G
M
ENERAL
AINTENANCE
Note
Harvey chamber cleaner works best in
a warm, dry chamber.
Hot Surface
Wear protective gloves and safety
glasses. Chamber and trays may
cause burns if touched.
Figure 31: Cleaning the Sterilizer Chamber and
Trays
Warning
Exposure Hazard: Do not operate the
valve plunger while the sterilizer is in a
cycle or under pressure. Exposure to
chemical vapor could result.
Figure 32: Checking the Pressure Relief Valve
50

Weekly

Cleaning Chamber and Trays
At least once a week, clean the chamber and trays with
Harvey Chamber Cleaner. Clean whenever slight staining
appears.
1.
Close the door.
2.
Turn ON the POWER switch. Wait two
minutes.
3.
Turn OFF the POWER switch.
4.
Open the door. The chamber should be warm,
not hot.
5.
Clean the chamber and trays with Harvey
Chamber Cleaner. Follow the instructions on
the container.

Monthly

Check Pressure Relief Valve
Manually operate the valve plunger of the pressure relief
valve on the back of the sterilizer (see Figure 32). This is
to be sure the plunger will move should the chamber
become over-pressurized.
To check the plunger:
1.
Pull on the ring several tomes to ensure free
movement.
2.
If the plunger does not move freely, contact
service personnel to replace the pressure
relief valve.