Commands - Overview - H&B DAD 141.1 Technical Manual

Load cell digitizing unit communication via rs 422/485 & ethernet port
Table of Contents

Advertisement

Commands – Overview
9.
Command Short description
AA
Get/set analog output action (base)
AD
Communication: Device Address
AG
Absolute gain calibration
AH
Get/set analog high
AI'n'
Assign Input 'n' to 1 out of 15 different functions/base
AL
Get/set analog low
AM
Analog Output Mode Current / Voltage
A'n'
Action for Setpoint 'n': choose 0 of 8 different base for setpoint 'n'
AS
Save analog output parameters
AZ
Absolute zero point calibration
BR
Communication: Baud Rate
CE
Calibration: Open Calibration Sequence; Read TAC Counter
CG
Calibration: Set Calibration Gain (Span) at Load > Zero
CI
Calibration: Minimum Output Value
CL
Communication: Close Device
CM
Calibration: Set Maximum Output Value
CS
Save the Calibration Data (CM, CI, DS, DP, etc.) to the EEPROM
Calibration: Set Calibration Zero Point – Scale Without Load
CZ
DP
Calibration: Set Decimal Point Position
DS
Calibration: Set Display Step Size
DX
Communication: Set full-duplex (1) or half duplex (0)
FD
Factory default settings: Write Data to the EEPROM (TAC protected)
FM
Read / modify filter mode: IIR (0) or FIR (1)
FL
Digital low pass filter: Filter Cut-off Frequency
GA
Output: Get Triggered Average Value
GG
Output: Get Gross Value
GH
Get Hold Value
Retrieves an image file from the DAD141.1's EEPROM
GI
GN
Output: Get Net Value
GM
Get Peak (Maximum) Value
GO
Get Peak tp Peak Value
GS
Output: Get ADC Sample Value
GT
Output: Get Tare Value
GV
Get Valley Value
Output: Get Data String "Net/Gros/Status"
GW
H'n'
Hysteresis for Setpoint H0 (S0) or H1 (S1) or H2 (S2)
HT
Trigger function: Hold time for Violation of Setpoint Limit
ID
Device information: Identify Device
IH
Hardware version
IN
Logic input: for each Input Status 0 or 1
IO
Logic output: for each Output Status 0 or 1
IS
Device information: Identify Device Status
IV
Device information: Identify Firmware Version
MT
Trigger function: Measuring Time for Averaging
NA
Network Address <aaa.bbb.ccc.ddd>
NR
Motion detection: No-motion Range
NT
Motion detection: No-motion Time Period
DAD 141.1 Technical Manual, Rev. 2.5 – April 2015
Parameter value
Page
0 through 8
48
0...255
47
± 32000
34
-999999 to 999999
48
0 to 15
43
-999999 to 999999
48
0 to 5
48
0 to 8
45
none
49
± 32000
34
9600...115200 baud
47
0...65535
31
1...999999
32
–999999...0
31
None
47
1...999999
31
None
34, 49
None
32
0...5
32
1, 2, 5, 10, ..., 500
32
0 or 1
47
None
33
0 or 1
36
0...8
36
None
41, 50
None
40
None
41
None
49
None
40
None
41
None
41
None
40
None
40
None
41
None
40
-9999...+9999
45
0...65535 ms
46
None
30
None
30
0000...0011
43
0000...0111
44
None
30
None
30
0...3000 ms
50
e.g. 192.168.0.100
47
0...65535 d
35
0...65535 ms
35
Page 28 of 54

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for H&B DAD 141.1