Cerlic FLX Manual

Level / flow sensor

Advertisement

Quick Links

FLX
FLX
Level / Flow
Sensor
2009-09-14
C70B5EN10

Advertisement

Table of Contents
loading

Summary of Contents for Cerlic FLX

  • Page 1 Level / Flow Sensor 2009-09-14 C70B5EN10...
  • Page 2: Table Of Contents

    Starting Up ..........................7 7. Removing the sensor ................. 7 8. Maintenance ....................8 9. Sensor display ................... 8 10. Menu for FLX sensor ................. 9 Level measurement ........................11 Parshall flume ........................... 11 Thompson Weir ........................12 Rectangular Weir ........................12 Rectangular Weir with End Contractions .................
  • Page 3: Introduction

    BB2 control box. 4. Measuring principle The FLX uses a pressure sensor to measure flow as a function of the level of the liquid in open channels. Different types of standardized weirs or flumes are used to achieve measurable rates of flow.
  • Page 4: Unpacking The Sensor

    Damages If damages occurred during shipment, immediately contact UPS or other truck line as well as your Cerlic representative. The shipment can be returned only after contact has been made with Cerlic. Packaging The original packaging is designed to protect the equipment and should be used for storage or if the goods must be returned.
  • Page 5: Mounting

    It is important that the sensor, once it has been removed, is re-positioned at the right height. An adjustable slide rail mounting is available as an option. In strong current, hydrodynamic effects may influence the measurement; contact Cerlic service for advice on current protection if problems occur.
  • Page 6: Mounting A Sensor In A Weir

    Mounting a Sensor in a Weir In a weir, the sensor should be placed at 4 times the maximum measuring height upstream from the overflow edge. If possible, measuring in a measuring well is preferably. However, this is not strictly necessary since the sensor ends up so far from the overflow that its interference with the flow profile is negligible.
  • Page 7: Starting Up

    If the sensor is held vertically in the air the level shall read zero (or close to zero). The FLX is pre-set at the factory, but sometimes the zero point still needs adjustment. (Refer “Adjusting the zero point”). NOTE, if an offset is entered in the sensors setting menu, the level shows the value of the offset instead of zero.
  • Page 8: Maintenance

    All sensors age, which eventually leads to inaccurate measurements. Check the sensor's zero point about once every six months by lifting the sensor into the air and observing the level the FLX registers. The level is shown in the sensors info display (opened by simultaneously pressing and ENTER).
  • Page 9: Menu For Flx Sensor

    10. Menu for FLX sensor to select the sensor in the main display. The menu for the selected sensor is accessed by pressing ENTER for five seconds. If the selected sensor is not active (the text No transmitter is shown) a warning is displayed that asks you to make another choice in order to show the sensor menu.
  • Page 10 MaxTemp The highest temperature the sensor has been exposed to, read only Info Press “ENTER” to go to “info” read only menu. This menu is for Cerlic internal use. Sa 0 Raw value for the zero level Sa 1 Raw value for the calibrated level...
  • Page 11: Level Measurement

    Level measurement FLX can be configured for level measurement by selecting the unit ”mm”or”inch” in the sensors settings menu. Type of flume, flume parameters and number of flumes selected are ignored. Preset, pulse relay and totalizer does not work for level measurement.
  • Page 12: Thompson Weir

    Thompson Weir When you select a Thompson weir, you must specify the following parameters: ANGLE The weir's α-angle (see the illustration Angle 20.0 - 150.0 ° above) Rectangular Weir When you select a Rectangular weir, you must specify the following parameters: WI i SILL Sill...
  • Page 13: Rectangular Weir With End Contractions

    Rectangular Weir with End Contractions When you select a Rectangular weir with end contractions, you must specify the following parameters: WI o WI i SILL Sill 50 - 8000mm The sill height of the weir (see the illustration above) (5.0 - 800.0 ") 50 - 32 000mm The inner width of the weir (see the WI i...
  • Page 14: Cipoletti Weir

    Cipoletti Weir When you select a Cipoletti weir, you must specify the following parameters: WI i WI i 50 - 32 000mm The inner width of the weir (see the illustration above). (5.0 - 3200.0 ") Sutros Weir When you select a Sutros weir, you must specify the following parameters: WI i Re h 0 - 4000mm...
  • Page 15: Venturi Flume With Rectangular Throat

    Venturi Flume with Rectangular Throat When you select a Venturi flume with rectangular throat, you must specify the following parameters: Sill 0 - 4000mm Sill height of the flume (see the illustration above). (0 - 400.0 ") WI i 50 - 32 000mm The throat width of the flume (see the illustration above).
  • Page 16: Venturi Flume With U-Shaped Throat

    Venturi Flume with U-shaped Throat When you select a Venturi flume with U-shaped throat, you must specify the following parameters: Sill 0 - 4000mm Sill height of the flume (see the illustration above). (0 - 400.0 ") 50 - 32 000mm The throat width of the flume (see the WI i illustration above).
  • Page 17: Calibration

    Zero calibration The FLX has a zero point calibration function, which is based on lifting the sensor into the air (zero pressure) and pressing a key. The BB2 registers the signal from the sensor and bases all subsequent measuring on that signal level.
  • Page 18: Scaling

    20.625”. Subtract 0,5” = 20.125”. Enter this value in “Cal level”.) • Exit the menu. • In the display mode of BB2, press ARROW DOWN + ENTER to get to the FLX information window. • Check that the level reading of the FLX corresponds to the water level in the bucket as per above.
  • Page 19: Resetting The Overflow Counter

    • Exit the menu. 14. Totalizer FLX has a built in totalizer, that registers the volume in m3 if metric units are used or thousands of Gallons if US units are used. The value of the totalizer is shown in the sensor information display.
  • Page 20: Technical Data Flx

    15. Technical data FLX FLX 1m P/N 11305551 Material Tube and electrode holder: SIS2343 (316SS) ∅28 x 1085 mm (∅ 1.1” x 43”) Dimensions Weight 1,8 kg (4 lbs) Process connection Immersion sensor Max immersion depth 94 cm (37“) Max temperature 60°C (140°F)
  • Page 21 Certificate of conformity: The FLX sensors along with their central unit BB2 are in conformance with the following EC Directive(s) when installed in accordance with the installation instructions contained in the product documentation: 73/23/EEC Low Voltage Directive as amended by 93/68/EEC...
  • Page 22: Dimensions

    Dimensions - 22 -...
  • Page 23: Appendix 1, Optional Slide Rail

    Appendix 1, Optional slide rail - 23 -...
  • Page 24: Appendix 2, Support Information

    Appendix 2, Support information Before calling Cerlic Support, please collect the following information and have it at hand. Company _____________________________________________________ Name _____________________________________________________ Phone _____________________________________________________ E-mail _____________________________________________________ Sensor Type _____________________________________________________ Position / Tag _____________________________________________________ First go to the BB2 menu, it is accessed by pressing and ENTER at the same time for five seconds.
  • Page 25: Appendix 3, Setup Information

    _____________________________________________________ Leave the BB2 menu by pressing and ENTER at the same time. Cerlic Controls AB, P.O. Box 5084, SE-141 05 Kungens Kurva Tel: +46 8 501 694 00, Fax: +46 8 501 694 29, sales@cerlic.se www.cerlic.se - 25 -...

Table of Contents