Enteral feeding pump
Table of Contents

Advertisement

kENDALL
KANGAROO JOEY
Enteral Feeding Pump
Operation and Service Manual
Enteral Feeding Pump — Operation and Service Manual
Nutripompe — Manuel d'utilisation et d'entretien
Pumpe für die enterale Ernährung — Benutzer- und Service-Handbuch
Pompa di alimentazione enterale — Manuale operativo e di manutenzione
Bomba para alimentación enteral — Manual de operación y servicio
Enteral matningspump — Användar- och servicemanual
Enterale voedingspomp — Gebruiks- en onderhoudshandleiding
Bomba de alimentação entérica — Manual de funcionamento e assistência
Enteraalinen ravitsemuspumppu — Käyttö- ja huolto-opas
Pumpe til enteral ernæring — Betjenings- og servicehåndbog
Aντλία εντερικής σίτισης — Εγχειρίδιο λειτουργίας και συντήρησης
Pumpa pro enterální výživu — Provozní a servisní příručka
Enterális táplálópumpa — Kezelési és karbantartási kézikönyv
Насос для энтерального питания — Руководство по эксплуатации и техническому обслуживанию
Pompa żywienia dojelitowego — Instrukcja obsługi i serwisowania
Enteral Besleme Pompası — Kullanım ve Servis Kılavuzu
Enteral ernæringspumpe — Bruker- og vedlikeholdshåndbok
Rx
ONLY
0123
™ Trademark of Tyco Healthcare Group LP or its affiliate.
™*SPRAYNINE is a trademark of SprayNine Corporation.
™*PHISOHEX is a trademark of The Mentholatum Company.
™*HIBICLENS is a trademark of Regent Medical Limited.
™*VESTA-SYDE is a trademark of Steris Inc.
TYCO HEALTHCARE GROUP LP • 15 HAMPSHIRE STREET • MANSFIELD, MA 02048 • MADE IN U.S.A. • © 2008 TYCO HEALTHCARE GROUP LP
ALL RIGHTS RESERVED • 080516 • PRODUCT INFORMATION IN U.S. 1-800-962-9888 • www.tycohealthcare.com
TYCO HEALTHCARE U.K. LTD • GOSPORT PO13 0AS, U.K.
182545

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Kendall Kangaroo Joey and is the answer not in the manual?

Questions and answers

vicki brooks
March 17, 2025

how do I reset the language

1 comments:
Mr. Anderson
March 17, 2025

To reset the language on a Tyco Kendall Kangaroo Joey pump, go to the “MoRe oPTIons” menu. The language selection screen only appears during the first power-up, but it can be changed later through this menu.

This answer is automatically generated

Lawrence A Palmer
May 1, 2025

How to use Kangaroo Joea Pump Backpack?

Subscribe to Our Youtube Channel

Summary of Contents for Tyco Kendall Kangaroo Joey

  • Page 1 ™*VESTA-SYDE is a trademark of Steris Inc. TYCO HEALTHCARE GROUP LP • 15 HAMPSHIRE STREET • MANSFIELD, MA 02048 • MADE IN U.S.A. • © 2008 TYCO HEALTHCARE GROUP LP ALL RIGHTS RESERVED • 080516 • PRODUCT INFORMATION IN U.S. 1-800-962-9888 • www.tycohealthcare.com TYCO HEALTHCARE U.K.
  • Page 2: Table Of Contents

    english Thank you for purchasing the KANGAROO JOEY™ Enteral Feeding Pump. With proper care, this device will provide you with years of precision service. Table of ConTenTs Page section I — General Information ................1 section II —...
  • Page 3 Table of ConTenTs Selecting Continuous or Intermittent Mode ............... . 13 Continuous Mode (Adjust Feed) .
  • Page 4 ................. 41 This product contains software solely owned by Tyco. Tyco grants the user a non-exclusive, limited license to use the software according to the operating instructions. A copy of the license can be obtained from Tyco. Enteral Feeding Pump ™...
  • Page 5: Section I - General Information

    section I – General Information The KANGAROO JOEY Enteral Feeding Pump is a simple-to-use, precision enteral feeding pump. It is easily programmed to provide patients with either continuous or intermittent feeding, and can provide automatic flushing capability when used with KANGAROO ePump™...
  • Page 6 service: • Serial I/O port for data retrieval and software updates • Uses replaceable, rechargeable Lithium-Ion Battery Pack for 18 hours of backup power • Easily removable power cord Status Line Menu Selections Status LEDs Input Settings Menu Selection Buttons Output Totals Power Button Graphic...
  • Page 7: Section Ii - Safety And Warnings

    figure 1b. KanGaRoo JoeY enteral feeding Pump set loading area. section II — safety and Warnings CaUTIon: U.S. federal law restricts the sale of this device to physicians or to their direct representatives. 1. Read this booklet thoroughly before using the KANGAROO JOEY Enteral Feeding Pump. 2.
  • Page 8: Section Iii - Icon Identification

    of this equipment, may lead to a reduced level of safety of the pump. While use of the pump’s serial port is not recommended for customers, the following should be considered in the event that the customer decides to connect to this port: •...
  • Page 9: Section Iv - Initial Setup

