MODEL 61302V
BAROMETRIC PRESSURE SENSOR

1.0 SPECIFICATIONS

Pressure
Digital Accuracy*
Analog Accuracy**
Analog Temperature
Dependence**
Long Term Stability:
*Defined as ±1 standard deviation from NIST-traceable
pressure reference in clean, dry air. Includes non-
linearity, hysteresis, repeatability, and calibration
uncertainty.
**Defined as ±1 standard deviation from ideal analog
output. Total analog output accuracy is the root sum
square of digital accuracy, analog accuracy, and
analog temperature dependence.
2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA
500 to 1100 hPa standard range
0.2 hPa (25°C)
0.3 hPa (-40°C to +60°C)
0.05% of analog pressure range
0.0017% of analog pressure
range per °C (25°C reference)
0.2% FS per year
R.M. YOUNG COMPANY
TEL: (231) 946-3980
Output Rate
1.8 Hz (max) to 1 per minute
Voltage Output
0 to 5000 mV (standard)
0 to 2500 mV
12-bit resolution (1 in 4000)
Selectable pressure range:
500 to 1100 hPa (standard)
Serial Output
Full duplex RS-232
1200 to 38400 baud
Continuous ASCII text
Polled ASCII text
NMEA
0.01 hPa resolution
Supply Voltage
7 to 30 VDC
2.8 mA with Vout (standard)
1.4 uA shutdown (Trig=0V)
7 mA with serial I/O
Case
Fiber-reinforced thermoplastic
Weight
44 g (1.5 oz)
FAX: (231) 946-4772
PN: 61302V-90
REV: I111215

Advertisement

loading

Summary of Contents for R.M. Young 61302V

  • Page 1: Specifications

    INSTRUCTIONS MODEL 61302V BAROMETRIC PRESSURE SENSOR 1.0 SPECIFICATIONS Output Rate 1.8 Hz (max) to 1 per minute Pressure 500 to 1100 hPa standard range Digital Accuracy* 0.2 hPa (25°C) Voltage Output 0 to 5000 mV (standard) 0.3 hPa (-40°C to +60°C)
  • Page 2 = 0.02 * mV + 950 Effective resolution is about 0.025 hPa Low power consumption and wide temperature range make the 61302V ideal for remote applications using battery or Please see section 4.3 below for details on changing the solar power.
  • Page 3 4.3 SOFTWARE COMMANDS CMD160 n sets the baud rate. Acceptable values are 38400, 9600, 4800, 2400, and 1200. Note that the Software commands sent via serial communication may be b a u d r a t e c h a n g e o c c u r s i m m e d i a t e l y a f t e r t h e used to set operational parameters.
  • Page 4 APPENDIX A 61302V BAROMETER TYPICAL WIRING CONNECTIONS...