Page 4
Maintenance Safety Information ..............89 Maintenance Intervals ..............89 Service at Leybold ..............89 Maintenance Work ..............90 4.4.1 Opening of the PHOENIX L300i ..........90 4.4.2 Exchanging the Filter Mats ............91 4.4.3 Exchanging the Oil ..............92 4.4.4 Cleaning ..................
Operating Instructions. Note The Leybold PHOENIX L300i leak detector has been designed for safe and efficient operation when used properly and in accordance with these Operating Instructions. It is the responsibility of the user to carefully read and strictly observe all safety precautions described in this section and throughout the Operating Instructions.
Page 6
Make sure that the mains voltage rating on the PHO- ENIX L300i coincides with the locally available mains voltage. When the PHOENIX L300i is running in closed rooms the exhaust has to be put out of doors so that the oil vapor can not be breathed in.
Page 7
Definition of Terms The range of the preamplifier and the vacuum ranges are selected automatically. The auto ranging feature of the PHOENIX L300i covers the entire range or leak Auto ranging rates depending on the selected operating mode. Not only the leak rate signal, but also the pressure in the test sample (inlet pressure P1) and the fore vacuum pres- sure (P2) are used for control purposes.
Page 8
Precision is a measurement mode for the PHOENIX L300i DRY only from an inlet PRECISION pressure < 0,01 mbar. In this mode the PHOENIX L300i DRY has the highest sen- sitivity, the minimum detectable leak rate is 3 · 10 mbar l / s.
The vacuum is achie- ved with a pumping system that is part of the PHOENIX L300i. In addition the vacuum can be generated by pumps which are set up in parallel to the PHOENIX L300i.
Description Another operating mode of the PHOENIX L300i is the Sniffer mode which can only be used when a sniffer line (See Chapter 1.1.6) is hooked up. 1.1.1 Vacuum Diagram PHOENIX L300i The vacuum diagram below shows the major components inside the PHOENIX L300i.
1.1.2 Vacuum Diagram PHOENIX L300i DRY The PHOENIX L300i DRY has a diaphragm pump as fore vacuum pump, making it suitable for applications where oil sealed systems can not be used. Furthermore the PHOENIX L300i DRY contains one more valve, the valve 4b. This valve opens step by step to regulate the inlet pressure into the turbo pump.
1.1.3 Vacuum Diagram PHOENIX L300i MODUL The PHOENIX L300i MODUL has no roughing pump integrated as the other models. Therefore it can be used with an external pump only. This pump can be oil sealed or a dry version with a roughing capacity between 2.5 and 65 m /h.
Because of the higher base pressure of the diaphragm pump the switching from GROSS to FINE mode of the PHOENIX L300i DRY is done by the valve V4b. When the pressure drops below 3,5 mbar the valves V1 and V2b will be close and V4b...
PHOENIX L300i dry and PHOENIX L300i MODUL also. However, in such a case the PHOENIX L300i will not be able to make measurements already at an inlet pressure of 1000 mbar.
To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero func-...
To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero func-...
To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero func-...
Page 25
Description Sniffer line Sniffer line With the use of a sniffer line the PHOENIX L300i can easily be converted to a sniffer leak detector. Available versions: SL300: 4 m SL301: 4 m and 10 m For further information on the sniffer lines SL300 and SL301 see the enclosed manuals.
(sniffer mode). The PHOENIX L300i is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.)
Page 28
The voltage must be in the range 230V (+/- 5%), 115V (+/- 5%) or 100V (+/- 5%) depending on the version. The mains voltage rating for the PHOENIX L300i can be read off from the name plate beneath the mains socket Fig.
Assignment +24 V, constantly applied, power supply for the Leybold partial flow valve or sniffer line. +24V switched by the PHOENIX L300i for an external venting valve 4, 5, 6, 7, 8 These pins are used in connection with accessories.
Page 30
Installation Digital In (Control 2) Digital In (CONTROL2) These inputs can be used to control the PHOENIX L300i with a programmable logic control (PLC). The contacts are numbered from bottom to top. Caution Maximum input voltage 35V. Assignment PLC in free selectable...