    Information Indicator This indicates information only, and does not require any immediate action. eZ Pump Mode Indicator Indicates that EZ Pump Mode feature is activated. Communication Indicato Indicates that communication feature is activated Volume to be Delivered Indicator Indicates that a VTBD has been programmed. Keep Tube open Indicates that Keep Tube Open feature is in operation.
  • Page 10: Battery Setup

    figure 2b. battery setup The battery is packed partially charged. Charge battery 6 hours prior to use off A/C power. It is also recommended that the battery charge be “topped off” before battery-only operation is utilized. The battery pack will charge continuously whenever the pump is plugged into a wall outlet, including during normal use of the pump with A/C power.
  • Page 11: Section V - Instructions For Use

    figure 3b. attaching and using the Pole Clamps. section V — Instructions for Use Quick start PoWeR on lower, right-hand corner of control panel. 1. Press 2. Open blue door enclosing pump set loading area. 3. To load pump set (refer to Figure 4): •...
  • Page 12: General Startup

    (note: display status line should read SET LOADED). 4. suspend the pump set bag so that the top of the starting volume of formula is 6 inches above the pump (figure 4b). PRIMe PUMP and aUTo PRIMe. For Feed & Flush sets, auto priming will prime 5.
  • Page 13: Power On/Off

    A fully-charged, new battery will supply 18 hours of backup power at 125 mL/hr before requiring recharging. Contact Customer Service, Section XI, to replace the battery pack if battery life degrades. note: The KANGAROO JOEY Enteral Feeding Pump will not operate unless an approved KANGAROO JOEY Enteral Feeding Pump battery pack is installed in the pump.
  • Page 14: Loading Pump Sets

    loading Pump sets The KANGAROO JOEY Enteral Feeding Pump will indicate LOAD A SET in the screen’s status line if a pump set is not installed. A blinking icon of a pump set will also appear in the upper left corner of the screen. Figure 5A shows the screen for continuous feeding mode and Figure 5B shows the screen for intermittent feeding mode (intermittent feeding mode not available in EZ Pump Mode).
  • Page 15: Prime Pump

    figure 7a. opening menu figure 7b. opening menu for Continuous feed mode for Continuous feed mode with a feed-only set. with a feed & flush set. figure 7C. opening menu figure 7D. opening menu for Intermittent feed mode with a for Intermittent feed mode feed-only set.
  • Page 16: Hold-To-Prime

    When auto-priming is completed, the status line will show AUTO PRIME COMPLETE and will no longer show the aUTo PRIMe option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the priming, as described below. Hold-to-Prime The hold-to-prime options allow for precise interactive control of the priming process.
  • Page 17: Selecting Ez Pump Mode

    Turning on eZ PUMP MoDe will automatically force the unit out of the InTeRMITTenT (bolUs feeD) mode if set. Although the Continuous Mode feed VTBD will be set to 0 while eZ PUMP MoDe is enabled, the previous value of the feed VTbD will be stored in the pump memory and recalled when “eZ PUMP MoDe”...
  • Page 18: Continuous Mode (Adjust Feed)

    Select ConTInUoUs/InTeRMITTenT under the MORE OPTIONS menu, Figure 10A. Then press ConTInUoUs MoDe InTeRMITTenT MoDe from the seleCT MoDe menu, Figure 10B. Press Done to exit. figure 10a. MORE OPTIONS menu. figure 10b. SELECT MODE menu for continuous or intermittent mode selection.
  • Page 19: Intermittent Mode (Adjust Bolus)

