Component List; Specifications - Prism Medical UK C1000 Owner's Manual

Table of Contents

Advertisement

Component List

The following components are included with your new
C1000
lift (Manual or Motorized traverse)
• Pneumatic Hand Control
• Lift Charger (mounted on the wall or ceiling at the end of the track)
• Owner's Manual
• Warranty Card
SLINGS: If a sling has been supplied with the lift refer to the instructions included with the sling.
ACCESSORIES: If additional accessories such as a turntable, or gate system have been supplied with the lift
refer to the instructions included with those items.
IMPORTANT: Before initial use, the lift unit must be charged for 4 hours. Refer to section titled
Charging Instructions
"
connected refer to the section titled "
Specifications of C1000 lift
Lift Motor:
Traverse Motor:
"H" Frame Traverse Motor:
Charger Input:
Charger Output:
Batteries:
Lift Case:
Hand Control:
Lifting Range:
Lift Weight:
Maximum Load:
Duty Cycle:
Rated Performance:
Type of Applied part:
Other Symbols Used:
This symbols is used to point out instructions and information related to work place safety where
injury may occur if the information is disregarded or ignored. Follow these instructions, be careful and
attentive at all times.
Body Weight
Class II Equipment - electrical equipment in which protection against electric shock does not rely on
basic insulation only.
C1000 - User Guide
". The hand control airline tube must also be connected to the lift. If it is not
Connecting airline to the lift
24 VDC
24 VDC
(Optional at time of Purchase)
24 VDC
(Optional at time of Purchase)
120VAC
1.0 Amps
24 VDC
1.5 Amps
24 VDC
(2 x 12 VDC) 5.0 AH, Sealed Lead Acid
Flame Retardant ABS
Pneumatic
Up to 96" (2438mm)
24.5 lbs
Standard maximum load in 1000 lbs.
1 minute ON, 9 minutes OFF
17-20 lifts at 1000 lbs, 10% duty cycle, each lift being 24 inches at the middle of the
lifting range (from 54" strap out to 30" strap out) per full battery.
Please note: the lift has a break in period; breaking in of the lift will need to be done
before these numbers will be achieved. The breaking in period will vary from lift to lift
and is dependent on the frequency of use and the types of load being applied, the higher
the load and a greater frequency of use will break in the lift faster.
B type
Rev: 9SEP2011
C1000
lift system:
".
Shipping/Storage Conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure: 500 to 1060 hPa
Maximum load of the installed lift is
determined by referring to the product
label located on side of lift.
Emergency Stop
-40 to +70 ºC
10 to 100% RH
Page: 4