Specifications - Arizant Ranger 245 Operator's Manual

Blood/fluid warming system
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Ranger Blood/Fluid Warming System

Specifications

Physical characteristics
WARMING UNIT
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)
CLASSIFICATION
Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as
Class I, Type B, Ordinary Equipment. Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous oxide. Classified by Underwriters
Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance
with UL 2601-1 and in accordance with Canadian/CSA C22.2 No. 601.1.
Electrical characteristics
INPUT VOLTAGE
100-120 or 220-240 VAC
OPERATING FREQUENCY
100-120 VAC, 50/60 Hz
220-240 VAC, 50 Hz
MAXIMUM HEATING POWER
900 W
FUSE
10 Amp
Leakage current
Meets leakage current requirements in accordance with UL 2601-1 and IEC 601-1.
Environmental conditions
OPERATING TEMPERATURE RANGE
15° to 40°C (59° to 104°F)
STORAGE TEMPERATURE RANGE
-20° to 45°C (-4° to 113°F)
OPERATING HUMIDITY
10 to 85% RH, noncondensing
ATMOSPHERIC PRESSURE RANGE
50 kPa to 106 kPa
Temperature characteristics
SET POINT TEMPERATURE
41°C
-
OVER
TEMPERATURE ALARM
43°C
-
UNDER
TEMPERATURE ALARM
33°C
-
OVER
TEMPERATURE CUTOFF
46°C
11