Calibrating the Scale
In the event that the scale loses accuracy , recalibration may be required if the load
cell of the scale is overloaded due to dropping something heavy on the scale or
dropping it.
NOTE: Normally the scale does not require calibration. You require a 100 gram, and
two 1 kilogram weights to perform the calibration – if you do not have these calibration
weights, please do not attempt to calibrate the scale.
Going to the Calibration Mode
1)
Remove the weight from the platform (if any), and switch the scale OFF.
2)
While pressing "FRANK LABEL" key, press and release "ON/OFF/TARE" key.
3)
Press "1" key to calibrate the scale.
4)
Now the scale is in "Calibration Mode", and the LCD will display "0kg Press
FkLtr" on the first line and "
around 2500.
4)
When the number XXXX display is stable, press the "Frank Letter" key to
confirm.
5)
At this moment the display will read "100g Press FrkLtr". Repeat step 4 for
1kg and 2kg weights. The scale will restart and is ready for weighing and
assessing postal rates.
Setting Franking Machine Options
To set the default franking machine, turn off the scale first. W hile pressing
"FRANK LABEL" key, press and release "ON/OFF/TARE" key. Press "2" key to set
the default franking machine type. Then choose the desired franking machine as
displayed on the LCD.
Displaying the Serial Number of the PS2000 UK Scale
To display the 8-digit serial number of the scale, turn off the scale first. While
pressing "EURO AIRMAIL LETTERS" key, press and release "ON/OFF/TARE" key.
The serial number will be displayed on the LCD. Press "ON/OFF/TARE" key to
restart scale or wait for it to start automatically.
ERROR DISPLAY
Overload Will be displayed when the weight on the platform exceeds 2000
grams.
ExcessWt Will be displayed when the weight on the platform exceeds the postal
service's maximum allowable weight.
XXXX" on the second line where XXX is
PS2000 UK
Electronic Postage Computing Scale
User's Manual
Specifications
?
Capacity
?
Graduation / Resolution
?
Minimum Weight
?
Accuracy
?
Power consumption
?
Tare
?
Weight of Unit Scale
?
Dimensions
?
Platform Size
?
Power supply
: 2kg excluding Tare
: 1g
: 1g
±
±
:
0.1% of reading
1 digit
: Approx. 0.35 watts
: 500g maximum
: 1134g
: 26.5cm X 20cm X 8.3cm
: 20cm x 15cm
: Mains via 9 Volts Main Adapter

Advertisement

loading
Need help?

Need help?

Do you have a question about the PS2000 UK and is the answer not in the manual?

Questions and answers

Summary of Contents for Elane PS2000 UK

  • Page 1 LCD. Specifications Capacity : 2kg excluding Tare Displaying the Serial Number of the PS2000 UK Scale Graduation / Resolution : 1g To display the 8-digit serial number of the scale, turn off the scale first. While Minimum Weight : 1g pressing “EURO AIRMAIL LETTERS”...
  • Page 2 Pressing the Metered (M)/Stamped (S) key will toggle metered/stamped rates to available domestic services. The PS2000 UK Keyboard Layout Programming the Postal Rates While pressing the “FRANK LETTER” key, press and release “ON/OFF/TARE” key to invoke the Programming Menu.
  • Page 3 Invoke the Programming Menu as described above and then press “2” to download the rates from the rate chip. Please wait for the scale to finish loading postal rates.