Ideal Life Gluco-Manager GMM0001 Instruction Manual

Ideal Life Gluco-Manager GMM0001 Instruction Manual

Blood glucose management system
Hide thumbs Also See for Gluco-Manager GMM0001:

Advertisement

IDEAL LIFE Gluco-Manager
Blood Glucose Management System
Model
GMM00010001
INSTRUCTION MANUAL
Read before use
May be used with IDEAL LIFE Pod™
(sold separately) for use with IDEAL LIFE
Health Management System™
Products may vary from images shown
*NOTE: The IDEAL LIFE Gluco-Manager™ is intended for use in the quantitative measurement of glucose in fresh capillary
whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm,the calf and the thigh. It is
intended for use by adults with diabetes mellitus in a non-clinical environment such as in the home, as an aid to monitoring the
effectiveness of a diabetes control program. It is not intended for the diagnosis of, or screening for, diabetes mellitus and not in-
tended for use on neonates. The alternative site testing can be used only during steady state blood glucose conditions. The system
is intended for information purposes only, for expert advice make sure to consult a healthcare professional. Any instructions for
change in treatment and health related information or advice should only be made directly with a healthcare professional.
Version 7.0 Jun 2013
TM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Gluco-Manager GMM0001 and is the answer not in the manual?

Questions and answers

Summary of Contents for Ideal Life Gluco-Manager GMM0001

  • Page 1 Products may vary from images shown *NOTE: The IDEAL LIFE Gluco-Manager™ is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm,the calf and the thigh. It is intended for use by adults with diabetes mellitus in a non-clinical environment such as in the home, as an aid to monitoring the effectiveness of a diabetes control program.
  • Page 2: Table Of Contents

    Additional Items ........8 Looking at Messages......22 IDEAL LIFE Pod™Location/ Looking at Graphics.......23 Interference..........9 Comparing Meter and Laboratory Before You Start Your IDEAL LIFE Results............24 Gluco-Manager™........10 Care and Maintenance......25 Your IDEAL LIFE Pod™......11 Cleaning and Desinfecting....26 Test Strip Description ......12 A List of the Signs of Potential Important Test Strip Information..12...
  • Page 3: Introduction

    Testing is done outside the body (in vitro diagnostic use). The IDEAL LIFE Gluco-Manager™ is not intended for the diagnosis of, or screening for, diabetes mellitus and not intended for use on neonates.
  • Page 4: Precautions

    • Do not use the device come into contact with surfaces which are too hot to touch. • Do not use the device where aerosol sprays are being used. • Do not use the device if it is not working properly, or if it has been damaged. IDEAL LIFE Gluco-Manager™...
  • Page 5: Important Information

    Transmitting Bloodborne Pathogens: Initial Communication” (2010) http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm “CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” (2010) http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html For any questions or comments, please call an IDEAL LIFE customer representative at 1-866-IDEAL-41 (866-433-2541) Instruction Manual...
  • Page 6: About Alternative Site Testing (Ast)

    • You think your blood glucose is low. • You are unaware of hypoglycemia. • Your AST results do not match the way you feel. • You are testing for hyperglycemia. • Your routine glucose results are often fluctuating. • You are pregnant. IDEAL LIFE Gluco-Manager™...
  • Page 7: Package Contents

    These products have been designed, tested and proven to work together as a system to produce accurate blood glucose test results. Use only the same brand-name test strips and control solution with the IDEAL LIFE Gluco-Manager™. YOUR SYSTEM INCLUDES:...
  • Page 8: Additional Items

    SOLUTION AC ADAPTER ILP 0001 IDEAL LIFE POD™ NOTE: Check your system to be sure that it is unopened prior to use and that it contains all parts listed above. If either of these conditions occurred, please return your system to the place of purchase.
  • Page 9: Interference

    While using an IDEAL LIFE Pod™ with telephone line connectivity, the telephone line will be in use. If the telephone line is in use while trying to use your IDEAL LIFE Pod™, the information will be stored in your IDEAL LIFE Manager™ and will be sent later.
  • Page 10: Gluco-Manager

