Specifications - Gilson dispensman User Manual

Sophisticated dispensing system that fulfills increasing requirements for quality, precision, and reproducibility in laboratories
Table of Contents

Advertisement

Noise During Aspiration and Dispensing
Noise while aspirating or dispensing is acceptable as it is because of the high quality
suction valve with valve ball.
If any problem persists, please return DISPENSMAN back to your local Gilson
service center.
by
ater
Return for Repair
of the
NOTICE
Forced opening of the instrument invalidates any warranty claim. Enclose with
the returned instrument a description of the trouble that has occurred and
specify which reagents were used.
t,
rmly
DISPENSMAN is a high quality dispenser that offers excellent accuracy and
se
precision; it is fully compliant with ISO 8655.
Each dispenser is inspected and validated by qualified technicians. Gilson
declares that its manufactured dispensers comply with the requirements of the
ISO 8655 standard, by type testing. The adjustment is carried out under strictly
defined and monitored conditions.
ter 4
MODEL
in its
DISPENSMAN 2.5 mL
DISPENSMAN 5.0 mL
DISPENSMAN 10 mL
DISPENSMAN 25 mL
DISPENSMAN 50 mL
LT801577_D _UG DISPENSMAN.indd 15
Please clean the instrument carefully, as described in Chapter 6
CLEANING &
STERILIZATION, page 12. Never send in instruments
filled with reagents. Returned instruments can only be
checked and repaired if they have been carefully cleaned and
decontaminated by the customer.
Chapter 9

SPECIFICATIONS

PART
VOLUME RANGE
NUMBER
F110101
0.25 – 2.5 mL
F110102
0.5 – 5 mL
F110103
1 – 10 mL
F110104
2.5 – 25 mL
F110105
5 – 50 mL
MAXIMUM PERMISSIBLE ERRORS
SCALING
GILSON
(mL)
SYSTEMATIC
RANDOM
ERROR (mL)
ERROR (mL)
≤ 0.002
0.05
±0.012
0.10
±0.030
≤ 0.005
0.20
±0.060
≤ 0.010
≤ 0.025
0.50
±0.150
1.00
±0.300
≤ 0.050
DISPENSMAN
ISO 8655-5
SYSTEMATIC
RANDOM
ERROR (mL)
ERROR (mL)
≤ 0.010
±0.030
±0.030
≤ 0.010
±0.060
≤ 0.020
≤ 0.050
±0.150
±0.300
≤ 0.100
|
®
USER'S GUIDE
09/11/2017 15:01:15
15

Advertisement

Table of Contents
loading