Page 31
Remote control REMOTE2 REMOTE2 Through this interface the PHOENIX L300i is controlled wirelessly via Bluetooth or WLAN. The Bluetooth transmitter connects to the remote control RC310 WL. The WLAN module provides the connection to Apple I-Pad or other handheld devices.
(see Chapter.Exhaust Oil Filter). When the PHOENIX L300i is running in closed rooms the exhaust has to be put out-of-doors using the provided adapter. So the oil steams that are harmful to health are lead off. With the provided connection a hose line can be connected to the exhaust of the PHOENIX L300i and lead off.
Page 33
Connection of external pumps The PHOENIX L300i MODUL offers two possibilities to connect the external fore vacuum pump to the DN 25 KF flange. One on the side of the PHOENIX L300i or one in the bottom (measurements see Fig. 5 Fig.
Operation Purge gas/Gas ballast Media Compatibility/Purge Gas The PHOENIX L300i is a leak detector for helium and hydrogen. Only air and clean gases must be used with the PHOENIX L300i. The leak detector is not suitable for - pumping liquids or gases containing dust...
This function can be chosen automatically for the PHOENIX L300i MODUL. Every time the unit changes into standby mode the purge starts automatically for 20 seconds.
Page 38
5. Press the START button. The inlet will be evacuated and if the inlet pressure drops below 15 mbar a measured leak rate will be displayed. 6. Press the STOP button, the PHOENIX L300i will go to standby. If you press STOP a few seconds the inlet of the PHOENIX L300i will be vented.
20. Remove the test object and put on a blind flange on the inlet port. Switch off Switch off 21. Switch off the PHOENIX L300i if the unit is in STANDBY or VENTED mode by using the mains switch Fig.
Page 40
Fig. 12 /9 is the communication interface to the operator. It displays the leak rates, the status report of the PHOENIX L300i, messages, warnings and errors. With the soft keys no.1 to no. 8 various functions which are shown in the...
Page 46
P1 Pirani PHOENIX L300i The inlet pressure P1 of the PHOENIX L300i will be shown logarithmic. Fig. 17 P2 Pirani PHOENIX L300i The fore vacuum pressure P2 of the PHOENIX L300i will be shown log- arithmic. Fig. 17 P1 (L200) The setting for the inlet pressure P1 is identical with those of the L200, i.e.
Page 49
Open when emission is on. Define PLC inputs Define PLC inputs Main menu > Settings > Interfaces > Define PLC inputs These inputs can be used to control the PHOENIX L300i with a programmable logic control (PLC). Maximum input voltage 35V. Caution...
Main menu > Settings > Interfaces > PLC Sample rate Softkey 2: Decreasing the PLC sample rate down to the minimum of 10 ms. This might be necessary if exchanging an L200 to the PHOENIX L300i to stay compatible. Softkey 3: Increasing the PLC sample rate to the maximum of 100 ms.
Switch off the PHOENIX L300i and turn it on again. During the run-up phase press the key no. 3 or 7 so long until the display can be read properly again. This setting is saved to the EPROM only after confirming this through the contrast menu.
Sources for residual gas are air or absorbed gases from the inner surfaces of the PHOENIX L300i. This internal background will never dis- appear totally. Very clean systems which have been pumped for a long time will show a background in the 10 mbar l/s range.
Trigger 1, 2 and Trigger 3 are programmable switching thresholds. When these thresholds will be exceeded the PHOENIX L300i reacts as follows: Display In the status line of the display the signs for Trigger 1, 2 and Trigger 3 are displayed inverted if the leak rate exceeds (becomes higher than) the programmed value.
Setting for the machine factor Softkey 8: Leak rate internal test leak Setting for the internal test leak The menu for the PHOENIX L300i DRY version allows the following vacuum set- Vacuum settings PHOENIX tings which varies to the PHOENIX L300i: L300i DRY...
