Item
Condition
6. Modulation
1) Set test mode
balanced
MIC input : OFF
Select "BAL" in tuning mode. meter
"L BAL"
Deviation meter filter
LPF : 15kHz
HPF : OFF
De-emphasis : OFF
PTT : ON
2) "C BAL"
PTT : ON
3) "H BAL"
PTT : ON
7. Maximum
1) Set test mode
deviation
Connect AG to the MIC
terminal.
Select "MAX" in tuning mode.
"L MAX"
AG : 1kHz/50mV
Deviation meter filter
LPF : 15kHz
HPF : OFF
De-emphasis : OFF
PTT : ON
2) "C MAX"
PTT : ON
3) "H MAX"
PTT : ON
8. MIC
1) Set test mode
sensitivity
CH : CH1 - Sig1
check
AG : 1kHz/5mV
PTT : ON
9. QT deviation 1) Set test mode
Select "FQT" in tuning mode.
"L FQT"
Deviation meter filter
LPF : 3kHz
HPF : OFF
PTT : ON
2) "C FQT"
PTT : ON
3) "H FQT"
PTT : ON
ADJUSTMENT
Measurement
Test-
Unit
Terminal
Unit
equipment
Power meter Rear
ANT
Deviation
panel
Oscilloscope
AF VTVM Front
MIC
AG
panel
TK-880/880H
Adjustment
Parts
Method
Make the de-
(Wide/Narrow)
modulation
waves into square
waves.
± 50Hz (Wide/Narrow)
3.8kHz (Wide)
1.75kHz (Narrow)
(According to the
larger +, –)
± 3kHz± 0.2kHz (Wide)
Check
± 1.5kHz± 0.05kHz (Narrow)
± 50Hz (Wide/Narrow)
0.75kHz (Wide)
0.35kHz (Narrow)
Specifications/Remarks