Z Modulation; Output Sockets - Philips pm3232 Instruction Manual

10 mhz dual-beam oscilloscope
Table of Contents

Advertisement

85
3.2.10.
Z
MODULATION
Depress
switch
SK4, +
switch
SK8,
AUTO
switch
SK10
and
BEAM SELECTOR
switch
A
SK14
Apply
a
positive-going
square-wave
voltage of
20
V,
repetition
rate
100
Hz, to
Z
MOD
socket
BU7
Apply
a signal
which
is
synchronous
with the
former square-wave
voltage
to Y/\ Input socket
BUS
Check
that the
trace
is
blanked during
the positive pulse
3.2.11.
OUTPUT SOCKETS
3.2.11.1. Calibration voltage
on
CAL
socket
BUI
(R761)
Check, with
the
aid
of an
accurate
peak-peak
voltmeter, that the voltage
on
CAL
socket
BUI
has
a
value
of
600
mVp.p
+
or
0,7
%;
if
necessary, readjust
potentiometer
R761
(Fig.
3.12.)
Check
that the pulse
does not
show
overshoot and
that the repetition
frequency
is
2
kHz +
or
20
%
3
2.11.2.
Time-base
voltage
on
TB
OUT
socket
BU6
Depress
AUTO
switch
SK10
Set
TIME/cm
switch
SK2
to
2
ms
Check
that the
amplitude
of the
sawtooth
voltage
on
TB
OUT
socket
BUG
is
6
V
+
or
20
%
86
Fig, 3, 12,
Adjusting
Fig,
3,
U.