Page 66
Main menu > Settings > Vacuum settings > Automatic purge Through this menu is it possible to start automatic purge for 20 seconds automatically. This setting is only possible for the PHOENIX L300i MODUL (refer to vacuum set- tings PHOENIX L300i MODUL) Softkey 3: The function automatic purge is off.
Page 67
This setting is only active in vacuum mode (see Chapter 3.5). Softkey No. 2: GROSS only In this mode the PHOENIX L300i remains at the inlet flange after falling below 15 mbar. When the pressure is increasing over 15 mbar the PHO- ENIX L300i switches automatically into evacuation mode.
Page 68
Operation If the PHOENIX L300i or the PHOENIX L300i MODUL is used with a par- tial flow system this vacuum mode must be enabled before. The setup for the partial flow mode is described in partial flow setup Softkey No. 7: Normal (default settings) This is the default setting.
Page 69
Operation The PHOENIX L300i MODUL allows additionally the settings for the forepump (oil sealed or dry) and selectable pumping speed for the forepump. Partial flow setup for the PHOENIX L300i. Partial flow setup PHOENIX L300i Softkey 2: The entry of the nominal pumping speed of the partial flow pump can be decreased.
Page 70
Settings > Vacuum settings > Vacuum ranges with softkey no. 6 partial flow enable and confirm with the soft key OK. Softkey 2: Pump setup Setting for the forepump if the PHOENIX L300i MODUL is operated with a partial flow system. Softkey 7: Partial flow setup Options for setting up the partial flow system.
Page 71
At T = endless valve V10 will open when pressing start. At an inlet pres- sure p1 < 15 mbar the PHOENIX L300i switches to measurement mode and display leak rates. This is recommended if it is acceptable to wait for a while until measurement mode is reached and leak rate reading at high inlet pressures are not needed.
Page 72
Setting between the default value 1 for the sniffer line SL300 or the cor- rection factor (1000) for the use of the Quicktest. For the use of a PHOENIX L300i with a Quicktest the setting for the Quicktest Quicktest setting under Main menu >...
If the use of the PHOENIX L300i DRY needs the possibility to shut off this function, fore ex. because of a high helium background in the ambient and no option to connect a hose line with fresh air to the gas ballast port, this function can be shut of here.
Page 74
„Calculate inlet area background“. Softkey 7: internal only (default) With start the PHOENIX L300i defines the internal offset (background) and subtracts this value from the leak rate signal, so that just the leak rate is shown in the display. This setting should be used as standard setting for the PHOENIX L300i.
Operation 3.10.3 Mass Main menu > Settings > Mass The requested mass of the measured gas can be selected. The PHOENIX L300i must be in standby mode for changing to another mass. Softkey 2: H (2 amu) Hydrogen with the mass of 2 amu will be measured.
Page 76
Setting for the service interval of the forepump. This setting depends on the use of the PHOENIX L300i but latest after 4000 running hours or one year the oil in the pump should be controlled. See also manual for the Trivac D2,5E which is included in the document folder.
Page 77
Here you can enter the service interval for the exhaust oil filter. This setting is only filter possible for the PHOENIX L300i. This setting depends on the use and application of your PHOENIX L300i and therefore no recommendations can be given (see chap.
Page 80
Decrease change over threshold GROSS-FINE Selectable between 0,2-0,05 mbar (Default value 0,2 mbar). For the PHOENIX L300i DRY and the PHOENIX L300i MODUL the changeover is between 0,1-0,05 mbar (Default value 0,2 mbar). Softkey No. 4 Adjustment for ARGON Selection between air or Argon Softkey No.
Page 81
Softkey No. 7: Increase change over threshold GROSS-FINE Selectable between 0,05 - 0,2 mbar, respectively 0,05 - 0,1 mbar for PHOENIX L300i DRY and PHOENIX L300i MODUL Pressure limits for sniff Pressure limits for sniff mode mode Main menu > Settings > Monitoring functions > Pressure limits for sniff mode This function is automatically activated in sniff mode.
