Table of Contents Introduction ................................. 4 General Specifications ..............................5 General Description ..............................6 Controls and Features ............................... 6 Power Switch ................................6 Control Panel ................................6 Chemi-Filter ................................8 Door ..................................8 Drain ..................................8 Pressure Relief Valve .............................. 8 Piping Components ..............................
Page 3
ABLE OF ONTENTS Pressure Gauge ..............................33 Pressure Relief Valve ............................33 Condenser ................................34 Condensate Tank ..............................34 Door ..................................34 Purge Pump ................................35 Air Filter ................................. 36 Check Valve ................................36 Electrical Components ............................36 Fuses ..................................36 Printed Circuit Boards (PCBs) ..........................
Introduction The Harvey Chemiclave sterilizer is a safe, fast and ® effective sterilizer for use in medical and dental offices, Note hospitals, clinics and other health care facilities. To save space, most figures in this manual show only one of the models.
General Description Controls and Features Power Switch The power switch on the front of the sterilizer turns power to the sterilizer ON (I) or OFF (O). See Figure 1. Control Panel Pressure Gauge Displays the chamber pressure. See Figure 2. Timer Times the 20 minute exposure phase and the purge phase for models EC5500 and EC6000.
Page 7
ENERAL ESCRIPTION Sterilant Level Switch Chemi-filter Pressure Relief Valve Baffle Plate Pressure Filling Air filter Switch Reservoir From Pump Filter Housing Condenser Exhaust to Condensate Chemi-filter Tank Exhaust to Chemi-filter Drain Metering Valve Door Cover Door Power Handle Cord Door Door Main fuse Latch...
Reservoir The reservoir holds about 1 liter of Vapo-Steril solution (see Caution Be sure you are using only Harvey “Technical Specifications” for exact amounts). The filling Vapo-Steril solution. Other solutions will cup has a spring-loaded valve. When the reservoir is full, cause damage to the Chemiclave and the valve stops the flow of solution from the filling bottle.
ENERAL ESCRIPTION With the control knob at DEPRESSURIZE, Vapo-Steril can enter through the fill port and fill the shot chamber, while the chamber itself connects through the exhaust port to the exhaust line. With the control knob at PRESSURIZE, the fill port and exhaust port are closed.
ENERAL ESCRIPTION Exhaust Vapors from the chamber that do not liquefy in the con- denser proceed to the exhaust connection on the filter housing. They exit through the Chemi-Filter into the atmo- sphere. Chemi-Filter The Chemi-Filter traps the formaldehyde from the exhaust vapors before releasing them to the atmosphere.
ENERAL ESCRIPTION Display Control Board Contains logic circuits and display components, including the timer and the buzzer. • Timer: On Display/Control Board. Counts to time the exposure and purge phases. For the exposure phase, the timer operates only if the valve shaft switch is closed and the pressure switch is closed.
ENERAL ESCRIPTION Overtemp Switch Opens to cut power to the heaters if chamber temperature reaches (163°C ± 4) 325°F. Overtemp switch has to be manually reset. Valve Shaft Switch Engaged by the cam on the valve shaft, this switch is closed only when the control knob is set to PRESSURIZE.
Power On Control knob is set to DEPRESSURIZE before turning Caution Use only harvey Vapo-Steril solution in the power ON. This vents any sterilant vapors into the the Chemiclave. Do not dilute, alter or condensate tank. Electrical power is applied to the unit otherwise change Vapo-Steril in any when POWER switch is turned ON.
RINCIPLES OF PERATION Exposure This timed function begins when chamber pressure reaches 20 psig (138 kPa). Pressure switch is then energized. The timer counts down the exposure time of 20 minutes. The decimal point flashes when the timer is active. Chamber pressure is displayed continually on the pressure gauge.
Page 15
RINCIPLES OF PERATION Figure 4: Metering Valve Function...
Maintenance and Servicing Piping Components Warning When troubleshooting the system, be Leaks sure that Vapo-Steril does not drip onto If the sterilizer cannot reach or maintain chamber pressure, power supply board. If the Vapo-Steril first inspect the door gasket and replace it if worn or drips or is spilled on the power supply damaged.
Page 29
AINTENANCE AND ERVICING Check for leaks around tube fittings and pipe threads at the pressure gauge, metering valve, and chamber. Do this by brushing the areas with soapy water. Leaks will appear as bubbles or foam that form and grow larger. If there is a leak, tighten the leaking part.
AINTENANCE AND ERVICING If the door tension is tight and there are still leaks, inspect the gasket. Replace if necessary (see “Gasket Replacement”). If there are leaks around the seal screws, check the door collar. Replace the seal screws, and the door collar if required.
Page 31
AINTENANCE AND ERVICING Loosen and remove bands holding reservoir (do not cut band). Remove reservoir. (Lift up and set on top of unit.) Remove “C” clip next to PC board and pull knob and shaft away from metering valve. Loosen the clamp next to the metering valve on tubing leading to the chamber.
Page 32
AINTENANCE AND ERVICING need new drawing Figure 17: Metering Valve...
AINTENANCE AND ERVICING Pressure Gauge The pressure gauge only indicates pressure. It has no effect on temperature. Failure of the gauge is highly un- usual. Before replacing the gauge, check other causes for pressure changes. Pressure Relief Valve The pressure relief valve releases pressure at the rating marked on the valve.
