INSTRUCTIONS
MODEL 61302L
BAROMETRIC PRESSURE SENSOR

1.0 SPECIFICATIONS

Pressure
500 to 1100 hPa standard range
Digital Accuracy*
0.2 hPa (25°C)
0.3 hPa (-40°C to +60°C)
Analog Accuracy**
0.05% of analog pressure range
Analog Temperature
Dependence**
0 . 0 0 1 7 % o f a n a l o g p r e s s u r e
range per °C (25°C reference)
Long Term Stability: 0.2% FS per year
* Defined as ±1 standard deviation from NIST-traceable
pressure reference in clean, dry air. Includes non-
linearity, hysteresis, repeatability, and calibration
uncertainty.
** Defined as ±1 standard deviation from ideal analog
output. Total analog output accuracy is the root sum
square of digital accuracy, analog accuracy, and analog
temperature dependence.
2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA
R.M. YOUNG COMPANY
TEL: (231) 946-3980
Output Rate
1.8 Hz (max) to 1 per minute
Current Output
4 to 20 mA
1 2 - b i t r e s o l u t i o n ( 1 i n 4 0 0 0 )
Selectable pressure range: 500 to
1100 hPa (standard)
Serial Output
Full-duplex RS-232
Half-duplex RS-485
1200 to 38400 baud
Continuous ASCII text
Polled ASCII text
NMEA
0.01 hPa resolution
Supply Voltage
7 to 30 VDC
25 mA max in 4-20 mA mode
7 mA with serial I/O only
Case
Fiber-reinforced thermoplastic
Weight
44 g (1.5 oz)
FAX: (231) 946-4772
PN: 61302L-90
REV: H111215

Advertisement

loading

Summary of Contents for R.M. Young 61302L

  • Page 1: Specifications

    INSTRUCTIONS MODEL 61302L BAROMETRIC PRESSURE SENSOR 1.0 SPECIFICATIONS Output Rate 1.8 Hz (max) to 1 per minute Pressure 500 to 1100 hPa standard range Digital Accuracy* 0.2 hPa (25°C) Current Output 4 to 20 mA 0.3 hPa (-40°C to +60°C)
  • Page 2 2.0 INTRODUCTION 4.2 SERIAL OUTPUT The 61302L may be configured for either full-duplex RS-232 MODEL 61302L Barometric Pressure Sensor is a versatile or half-duplex RS-485 serial communication. For RS-232 electronic barometer featuring high accuracy, low power, serial communication, both jumpers at P3 must be in position wide operating temperature range, and calibrated 4-20 mA ‘2’.
  • Page 3 T h e M O D E L 6 1 3 0 2 L b a r o m e t e r r e q u i r e s n o r e g u l a r maintenance. Periodic calibration certification, if required, YOUNG 61302L SERIES BAROMETER V2.02 is available from the factory.