Information on measured internal data is provided on 10 screens. Softkey 4: Vacuum diagram The vacuum diagram of the PHOENIX L300i is shown. Here you can see which valves are opened or closed momentarily. See chapter 4.1.1 Softkey 5: View error list The list of occurred errors and warnings will be displayed.
Calibration 3.13.1 Introduction The PHOENIX L300i can be calibrated in two different ways: Internal calibration by means of a built-in test leak external calibration by means of an additional test leak which then is attached to the inlet port or the component under test.
Although the PHOENIX L300i is a very stable inst- rument a calibration is recommended every day with heavy use, or before using the PHOENIX L300i from time to time, to make sure that ambient temperature changes or dirt or other impacts don’t adulterate the measurements.
3.14 Switching Off/Shutting Down The PHOENIX L300i can be switched off any time by using the mains switch. The turbo pump will be decelerated automatically. It is recommended to put the leak detector into standby and vented mode. Approximately after 30 seconds the turbo pump is decelerated sufficient to move the PHOENIX L300i.
Only Leybonol LVO 310 (Cat. no. L31001) must be used in the TRIVAC D2,5E in the PHOENIX L300i. The monthly interval for the check is just a nominal period. If the PHOENIX L300i is used heavily, in particular in sniffer mode, then this check should be performed more frequently.
Maintenance Fig. 41 Opening of the mechanical hood 4. Turn the PHOENIX L300i so that it is orientated in the same way as shown in Fig. 5. Use two flat blade screw drivers and insert these into the openings (Fig.
4.4.4 Cleaning The housing of the PHOENIX L300i is made of painted plastic parts. Thus for the purpose of cleaning, only such agents should be used which are generally also used for other painted or plastic surfaces (mild household cleaning agents, for example).
Page 93
6. After having exchanged the fuse (s) press the lid of the mains socket firmly back 7. Insert the mains cord into the PHOENIX L300i and switch the instrument on. Beside these mains fuses several internal circuits are fused separately. These fuses are listed in the following table.
Fig. 42 Assembly fuses 4.4.6 Exhaust Oil Filter After using the PHOENIX L300i for a longer time there can be oil accumulated from the pump. In this case please do the following: 1. Switch off the PHOENIX L300i. 2. Remove the mechanical cover according to Chapter 4.4.1.
Calibrated Leak TL7 The Calibrated leak TL7 with the helium reservoir is used for alignment of the mass spectrometer in the PHOENIX L300i as well as for calibration the leak rate indica- tion. It is equipped with a solenoid valve which is actuated via the control electronics of the PHOENIX L300i.
The PHOENIX L300i will indicate a message on the LC display and will remain in the standby or the measurement mode. Warning messages will remain on the LC display until the warning has been acknowledged by pressing „OK“...
Page 99
W46 Emission for cathode 2 can can Signal MSIBE on MSV board is not active. Emission for cathode 2 can not be not be switched. switched on. PHOENIX L300i switches to cathode 1. Order a new ion source. Cathode 2 is defective.
Page 100
Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of the PHOENIX L300i Fan defective: Replace the fan Ambient temperature too high: Feed cooler air to the PHOENIX L300i Bearing defective: Inform Leybold service for repair E67 TMP error: Short circuit in TMP- Short circuit in the pump’s motor or the connecting cable...
Page 101
Maximum temperature for the converter has been exceeded. ter too high (>75°C) Ambient temperature too high: Feed cooler air to the PHOENIX L300i Fan defective: Replace the fan Fore vacuum or high-vacuum pressure to high: reduce the inlet pressure of...
Page 112
Sales and Service Germany Great Britain America Leybold Japan Co., Ltd. Tsukuba Technical Service Center 1959, Kami-yokoba Leybold UK LTD. Leybold GmbH Tsukuba-shi, Ibaraki-shi 305-0854 Unit 9 Sales, Service, Support Center (3SC) Japan Silverglade Business Park Bonner Strasse 498 Leybold USA Inc.
Need help?
Do you have a question about the PHOENIX L300i and is the answer not in the manual?
Questions and answers