AINTENANCE AND ERVICING Condenser In models EC5500 and EC6000 the condenser has a heatsink and a coil. Removing the Coil Remove the cover from the unit. Disconnect the fitting at the heatsink (see Figure 19). Disconnect the tubing from the other end of the coil.
AINTENANCE AND ERVICING Door Replacement If replacement is necessary (see Figure 20). Door Collar If you can change the door adjustment by spinning the door, the door collar must be replaced. Replace the seal screws also (see Figure 20). Seal Screws Replace if the unit leaks from the door collar, or if the door collar is replaced (see Figure 20).
AINTENANCE AND ERVICING Air Filter Replace the air filter once a year. Mark the replacement date on the air filter for reference. To replace the air filter, disconnect tubing (from Purge Pump), install replacement air filter in the correct direction, and reconnect tubing (see Figure 21).
AINTENANCE AND ERVICING Fuse Replacement Disconnect unit from power supply. Remove unit cover. Remove the fuse. Check the fuse with an ohmmeter. Replace fuse if necessary. Reinstall fuse. Replace the unit cover. Reconnect unit to power supply. Printed Circuit Boards (PCBs) There are printed circuit boards: •...
Page 38
AINTENANCE AND ERVICING Testing the Printed Circuit Boards (PCBs) Check for voltage to Power Supply Board. • Set the probes of a DC voltmeter across test points TP1 and TP2 (see Figure 22). • Check for 5 VDC ± 0.25V. •...
Page 39
AINTENANCE AND ERVICING Figure 22: Power Supply Printed Circuit Board...
AINTENANCE AND ERVICING Timer The timer is part of the Display/Control Board. It indicates the time only. It cannot interfere with the sterilization cycle. When chamber pressure increases to about 20 psig (138 kPa), Pressure Switch closes. This enables the timer to count down.
AINTENANCE AND ERVICING Heater Warning Shock Hazard: Unplug power cord Functional Check before testing or replacing the heater. Test the continuity of the heater. 100V/230V Units Unplug unit from the power supply. Remove the top cover. Remove the Power Supply Shield over Power Supply Board.
AINTENANCE AND ERVICING Replacement Heater is not field replaceable. Return the unit to the factory for repair. Chamber Temperature To verify the temperature, a thermometer assembly (Part #267569) is required. Remove the tray from the unit. Turn unit “ON” and allow to heat foe a mini- mum of 90 minutes.
AINTENANCE AND ERVICING Locate thermostat: EC5500 — Remove top cover (see Figure 25). EC6000 — On the back of the unit, remove the 1-1/4” cover located in the bottom center of the back. Loosen the locking screw on the thermostat (see Figure 25).
AINTENANCE AND ERVICING Pressure Switch Note The pressure switch is set to close at 20 psi (138 kPa). If Be sure unit is cool and at room it does not, adjust it. temperature. Unplug unit from power switch and remove top cover.
AINTENANCE AND ERVICING Check Valve If the unit does not maintain pressure during operation, the check valve may not be seated properly. To reseat the check Valve: Run the unit in Purge for one minute to loosen the check valve ball. Run a test cycle.
• Always line the instrument tray with chemically pure could create a slight vacuum and pull Harvey Tray Liners. Certain papers (such as paper towels) contain paper-processing impurities. If used waste Vapo-Steril solution back into the as liners, they may stain or cause deposits in the chamber from the waste tank.
ENERAL AINTENANCE Draining the Waste Tank • Drain the waste tank when the VAPO indicator Warning lights on the control panel (see Figure 30). Vapo-Steril solution causes eye damage and may cause skin irritation. • If preferred, drain the waste tank daily—but also Do not get in eyes or on skin or whenever the VAPO indicator lights.
Cleaning Chamber and Trays At least once a week, clean the chamber and trays with Hot Surface Harvey Chamber Cleaner. Clean whenever slight staining Wear protective gloves and safety appears. glasses. Chamber and trays may cause burns if touched.
Do not attempt to remove the cartridge while the sterilizer is processing. Remove the new Chemi-Filter from its package. Replace only with a Harvey Chemi- Mark the replacement date in the “Replace By” Filter. box on the Chemi-Filter label. To calculate the...
ENERAL AINTENANCE Months Until Replacement Calculation Values Needed: Assign proper values to c, D and T. c = no. of cycles/workday (operator defined) D = no. of workdays/week (operator defined) T = Total no. of cycles before filter needs replacement (see chart below). Total Cycles Before Replacement (T) Model EC5500...
ENERAL AINTENANCE Wipe the replacement gasket with a clean, damp cloth. Then insert the gasket into the gasket groove. • Seat the gasket in the groove at 12 o’clock, 6 o’clock, 3 o’clock and 9 o’clock. • Work the gasket into the groove in each of the four quadrants.
Barnstead|Thermolyne or on the date it is purchased by the original retail consumer, whichever date occurs first. Use only HARVEY VAPO-STERIL solution in this sterilizer. DO NOT dilute, alter or otherwise change VAPO-STERIL solution in any way. Do not use water in the CHEMICLAVE sterilizer.