    Before You Start Before You Start Your IDEAL LIFE Gluco-Manager Test Port Blue LED LCD Display Right “Soft”Button Left “Soft”Button START/ESCAPE Button AAA Batteries Battery Cover Battery Cover IDEAL LIFE Gluco-Manager™...
  • Page 11: Your Ideal Life Pod

    Your IDEAL LIFE POD Your IDEAL LIFE Pod ™ (*An optional accessory sold separately) LED Display AC Adapter Jack USB Port Ethernet Cable RJ45 Jack Connection Telephone Cord RJ11 Jack For Telephone Connection (sealed for cellular internet connectivity) (Sealed with an antenna for cellular Internet connectivity)
  • Page 12: Test Strip Description

    • Do not use test strips beyond the expiration date printed on the package since they may cause inaccurate results. • Do not bend, cut or alter a test strip in any way. IDEAL LIFE Gluco-Manager™...
  • Page 13: Powering Up Your Ideal Life Gluco-Manager

    Powering Up Your IDEAL LIFE Gluco-Manager ™ INSTALLING THE BATTERIES (INCLUDED) • Remove the battery compartment door on your IDEAL LIFE Gluco-Manager™ by lifting the battery door using the notch. • Place the batteries as instructed on the device. • Replace the battery compartment door and click into place.
  • Page 14: First Time Power Up And Activation

    Make sure to call IDEAL LIFE before you begin. Then make sure that the IDEAL LIFE Pod™ is connected to a telephone line cord and RJ11 jack (or an Ethernet cord and RJ45 jack), “Powered ON”! Hold down any button on your IDEAL LIFE Gluco-Manager™...
  • Page 15: About Control Solution

    When should the control solution test be performed? • When you first get your IDEAL LIFE Gluco-Manager™. • When you begin using a new vial of test strips. • Whenever you suspect that the meter or test strips are not working properly.
  • Page 16: How To Do A Control Solution Test

    Shake the control solution vial well. Remove the cap from the control solution bottle. Place cap on flat surface. Squeeze the vial, discard the first drop, and wipe off the dispenser tip to prevent contamination. Squeeze the vial again to get another drop and apply the drop to the top of cap. IDEAL LIFE Gluco-Manager™...
  • Page 17 If you continue to get out of range results, it means that the system may not be work- ing properly. Do NOT test your blood. Contact an IDEAL LIFE customer representative at 1-866-IDEAL-41 (866-433-2541) to resolve the issue.
  • Page 18: Functions

    Functions Functions (when using the IDEAL LIFE POD™, an optional accessory sold separately) Main Screen The Main Screen is your default screen. This is where you will be directed to after completing functions and where you will find status indicators for the functioning of your IDEAL LIFE Gluco-Manager™.
  • Page 19: Taking Your Blood Glucose Reading

    Taking Your Blood Glucose Reading Taking Your Blood Glucose Reading Be sure to read this section and any inserts found in the test strip box carefully before taking a reading. Wash and dry your hands before starting. Getting a Drop of Blood Step 1.
  • Page 20: Taking A Reading

    Each time your IDEAL LIFE Gluco-Manager™communicates successfully to IDEAL LIFE through your IDEAL LIFE Pod™ the time and date are set to ensure reliable date and time stamping of all readings.
  • Page 21: Looking At Previously Taken Readings

    Looking At Previously Taken Readings Looking At Previously Taken Readings Your IDEAL LIFE Gluco-Manager™ will store up to 575 of your latest readings. Press Menu to view the feature list. SCROLL to highlight READINGS (it should already be highlighted as it is the first line item in the list).
  • Page 22: Looking At Messages

    Your IDEAL LIFE Gluco-Manager™ is equipped for 2-way, interactive communication when using a communication gateway or access point such as a cellular phone, computing device or the IDEAL LIFE Pod™. Messages can be sent remotely to your IDEAL LIFE Gluco-Manager™ and responded to directly.
  • Page 23: Looking At Graphics

    When using a communication gateway or access point such as a cellular phone, computing device or the IDEAL LIFE Pod™ your IDEAL LIFE Gluco-Manager™will store different types of graphics downloaded from the IDEAL LIFE website. (All the graphics can be altered through the IDEAL LIFE website, refer to www.ideallifeonline.com for details).
  • Page 24: Comparing Meter And Laboratory Results