    Press flUsH InTeRVal in Figure 12A to define the time interval between starts of flushing. Press the buttons in Figure 12C to program the time interval from 1 to 24 hours in increments of 1 hour. Press enTeR (top button) to exit the menu. The pump will automatically limit flushing settings which exceed pump capabilities to deliver.
  • Page 20: Intermittent Mode Feeding - Bolus Max

    noTe: If the nUMbeR of bolUses is reprogrammed during a feeding (from HOLD mode), the pump will restart the bolus feeding and will deliver all the boluses as defined by the new setting. figure 13a. ADJUST BOLUS menu. figure 13b. setting the bolus Rate. figure 13C.
  • Page 21: Run

    figure 14b. setting the flush figure 14C. setting the flush Time figure 14a. ADJUST FLUSH Volume. Interval. menu, intermittent mode. note: The nature of two intermittent processes, periodic bolus feeding and periodic flushing, is such that occasional conflicts can arise where each process needs to be running at the same time. The flushing process is assigned precedence over the feeding process for Feed &...
  • Page 22: Kto

    figure 15a. RUNNING mode screen figure 15b. KTo. for continuous feed mode. figure 16. Program locked screen, which indicates the pump is in loCK seTTInGs mode. (Not Available in EZ Pump Mode) The Keep Tube Open (KTo) function within the KANGAROO JOEY Enteral Feeding Pump will help prevent a blockage of the feeding set by moving the feeding solution periodically within the tube.
  • Page 23: Hold

    Hold HolD while it is RUNNING, such as in Figure 15A. The status line will indicate Pump operation can be paused by pressing HOLDING and the yellow LED indicator will be lit up to give a quick visual indication of an informational status in a darkened room. Note: the pump’s alarm will sound if it has been left HOLDING for 10 minutes without further input.
  • Page 24: Flush Now

    flush now (Not Available in EZ Pump Mode) During the RUNNING mode, the pump can be diverted to immediately perform a flush (flush on demand) when a feed/flush set FLUSH NOW in the RUNNING menu, input a desired flush volume, then press enTeR.
  • Page 25: Language Selection

    figure 20. feed and flush totals for up to 72 hours of previous history, excluding data from the current hour. languages lanGUaGes in the MORE OPTIONS menu to change the preferred language for the screen interface. Press Continuous/Intermittent (Not Available in EZ Pump Mode) Press ConTInUoUs/InTeRMITTenT in the MORE OPTIONS menu to access the SELECT MODE menu, Figure 10B.
  • Page 26: Biotech Options

    CeRT InTV indicates the number of times the pump has been powered on since the last re-certification. See Section VI for details of re-certifying the pump. bioTech options figure 22. bioTech option screen, with lock setting, eZ PUMP MoDe, and Communications features activated.
  • Page 27: Section Vi - Re-Certification Of Performance

    section VI – Re-Certification of Performance system Performance Re-Certification A series of self-diagnostic checks can be run to re-certify pump performance. It is recommended that this re-certification procedure be run every two years, or any time the pump is suspected of having improper performance. See Section XIII – Service Part Numbers for ordering new Re-Certification Pump Sets.
  • Page 28: Section Vii - Cleaning

    section VII — Cleaning Cleaning should be performed as needed. It may also be desirable to define cleaning intervals based on knowledge of the environment in which the pump is used. Only personnel trained in the cleaning of medical devices should perform cleaning. CaUTIon: Do not immerse pump or power cord in water or other cleaning solution;...
  • Page 29: Section Viii - Battery Replacement

    Re-certification may be done at the user’s Biomedical Engineering Department, an outside service, or by Tyco Healthcare Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service in the US, call 1-800-448-0190. In Canada, call 1-877-664-TYCO(8926). Outside of the US and Canada, please contact your local Customer Service for more information.
  • Page 30: Section Ix - Error/Warning/Informational Screens And Troubleshooting

    24. Installation of battery Pack. WaRnInG: USE OF A BATTERY PACK NOT SUPPLIED BY TYCO HEALTHCARE GROUP LP CAN BE DANGEROUS AND VOIDS ALL WARRANTIES AND PERFORMANCE SPECIFICATIONS. The battery will charge continuously whenever the pump is plugged into a wall outlet. 6 hours of charging is required to fully recharge the Battery Pack.
  • Page 31: System Error

    The Error, Warning, and Information Screens are described below: system error (Fixed Audible Volume) The system error screen is the most general form of error, Figure 25. The only way to exit from a System Error is to power down. No ConTInUe option is allowed because of the severity of the error.
  • Page 32: Hold Error

    Turn the power off and back on to try to clear the error. If the error cannot be resolved in a timely manner, press PoWeR to stop operation of the pump and put a different pump into service. Hold error (Adjustable Audible Volume) The HolD eRRoR screen, Figure 26, will appear if the pump has been inactive, without input, for more than 10 minutes.
  • Page 33: Flush Error

    figure 28. feed error screen. flush error (Adjustable Audible Volume) The flUsH eRRoR screen appears when the flushing solution is no longer being delivered because the flush bag is empty or because of a clog between the pump and flush bag. Check the flush line to find the occlusion causing the blockage. If the occlusion cannot be cleared, load a new pump set, prime it, and press ConTInUe to restart the feeding, Figure 29 If the error still cannot be resolved, press...
  • Page 34: Battery Low

