Valid for serial number: FS-02555 - (2004-06-03). Revision history for this document. Revision AK0. 2004-05-10, AnNy, document created. Revision AK1. 2017-05-17, AnNy, text corrected. Document properties. Last saved: 2017-05-17 Filename: Technical Manual 90-20-DA-36 AK1.doc Author. Created by Anders Nyström. Last saved by Ewa Bartłomiejczyk...
– Disconnecting the line cord – Clearly marking the product to prevent its further operation – Informing your Pendulum representative. For example, the product is likely to be unsafe if it is visibly damaged.
General description Functional The DA-36, Fiber & Coax Distribution Amplifier is intended to extend distance and galvanic isolate distribution of 10MHz reference frequency. Fiber optic distribution. In coaxial distribution nets, there is always a risk for unwanted ground loops and other types of EMI. Also distances are limited to a few meters depending on environment and the quality of the coax cables used.
If both LED’s are green, the coax input signal is internally blocked and the fiber input signal is switched to the outputs, both fiber and coax outputs. If both LED’s are turned off , there is no power supplied to the DA-36. Grounding...
Wall Mounting Bracket 50-65-1673 Screw, MFX-H M3x5 FZB 50-65-5030 Rubber feet 50-65-5101 Isolating washers for BNC 60-00-5538 50-65-5128 Power cord 90-20-DA-36 This manual DA-36, (Part number includes all parts in list above). Wall mounting kit, rubber feet and isolation washers.
Installation. Serial number. The products serial number is the best way for Pendulum Instruments to identify the product. If the serial number is not noted on your delivery notes, please add the serial number to your own product documentation. This will be useful at future contact with Fiberdata System.
Make sure that the source fiber optic transmitter, marked Tx, is connected to the remote units fiber optic receiver, marked Rx. Functional earth/ground. Functional ground Connect to the functional ground screw if DA-36 chassis needs to be referenced to ground. This connection must be low impedance also at RF-frequencies. BNC-ports All BNC-ports are referenced to DA-36 chassis at delivery.
Page 11
Loosen nut at BNC – 14mm nut. Add one of the isolation washers delivered with the DA-36. Isolation washer The isolation washer shall be added closest to the panel. Reassemble the washers and the nut in reverse order. Isolation washer...
The DC-cable from the power supply shall be connected to DA-36. Connect power supply AC-cord to mains. Allow DA-36 to warm-up for some minutes. Wait to connect fiber- or coax- signal to input until both Input Status indicators are red.
If both LED’s are green, the coax input signal is internally blocked and the fiber input signal is switched to both fiber and coax outputs. If both LED’s are turned off/black , there is no power supplied to the DA-36. Note that the Input Status indications do not evaluate the quality of the input...
Service Sales and Service office For additional product information, customer support and service, please contact Pendulum Instruments at the following address: Altaria Services Sp. z.o.o ul. Lotnicza 37 80-298 Banino Poland Visiting address: as above Delivery address: as above Phone:...