Issue 1
(2) Set the Leak Test
options
TAP the window
(Maximise)
1
Steps 5 and 6: Set the Test Time
+ Set the Wait Time (if necessary)
5

5.3.5 DPI 620 calibrator: Set the PM 620 module to zero

5-6 Pressure calibrator operation (DPI 620)
When you have set the Utility to Leak Test, you can set these
options:
Wait Time: The time before the test starts in
hours:minutes:seconds (hh:mm:ss)
Test Time: The period of the leak test in
hours:minutes:seconds (hh:mm:ss)
Note: To set the Leak Test options, you must have a PM 620
module correctly installed (Section 5.2.3).
Select Settings
2
Go back to the Home window. Start the test
when the pressure system is set up.
6
Use this option to write a new zero pressure value to the
PM 620 module you are using. This option only affects gauge
type modules.
Note: To make a temporary adjustment for zero, you can use
the Tare function: Select Settings > Process > Tare (refer to the
user manual - K0449).
Select Leak Test
Select Test Time
3
7
4
Start the test and let it
complete.
Stop the test (with no results)
K0457 - [EN] English