substances hazardous to health may have been used, you are requested to provide the following information: CUSTOMER DETAILS Company:-...
  • Page 20: Section 11 Spares And Accessories

    11. SPARES AND ACCESSORIES. Neon Clear CRM5607 Amber CRM5608 Extension Lead M6332 (Europe) M6902 (UK) Plug and lead set (UK) AZ9165 Plug and lead set AZ6746 (USA) Page 20 of 24 M7044 iss 3.3...
  • Page 21: Section 12 Notes

    12. NOTES. Page 21 of 24 M7044 iss 3.3...
  • Page 22 Page 22 of 24 M7044 iss 3.3...
  • Page 23 Page 23 of 24 M7044 iss 3.3...
  • Page 24: Section 13 Ec Declaration Of Conformity

    General enquiries : • info@bibby-scientific.com Order enquiries : • sales@bibby-scientific.com Technical support : • electrothermalhelp@bibby-scientific.com www.electrothermal.com • Part of the Bibby Scientific Group © © © © 2013 Bibby Scientific Limited. All rights reserved Page 24 of 24 M7044 iss 3.3...