Adjusting An Attenuator Probe - Philips pm3232 Instruction Manual

10 mhz dual-beam oscilloscope
Table of Contents

Advertisement

100
3.5.3.
1.
Technical data
Attenuation
1:
10
±3
%
Input
impedance
10
MOhm//10
pF
Max.
permissible input voltage
1000 Vp.p
Maximum
d.c.
component 500
V
with
the
blocking
capacitor included
into
the
circuit.
3.5.3.
2.
Adjustment
(Fig.
3.25.)
Unscrew
the
probe
locking
nut by
turning
it
anit-clockwise
Rotate the probe
body
to
change
the capacitance while
watching
the display for the desired
waveform
When
compensation
is
completed,
carefully turn the locking
nut clockwise,
to
lock
it
without
disturbing
the
adjustment
The
attenuator
probe
causes distortion
if it
has not
been
properly
adjusted.
over
compensated
correct
under
compensated
Fig.
3.25.
Adjusting an atttenuator
probe