Safety Considerations Before Initial Operation
•
Check instrument for damage caused in transit.
•
The telescopic tube
•
Do not use the outer housing
need to carry the instrument, hold it by the base of the instrument.
•
Attach discharge tube
damage.
•
When screwing on/off the bottle do not hold the instrument at its outer
3
housing
, but at the screw base.
•
Do not pump before the instrument has been assembled completely and a
collecting vessel has been placed underneath.
Operating Limitations
Do not use for:
•
PTFE swelling solvents
•
Hydrofluoric acid
•
Chemical solutions that react with platinum-iridium alloys
Liquids used with DISPENSMAN must not exceed the following limits:
OPERATION LIMIT
Vapor pressure
Viscosity
Density
Temperature
Liquid Contact Materials
The components having direct contact with the reagent are made of chemical
resistant materials: platinum spring, ceramic valve balls, PTFE piston, and borosilicate
glass 3.3 cylinder.
LT801577_D _UG DISPENSMAN.indd 7
21
should reach the bottom of the bottle.
3
for carrying the assembled dispenser. If you
9
and clip on protective sleeve
8
carefully to avoid
MAXIMUM
600 mbar
500 mm
/s
2
2.2 g/cm
3
<+15°C and >+40°C
<+59°F and >+104°F
|
DISPENSMAN
®
USER'S GUIDE