Cel-63x series. sound level meters

Advertisement

Fit the batteries
The CEL-63x instrument requires three AA batteries.
Remove the battery compartment cover and fit the batteries using the correct
polarities. Refit the battery compartment cover.
Note: The batteries can be zinc carbon, alkaline or rechargeable. DO NOT mix
battery types.
Switch the instrument ON
Press the On/Off key. Confirm the battery condition symbol shows there is a good
level of charge in the batteries.
The initialisation screen is displayed for approximately 10 seconds, and then the
measurement screen is displayed in the STOP mode (with red bars at the top and
the bottom of the screen). Press the Menu key to access the settings, memory
and instrument status. The screens displayed depend on the model of instrument
in use.
Settings screen
MENU
SETTINGS
LANGUAGE
4
1
2
3
5
C
C
A
B
A
B
A
B
C
C
C
1. Settings
2. Memory results 3. Instrument status
A. Exit / Back B. Select / Edit
C. Navigation keys
Figure 2. Settings screen
Set language
(Refer to Figure 2)
Select
1
Settings and press
B
to continue. Use navigation keys
C
to select the
Language symbol . Press to continue. Use navigation keys to select the correct
4
B
C
language. Press Select ( ). Press Exit ( ).
B
A
Set frequency modes
(Refer to Figure 2)
Some instruments in the CEL-63x range allow you to select either Octave or
1/3
O ctave with which to take measurements. Select
1
Settings and press
B
to
continue. Use navigation keys to select the Octave symbol . Press to continue.
C
5
B
Use navigation keys to select Octave mode. Press Select ( ). Press Exit ( ).
C
B
A
Set the time and date
(Refer to Figure 3)
Select Settings and press
1
B
to continue. Use navigation keys
C
to select the Set
Clock symbol . Press to continue. Use navigation keys to select the time or date.
2
B
C
Press to edit. Use navigation keys to adjust time/date. Press Save ( ). Press Exit
B
C
B
( ).
A
2
1
Figure 3. Setting the time and date
Calibration
(refer to Figure 4)
Calibration is auto-set to the reference level as set within the Settings menu. Refer to
the manual to set the calibration reference level.
1
A
B
A
Figure 4. Calibration
Select the Measurement Stop screen. Fit the acoustic calibrator fully over the
removable microphone. Press the calibrator's On/Off key . The CEL-63x
1
instrument detects the signal and activates the calibration screen (with yellow bars at
the top and bottom). Press
B
to start the calibration. The instrument calibrates and
displays the word PASSED when finished. Press Exit ( ) to return to the Stop
A
screen. Press and hold the calibrator's On/Off key to switch the calibrator off.
Remove the acoustic calibrator.
Setup
(Refer to Figure 5)
Select Settings and press to continue. Use navigation keys to select the Setup
1
B
C
symbol . Press
2
B
to continue. Use navigation keys
C
to select the required
measurement view for your local legislation. Refer to the User Manual for detailed
information about the measurement views.
Note: This instrument simultaneously records all noise functions for every
measurement it completes. The Measurement View screens simply define which
functions are displayed on the measurement screens.
1
2
Figure 5. Settings options
Measurement control
(Refer to Figure 6)
Select Settings and use the navigation keys to select the Measurement control
symbol . Press
1
B
to continue. Use the navigation keys to select the Measure -
Start/Stop symbol . Press Edit ( ). Use the navigation keys to select the Mode (Key
2
B
Press / Fixed Duration / Timers). Press Save ( ). Use the navigation keys to select
B
additional Fixed duration or Timers options. Press Edit ( ), and use the navigation
B
keys to set the run duration. Press Save ( ) to continue (if required).
B
2
1
Figure 6. Run duration
Measurement data sets
(Refer to Figure 7)
Cumulative measurements made by the instrument apply for the entire measurement
duration. They do not show how the sound level changed during the period of
measurement. The instrument can also make Periodic measurements, which store
measurements acquired during a repeating interval of time. When one interval ends,
the next one begins until the periodic measurements stop.
The instrument's Profile recording function provides an additional time-history
recording channel, which may operate in parallel with either Cumulative or Periodic
data recording. The Profile time history channel can provide faster recording
resolution but for a reduced set of noise parameters.
Select Settings and use the navigation keys to select the Measurement control
symbol . Press
1
B
to continue. Use the navigation keys to select the Data Sets
symbol . Set the Periodic and Profile measurement intervals and press Exit.
2
2
1
Figure 7. Measurement data sets
1
Figure 8. Memory results
Memory results
(Refer to Figure 8)
Press the Menu key to access the settings, memory and instrument status. Use the
navigation keys to select the Memory Results symbol and press Select ( ) to
1
B
continue. Use the navigation keys to select a result in the list, and press Select ( ) to
B
select the measurement.
Use the navigation keys to select View Results (see Figure 9) and press Select ( )
1
B
to display the selected memory results. For additional data press View ( ). The
B
screen displays detailed, read-only data for the selected measurement. Press View
( ) to see the data in a graphical format, using the Octave format saved in the
B
Memory Results. Use the navigation keys to move the vertical line across the graph.
Each graph bar represents an octave level measured in Hz or kHz. Press View ( )
B
again to view tabular octave results.
Use the navigation keys to toggle between functions such as L and L . Press View
max
Aeq
( ) to view the results in tabular form. The results stored for the octave band graph
B
display functions L and L .
max
eq
Press Exit ( ) to leave the Memory Results, or press View ( ) to cycle the Memory
A
B
Results screens again. Note that the display screens will vary according to the model
of CEL-63x instrument you are using.
Delete Memory Results
Refer to the User Manual for details and instructions to delete the results stored in the
instrument's memory.