    References 1. Surwit, R.S., and Feinglos,M.N.: Diabetes Forecast (1988), April, 49-51 2. Sacks, D.B.: “Carbohydrates.“Burtis, S.A., and Ashwood, E.R. (ed.),Tietz Textbook of Clinical Chemistry. Philadelphia: W.B. Saunders Company (1994), 959. IDEAL LIFE Gluco-Manager™...
  • Page 25: Care And Maintenance

    Care and Maintenace Care and Maintenance • Handle the meter with care. Dropping or throwing the meter may cause damage to the device. • Don’t expose the Blood Glucose Monitoring system, Test strip, and Control solution, to extreme conditions, such as high humidity, heat, freezing cold or dust. Storage conditions for the blood glucose Monitoring system (meter and test strips) are 39.2°F ~86°F Humidity <80% RH.
  • Page 26: Cleaning And Desinfecting

    Step 4 Let meter dry. Dispose of the wipe. With all the recommended glucose meter cleaning methods, it is crucial that the IDEAL LIFE Gluco- Manager™ is completely dry before taking a reading. Please follow the disinfectant’s product label instructions to ensure proper drying time.
  • Page 27: A List Of The Signs Of Potential Deterioration

    2. The users should wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips. 3. If you have any questions you can contact an IDEAL LIFE customer representative at 1-866-IDEAL-41 (866-433-2541) to resolve the issue.
  • Page 28: Specifications

    Auto Shut Off: After 2 minutes without action Measurement Units: Mg/dL Note: These specifications, to improve performance, are subject to change without notice. IDEAL LIFE Pod™ Model: ILP 0001 IDEAL LIFE Pod™ Display: LED’s Power Source: AC adapter OperatingTemperature/Humidity: 50°F to 104°F (10°Cto 40°C)/Below 85% relative humidity Storage Temperature/Humidity 23°F to 122°F (-5°C to 50°C)/Below 90% relative humidity...
  • Page 29: Limitations Of Use

    • The blood glucose monitoring system is not intended for use on artery blood, neonates’ serum and plasma. • IDEAL LIFE Gluco-Manager™ should be used with IDEAL LIFE test strip. • The system can be used up to an altitude of 10744 feet.
  • Page 30: Problem Solving Guide

    Improper use may cause an inaccurate result without producing an error message or a symbol. In the event of a problem, refer to the information under action. If you follow the recommended instructions but a problem still exists, please contact an IDEAL LIFE™ customer representative at 1-866-IDEAL-41 (866-433-2541) to resolve the issue.
  • Page 31: Error Message

    Problem Solving Guide Error Message MESSAGE WHAT IT MEANS ACTION Appears when the batteries Replace the batteries immediately. can not provide enough LOW BATTERY power for a test. Appears when inserting a Test with a newtest strip. used test strip. USED TEST STRIP Remove the strip after Re-test with a new test strip.
  • Page 32: Summary Of Operation

    Step 3. Obtain accurate results The meter will display an hour glass and then display the test result. The IDEAL LIFE Gluco-Manager™ will automatically return to sleep mode once the reading has been transmitted. IDEAL LIFE Gluco-Manager™...
  • Page 33: Fcc Statement

    Federal Communications Commission (FCC) Statement Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user’s authority to operate the equipment. 15.105(b) This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.
  • Page 34: Customer Information

    In most areas, the sum of the RENs of all devices on any one line should not exceed five (5.0). If too many devices are attached, they may not ring properly. IDEAL LIFE Gluco-Manager™...
  • Page 35: Service Requirements

    Service Requirements Service Requirements In the event of equipment malfunction, all repairs should be performed by our Company or an authorized agent. It is the responsibility of users requiring service to report the need for service to our Company or to one of our authorized agents. Please refer to the location you purchased or received your equipment for details of where service can be facilitated through.
  • Page 36 At all other times, you should contact your healthcare professional for assistance. © 2011 - V1 IDEAL LIFE INC. All rights reserved. IDEAL LIFE is a registered trademark in the United States for weight scales, blood glucose me- ters, blood pressure cuffs and meters, heart monitors. IDEAL LIFE SpO2 MANAGER, IDEAL LIFE STEP MANAGER, IDEAL LIFE BREATH MANAGER, IDEAL...