Specifications - PRO-flex ACP3000_4/36 Operating Instructions Manual

Ac current probes
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
______________________________________________________________
3.

SPECIFICATIONS

Measuring ranges
Output sensitivity
(AC coupled)
Accuracy (45-65Hz)
Minimum load
Frequency range
Phase (45-65Hz)
Position sensitivity
External field
from the probe
Noise
Temperature coeff.
Power supply
Battery life
External Power
Low battery
Overload
Enclosure
Material
Degree of Protection
Dimensions
Output connection
_______________________________________________________________
ACP3000_4
30A / 300A / 3000A AC
ACP6000_4
60A / 600A / 6000A AC
ACP3000_4
100mV / 10mV / 1mV per A
ACP6000_4
50mV / 5mV / 0.5mV per A
± 1% of reading ± 0.1A 30/300A or 60/600A ranges
± 1% of reading ± 1A
3000A or 6000A range
100kΩ for specified accuracy
10Hz to 10kHz (-1dB)
<± 1°
± 2% of reading
± 0.2% of range with cable >200mm (8")
8mV rms in 30A / 60A ranges
2mV rms in 300A / 3000A and 600A / 6000A ranges
± 0.1% /°C
Two AA MN1500 LR6 alkaline batteries
or an external power supply (+3V/100mA)
2000 hours typical
Via 2.1mm (0.08") mini power connector
Indicated by an orange LED
Indicated by a red LED
ARNITE T06-200SNF, UL94 V0
IP40 (not certified by UL)
110 (l) x 65 (w) x 23 (d) mm
0.5m coax cable terminated with a safety BNC