Installation and operating manual
6.2. Solenoid valve test function
Pressing the >>Test<< button will control the solenoid valve and switches on 3 cycles for a length of 2 seconds.
The red warning LED flashes in cycle with the solenoid valve.
Solenoid valve lapsed → red LED off → complete purge air exits
Solenoid valve closed down → red LED on → only measuring gas exits
...
Power
∆
10
20
30
Test
+10
+5
0
°C
P
ressure
D
ew
P
oint
...
∆
35 45 55
Value
-5 -15 -26
°C
The system subsequently returns to normal operating mode.
6.3. Acquiring settings (setup mode)
The software must be initially unlocked to be able to acquire settings. The unlocking will be executed by simultaneously
pressing the >>Value<< + >>Test<< buttons: All green LEDs flash to confirm the activation.
You now have 10 seconds to acquire the settings. Every actuation of a button during this time span will result in the time
span being available again for the complete duration.
...
Power
∆
10
20
30
Test
+10
+5
0
°C
P
ressure
D
ew
P
oint
...
∆
35 45 55
Value
-5 -15 -26
°C
All settings apply until they are amended. Settings can be acquired via the >Value<< and >>Test<< buttons.
6.3.1. Altering the
operating mode
...
Power
∆
10
20
30
Test
+10
+5
0
°C
P
ressure
D
ew
P
oint
...
∆
35 45 55
Value
-5 -15 -26
°C
Once the action or amendment has been finalised and no button has been pressed for a time period of 10 seconds, then the
confirmation for finalising the action / acquisition of the amended values will be executed: All green value LEDs flash. The
system then transfers into normal mode and will be controlled on the basis of the prescribed settings.
26
Press >>Test<< button for 3 seconds.
Activation will be displayed (all green LEDs flash)
Release the >>Test<< button valve switches for 3 cycles
The system subsequently returns to normal operating
mode.
Press >>Test<< + Taste >>Value<< button for 3 seconds
Activation will be displayed (all green LEDs flash)
Release both buttons: Setup mode is activated
Acquire amendments / settings
Press the >>Test<< button.
The power LED illuminates
continuously:
Constant mode
Power LED flashes
[interval 2 sec (on-off)],
Dynamic mode
6.3.2. Altering the values
...
Power
∆
10
20
30
Test
+10
+5
0
°C
P
ressure
D
ew
P
oint
...
∆
35 45 55
Value
-5 -15 -26
°C
...
Power
∆
10
20
30
Test
+10
+5
0
°C
P
ressure
D
ew
P
oint
...
∆
35 45 55
Value
-5 -15 -26
°C
Press the >>Value<< button.
1 value LED illuminates:
35K / -5 °C (example)
Press the >>Value<< button
again.
2 value LEDs illuminate:
Intermediate value
40K / -10 °C (example)
DRYPOINT
M eco control
®
EN