    figure 31. Pump set Dislodged error screen. battery low (Fixed Audible Volume) The baTTeRY loW screen appears and the alarm beeps continuously when the battery needs to be recharged. There is approximately 15 minutes of battery life remaining when this screen appears, Figure 32. Plug the power cord in to begin recharging.
  • Page 35: Led Array

    figure 34. Running screen with the > 24 HRs of Pump set Usage Indicator in upper left of the screen. leD array The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in darkened rooms.
  • Page 36: Section X - Specifications And Symbols

    section X — specifications and symbols specifications Ul listing Medical equipment KANGAROO JOEY™ Enteral Feeding Pump (1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with UL60601-1. (2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance with CAN/CSA C22.2 No.
  • Page 37 occlusion Pressure 15 psi (103 kPa) Nominal Dimensions Height: 10.4 cm (4.1”) Width: 13 cm (5.1”) Depth: 9.1 cm (3.6”) Weight 0.77 kg (1.7 lbs), 1 kg (2.3 lbs) with pole clamp Material Soft-Touch Coating: Latex-Free Thermoplastic Urethane Housing: Polyester/Polycarbonate blend Power Use AC adapter for wall outlet usage.
  • Page 38 alarm Volume Minimum of 65 dBA at 1 meter in maximum volume orientation operating Temperature 10° - 40° C (50° - 104° F) 75% R.H. non-condensing storage Temperature 0° - 50° C (32° - 122° F) 95% R.H. non-condensing Type of Protection against electrical shock Class II, Internally-powered Equipment Degree of Protection against electrical shock Type BF...
  • Page 39: Symbols On Pump And Accessories

    symbols on Pump and accessories The following symbols are found on the pump or on the accessories, such as the A/C adapter. Refer to Accompanying Buttons for Selecting Documents Screen Prompts See Accompanying VDE in Europe Documents Power Source Type BF Protection (Degree of 5 VDC 2.4 A protection against electrical UL in USA...
  • Page 40: Section Xi - Customer Service

    1. Call Customer Service for an Authorized Return Number and shipping instructions, using the appropriate phone number below. 2. Pack the instrument carefully and ship the insured parcel to the following locations (or your local service center): United states Canada Tyco Healthcare Group LP Tyco Healthcare Canada 98.6 Faichney Drive 7300 Trans Canadian Highway...
  • Page 41: Battery Pack

    figure 35. blue Door Replacement. battery Pack See Section VIII for full instructions on replacing Battery Pack. During long-term non-usage or storage (greater than 1 year) of the pump, battery discharge and deactivation may tend to occur. It is recommended that the battery be disconnected during this time. The battery may need to be charged and discharged several times to restore optimal capacity.
  • Page 42: Section Xiii - Service Part Numbers

    section XIII – service Part numbers To place an order for parts, or if technical assistance is required, call customer service. The KANGAROO JOEY Enteral Feeding Pump contains a limited number of serviceable parts, Figure 37. User maintenance is to be performed only by appropriately qualified technical personnel.
  • Page 43: Section Xiv - Warranty

    Customer the purchase price for the defective part or product. Dated proof of original purchase will be required. Tyco Healthcare Group LP does not assume any liability for loss arising from unauthorized repair, misuse, neglect, or accident.
  • Page 44: Section Xv - Electromagnetic Conformity Declaration

    section XV – electromagnetic Conformity Declaration The KANGAROO JOEY Enteral Feeding Pump has been built and tested according to UL60601-1, E189131 CSA-C22.2 No.601.1-M90, and EN60601-1-2 Standards. The KANGAROO JOEY Enteral Feeding Pump is intended for use in the electromagnetic environment specified below. The user of the pump should assure that it is used in such an environment.
  • Page 45: Section Xvi - Accusystem

    Tyco Healthcare Group LP KANGAROO ePump AccuSystem√ feeding sets. The use of generic feeding sets or other feeding sets not produced by Tyco Healthcare Group LP is not supported or sanctioned by Tyco Healthcare Group LP. The use of unauthorized feeding sets could jeopardize the accuracy of the KANGAROO JOEY Enteral Feeding Pump.
  • Page 46 ™*SPRAYNINE is a trademark of SprayNine Corporation. TYCO HEALTHCARE GROUP LP • 15 HAMPSHIRE STREET • MANSFIELD, MA 02048 • MADE IN U.S.A. • © 2008 TYCO HEALTHCARE GROUP LP • ALL RIGHTS RESERVED • 080430 • IN U.S. 1-800-962-9888 • www.tycohealthcare.com ™*PHISOHEX is a trademark of The Mentholatum Company.

Table of Contents