KERN EWJ 300-3 Operating Instructions Manual page 73

Precision balances ewj series
Hide thumbs Also See for EWJ 300-3:
Table of Contents

Advertisement

Available languages

Available languages

F4 SPD
Display speed
F5 HoUr
P6 bAtt
SALE m
The SALE mode is locked for balances with type approval.
To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chap. 6.11
Attention:
After destruction of the seal the balance must be re-verified by an authorised
agency and a new verification wire/seal mark fitted before it can be reused for
applications subject to verification.
EWJ-BA-e-1615
tP
Standard printer setting
LP 50 Not documented
Acknowledge selection by
Eng*
Standard stetting English.
only displayed for setting „LP 50"
chi
Not documented
Verifiable models:
SPd L
"Low"
SPd n
"Normal"
SPd H
"High"
oFF
Automatic adjustment switched off
1 HoUr
Time interval after which the automatic adjustment
starts, selectable 1, 2, 3,4, 5, 6, 7, 8 hours.
8 HoUr
bAttoF
Mask loading status display „Recharg.battery"
bAtton
Display loading status display „Recharg.battery"
Sale mode no: d= 0.1g
SALE n
EWJ 300-3:
d= 0.001 g
Sale mode yes: d= 0.01 g
SALE y
EWJ 300-3:
d= 0.01 g
.
Non-verifiable models:
SPd 1
SPd 2
SPd 3
SPd 4
EWJ 600-2M:
EWJ 6000-1M:
d= 0.01 g
d= 0.1 g
EWJ 600-2M:
EWJ 6000-1M:
d= 0.1 g
d= 1 g