1 Image, Symbol & abbreviation..........................2 2 Product introduction............................2 2.1 Product features and specifications......................2 2.2 Application, Contraindication and notices....................3 2.3 Main structure............................... 4 2.4 Working principle and characteristics......................5 2.5 Main performance index..........................1 2.6 Normal operating condition..................错误!未定义书签。 3 Installation &...
Preface “ HR-OF808 ” is a dedicated device designed for hair removal in a safe and efficient way. Please read this User Manual very carefully before operation and keep it for future reference. “ HR-OF808 ” must be operated by qualified doctors who have been trained by our ...
Image, Symbol & Abbreviation Image or symbol Meaning BF Equipment Danger, high voltage Attention!Refer to the documents coming along with the machine. Waterproof IPX1 Laser Fragile Object This way up. Away from rain. Manufacturer Production Date Disposal of Discarded Electronic Items Emergency Laser Stop 2 Product introduction 2.1 Product features and specifications...
Operator should have got through training courses offered by supplier. c) Clean the treatment area thoroughly, and shave the hair. “HR-OF808” d) Do not pull out hair in the way like waxing prior to use within 6 weeks; or...
the laser therapy does not give best results. e) Avoid sun burning especially for dark skin. It is highly suggested to use sun lotion for about 4-6 weeks before treatment or use some medicine like hydroquinl. f) Try to avoid sun exposure to decrease rate of side effects. g) It is good to take test treatment on small skin site first and observe for 15 minutes before taking large area treatment.
Figure 2.3.1 2.4 Working principle and characteristics 2.4.1 Working principle “ HR-OF808 ” working principle is based on the photothermic theory. There is plenty of melanin in hair follicles and hair shafts. Melanin scatters among hair bulbs and hair shaft “...
Page 8
3.1.2 Assemble methods a) Lift main machine up from below, instead of forcibly pulling it out by holding the hanger. b) As shown Figure 3.1.2.1, put the hand piece onto the hanger. c) Show as figure 3.1.2.2,connect the foot switch to the corresponding socket “ FOOT SWITCH”...
Page 9
"Welcome" shows on screen interface after powering on(see Figure 3.3.1-1) ,your logo can be added here. The system enters equipment interface automatically 10 seconds later. (see Figure 3.3.1-2). Press " ",the system enters equipment settings interface. (see Figure 3.3.1-3) Figure 3.3.1-1 Figure 3.3.1-2 Touch "LANGUAGE"...
Page 10
before sending! So customers don't need to set the "factory settings". Chose the mode to go to the next operation interface,as Figure 3.3.1-9. 3.3.1-9 Chose the “Skin Type”“Hair Color” and “Hair Texture” depending on your customer’ s actual situations. Then press “Next” enter to the final interface, as Figure 3.3.1-10 3.3.1-10 3.3.2 Operation interface and parameters setting 3.3.2.1 working mode setting...
Page 11
Skin Type I Skin Type II Skin Type III Skin Type IV Skin Type V Skin Type VI 40ms 80ms 200ms 200ms 300ms 400ms black 10Hz 4-6 8-12 2Hz 26-30 1Hz 28-32 1Hz 40-44 1Hz 30-34 J/ cm J/ cm J/ cm J/ cm J/ cm...
Manual mode: user could turn on or turn off cooling system according to actual situation. 3.3.2.6 Number of Flash times The screen would show the number of the flash times since the device is on. Two windows: one is the counter for the single treatment time;the other one is the total number. 3.3.2.7 Ready/Standby Enter Ready status: The device is defaulted on Standby status when it is turned on.
4.3.2 Please keep touch screen free from drugs or other dirt. 4.3.3 Please keep touch screen free from wet goods. 4.3.4 Don’t scrub touch screen by alcohol or washing powder etc. If needed please scrub it slightly by soft dustless cotton fabric. 4.3.5 Don’t move touch screen and stretch data wire.
Swelling will disappear in 24 h and red will fade in 48 h. There is scab after the operation and pruritus can happen during the operation which all are normal phenomenon. Please pay attention to keep skin humid. And skin maybe peeled slightly when more than 2 times treatment caused by skin’s metabolism, which is normal phenomenon.
Page 16
specification of fuse is T.a.c.250V/10A and when power is a.c.110V/a.c.120 specification of fuse is T.a.c.250V/20A. Put fuse holder into socket. 1、Prize up 2、Pull out fuse box 3、Exact fuse holder 4、 Change new holder 7.7 Water circulation system and water shortage alarm 7.7.1 Water circulation system alarms due to water shortage in water tan.It is forbidden to use laser probe head before filling pure water in water tan.
Page 17
Overflow Water Level Inlet / Outlet 7.7.2.4 7.8 Cold water, hot water system defect alarm In case of cold or hot water system defect alarm shown on screen, it means water flows too slowly or water temperature is too high which is not up to working condition standard. In this case, it is forbidden to use laser probe head.
8 Warranty 8.1 Warranty term We provide one year guarantee from the purchase day. 8.2 Warranty rang We provide free maintenance or replacement in warranty term but exclude the damage caused by abnormal use and force majeure. 8.3 Warranty limits 8.3.1 User should provide purchase invoice and warranty card when applying free warranty service in warranty term.The initial date of products ’...
e. The faults and damages caused by un-prescribed environment. f. Improper keeping( damages caused by mouse and liquid inleakage). g. The machine can’t be operated normally caused by using self-organizing or non-public procedures. 8.3.8 The machine’s lifespan is 5 years. 8.3.9 Please inquiry our customers service website or contact us for further information of machine’s technical issue.