10.3 Overview

Factory settings are marked by *.
Menu block
Menu item
F1 Unt
Weighing Units
EL AU*
F2 bl
Display
EL on
background
illumination
EO oFF
F3 Com
Interface
S 232
parameters
S USb
PFB-BA-e-1741
Available settings / explanation
g, ct, lb, oz, d, yn, ozt, dwt, mom, tl h, tl c, tl
t, t, bt, n
Automatic background illumination on when
weighing pate is loaded or key pressed.
Background lighting of display is switched on
permanently
Display background illumination off
Use
to select the interface: RS232 or USB
P Prt*
Output of stable weight value after pressing
F4 Acc"  „off")
("
Manual totalizing, see chap.9.3.1
„F4 Acc"  „on")
(
Press
added to the summation memory and
issued.
P Cont
Continuous data output
P AUto
For automatic totalizing see chap. 9.3.2
This function is used to issue and add individual
weighing values automatically to the summation
memory on unloading of weighing scale.
wirel
Not documented
.
and the weighing value will be