PerkinElmer 5184 Instruction Manual
PerkinElmer 5184 Instruction Manual

PerkinElmer 5184 Instruction Manual

Ultra low-noise voltage preamplifier

Advertisement

Quick Links

Model 5184
Ultra Low-Noise
Voltage Preamplifier
Instruction Manual
222516-A-MNL-D
Copyright © 2000 PERKINELMER CORPORATION

Advertisement

Table of Contents
loading

Summary of Contents for PerkinElmer 5184

  • Page 1 Model 5184 Ultra Low-Noise Voltage Preamplifier Instruction Manual 222516-A-MNL-D Copyright © 2000 PERKINELMER CORPORATION...
  • Page 2 Company Names PerkinElmer Instruments is part of Perkin Elmer, Inc., and includes the businesses formerly trading as EG&G Princeton Applied Research, EG&G Instruments and EG&G Signal Recovery. Declaration of Conformity...
  • Page 3: Table Of Contents

    Table of Contents Table of Contents Chapter One, Description 1.1 Description ..............................1-1 1.2 Options............................... 1-2 Chapter Two, Initial Checks 2.1 Introduction ..............................2-1 2.2 Equipment Needed ............................2-1 2.3 Procedure..............................2-1 Chapter Three, Operating Instructions 3.1 Introduction ..............................3-1 3.1.01 Power Switch .............................
  • Page 4 TABLE OF CONTENTS...
  • Page 5: Description

    FET input stage (800 pV/√Hz at 1 kHz). It has a frequency range of 0.5 Hz to 1 MHz and a fixed gain of 60 dB (×1000). The 5184 is ideal for use with medium impedance cryogenic sources and IR detectors such as HgCdTe, InSb and InAs.
  • Page 6: Options

    Chapter 1, DESCRIPTION 1.2 Options The model 5184 is supplied complete with four 9 V alkaline batteries. The following optional extras are available. Model PS0108 Remote Line Power Supply for the model 5184. See Appendix B for details of this unit.
  • Page 7: Introduction

    POWER TEST button. The associated indicator should light, indicating that the external supply is functioning properly. 2) Set the 5184 input GROUND / ISOLATE switch to the down (GROUND) position. 3) Connect the oscilloscope to the OUTPUT BNC connector.
  • Page 8 6) Monitor the 5184 output; the output level should be 10 V pk-pk. 7) Set the 5184 input GROUND / ISOLATE switch to the up (ISOLATE) position.
  • Page 9: Introduction

    100 kΩ. The 450 ohm output impedance provides a convenient 10:1 output attenuator if the amplifier is loaded with 50 ohms, which can be useful if the 5184 is required to drive coax cable greater than one meter in length at signal frequencies...
  • Page 10: Battery Operation

    Chapter 3, OPERATING INSTRUCTIONS 3.2 Battery Operation In battery powered operation, the 5184 requires four 9 V alkaline cells (type NEDA 1604, PP3 or equivalent). Each battery fits into its own removable tray accessible from the rear panel. When they are exhausted, all four batteries should be replaced together.
  • Page 11: Appendix A, Specifications

    Max input without damage ±15 V DC or 10 V rms. AC @ 50 Hz Noise See Figure A-1 ; typ 800 pV/√Hz @ 1 kHz Figure A-1, Model 5184 Noise Figure Contours (Typical) Gain x1000 (60 dB) fixed Gain Accuracy ±1%...
  • Page 12 Appendix A, SPECIFICATIONS 450 Ω Impedance Max voltage swing >10 V pk-pk Slew rate > 22 V/µs Polarity Non-inverting Distortion < 0.1% T.H.D. Power Internal Four 9 V alkaline batteries provide approximately 8 hours of use External ±15 V or ±18 V DC @ 40 mA 110 V AC or 240 V AC via optional external model PS0108 power supply...
  • Page 13: Power Supply Unit

    Power Supply Unit (Model PS0108) Appendix Figure B-1, Model PS0108 External Power Supply Fuse Rating 115 V operation 200 mA 230 V operation 100 mA 5 mm × 20 mm, 250 V, Slow Blow Fuse Type...
  • Page 14 Appendix B, POWER SUPPLY UNIT (MODEL PS0108)
  • Page 15 WARRANTY PerkinElmer Instruments Corporation warrants each instrument of its own manufacture to be free of defects in material and workmanship for a period of ONE year from the date of delivery to the original purchaser. Obligations under this Warranty shall be limited to replacing, repairing or giving credit for the purchase, at our option, of any instruments returned, shipment prepaid, to our Service Department for that purpose, provided prior authorization for such return has been given by an authorized representative of PerkinElmer Instruments Corporation.

Table of Contents