Page 1
Yesterday’s challenge Today’s solution Service Manual innova.uk.com...
Page 2
....................... 2 ..........................3 ....................3 ......................4 ......................5 ..................... 6 ........................6 ..................7 ..................8 ..................9 ..................... 10 ..................11 ......................12 ............14 Explanation of symbols Warning Read User Manual Product for indoor use...
Page 3
IMPORTANT: Integralift™ must only be operated by personnel with the proper training to do so. Read the User manual before operating Integralift™ and please keep the User manual at hand for everyone using Integralift™. Don’t leave the User suspended in the sling. This product must be installed by authorized personnel only.
Page 4
Integralift™ creates a more attractive room for the User and gives a subtle Non-intrusive expression. Integralift™ appears as elegant designed furniture which encloses the user’s bed and becomes a natural part of the room’s interior. Cabinets can be special ordered in every colour to fit its environments ...
Page 5
1. Vertical structure 8. Top Cabinet 15. Lifting tape Brackets forstructure 9. Handle 16. Spreader bar 3. Backjib 10. Rear cover 17. Remote control 4. Shaft for back jib 11. Charger 18. Emergency stop 5. Frontjib with engine 12. Vertical cabinet 19.
Page 6
Part.no Pos nr Name Lift strap Sling bar Remote control Battery Charger Mascot 9940 24V max 1300mA Power Cord Connector cable Drum house Motor module Control board Interface card Battery set: 2x12V 5,4Ah Model Capacity WM-230 230 kg FC-230 230 kg 230-330 IW-230 230 kg...
Page 7
If the lifting tape is visibly worn it should be replaced. It is nevertheless recommended to change the lifting tape after 4000 lifting cycles or 4 years usage. 1. Remove the rear cover 2. Drive lifting tape out until the fastening part of the drum is serviceable. 3.
Page 8
1. Remove the protection from the screw that attaches the band to the spreader bar. 2. Unscrew the screw and place the new spreader bar in the lifting tape. 3. The spreader bar screw is to be tightened thoroughly with a screwdriver, 15 nm moment minimum 4.
Page 9
1. Remove the rear cover 2. Unscrew the 4 screws that fixes the control board to the front jib. 3. Pull control board down gently and remove all cables. 4. Place new control board and attach all cables. a. Be careful with polarity, in conjunction with battery and motor. b.
Page 10
1. Remove both rear cover and the top cabinet. Tip: With a vertical elevator-support the top cabinet can be removed by one person. 2. Remove the connector cable from the connector card. 3. Unscrew the four holding screws/locknuts pictured in top picture, left. 4.
Page 11
7. Test Integralift™ functionality. 8. Register the change of motor module in the log. Batteries Control board w/motor switch Gear motor w/break NB: If the motor, controller card or batteries have errors, these parts are to be sent to Innova™ for inspection.
Page 12
Replace control board. 2. If Integralift™ is still not working, contact Innova™ or your Innova™ Dealer. Integralift™ lifts up, but will not move downwards. Gently pull the lift strap while driving the lifter downwards. Check the remote control is correctly attached.
Page 13
Perform a visual inspection of lifting strap with drum and attachments, and make sure everything is working accordingly. Innova™ c. Check that noise disappears during usage, if not; contact or your local retailer. Innova™ or your Innova™ Dealer. In the event of other errors; contact...
Page 14
Integralift™ Customer Model Name Serial number Adress Production year Zip number Installed date Placement (room) Control Comment Fastening Comments about room/house Specify loadbearing structure & quality Distance from floor to ceiling Specify chosen fastening, wall/ceiling Screws used Structure Vertical structure Jibs Bracketts Driving unit and lifter...
Page 15
0845 034 1450 0845 034 1451 Yesterday’s challenge, enquiries@innova.uk.com Today’s solution www.innova.uk.com Innova™ is a trading name of CoreCare Global Ltd,a company registered in England & Wales at Linkside House, Forest Lane Head, Harrogate, HG2 7TE. Registered Company Number: 08931470.
Need help?
Do you have a question about the Integralift FC-230 and is the answer not in the manual?
Questions and answers