Main menu ..........................41 5.5.2 “Settings” submenu ......................... 42 5.5.3 “Service” submenu ........................42 Principle of controller entries ......................43 Performance during and after power failures ..................43 Performance when opening the door ....................44 Datacolor CONDITIONER™ (MB2) 06/2017 page 2/126...
Page 3
10.6 Value entry for a program section ..................... 70 10.6.1 Set-point ramp and set-point step modes ................70 10.6.2 Weekday ..........................71 10.6.3 Start time ..........................71 10.6.4 Setpoint entry .......................... 72 10.6.5 Special controller functions via operation lines ............... 72 Datacolor CONDITIONER™ (MB2) 06/2017 page 3/126...
Page 4
16.1 Views ............................... 101 16.1.1 Show and hide legend ......................101 16.1.2 Switch between legend pages ....................101 16.1.3 Show and hide specific indications ..................102 16.1.4 History display ........................102 16.2 Setting the parameters ........................105 Datacolor CONDITIONER™ (MB2) 06/2017 page 4/126...
Page 5
All efforts have been made to ensure the accuracy of the information presented in this format. However, should any errors be detected, Datacolor appreciates your efforts to notify us of these oversights. Changes are periodically made to this information and are incorporated into forthcoming versions. Data- color reserves the right to make improvements and/or changes in the product(s) and/or program(s) de- scribed in this material at any time.
Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall Datacolor be held liable for any dam- ages, direct or incidental arising out of or related to the use of this manual.
The following labels are located on the cabinet: Pictograms (warning signs) Hot surface (inner glass door above the glass door handle) Burning and scalding hazard (cabinet rear) Figure 1: Position of labels on the cabinet rear Datacolor CONDITIONER™ (MB2) 06/2017 page 8/126...
Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact your local Datacolor service representative for these replacements. Type plate The type plate sticks to the left side of the cabinet, bottom right-hand. Nominal temp.
Datacolor is only responsible for the safety features of the cabinet provided skilled electricians or qualified personnel authorized by Datacolor perform all maintenance and repair, and if components relating to cabinet safety are replaced in the event of failure with original spare parts.
Page 11
Damage to the cabinet and the charging material. Housing cover breakaway. ∅ Do NOT climb on the lower housing cover. ∅ Do NOT load the lower housing cover and the door with heavy objects while the cabi- net door is open. Datacolor CONDITIONER™ (MB2) 06/2017 page 11/126...
Intended use Datacolor CONDITIONER™ sample conditioning cabinets are suitable for exact conditioning of harmless materials. A mixture of any component of the charging material with air must NOT be explosive. The op- erating temperature must lie below the flash point or below the sublimation point of the charging material.
• Examinations The chamber has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. (DGUV) (German Social Accident Insurance (DGUV)” (German Social Accident Insurance (DGUV), Testing and Certification Body Foodstuffs and Packaging Industry in DGUV Test). Datacolor CONDITIONER™ (MB2) 06/2017 page 13/126...
• Vapors of oil and fat are dangerous for the sensor because they may condensate at the sensor and thus prevent its function (insulating layer). For similar reasons it is not possible to measure smoke gases. Datacolor CONDITIONER™ (MB2) 06/2017 page 14/126...
Cabinet description The Datacolor CONDITIONER™ sample conditioning cabinet is equipped with a multifunctional micro- processor display controller with 2-channel technology for temperature and humidity plus a digital display accurate to one-tenth of a degree resp. 0.1% r.H. With its comprehensive program control functions, the display program controller MB2 permits the high precision performance of temperature and humidity cy- cles.
For the control ranges of temperature and humidity, see climatic diagram (chap. 17). Standard temperature and humidity settings for sample conditioning are typically 21 °C and 65 % r.H. Cabinet overview Figure 3: Datacolor CONDITIONER™ sample conditioning cabinet Instrument box Door handle...
Figure 4: Instrument panel with MB2 program controller and USB interface Lateral control panel Figure 5: Lateral control panels at the right side of the refrigerating / humidity generation module Main power switch Ethernet interface Datacolor CONDITIONER™ (MB2) 06/2017 page 17/126...
Power cable Purging outlet for humidifying module – for service purpose only Freshwater connection “IN” with screw thread ¾’’ for hose ½“, with union nut Wastewater connection “OUT” with hose olive for hose ½“ Datacolor CONDITIONER™ (MB2) 06/2017 page 18/126...
∅ Do NOT lift or transport the cabinet using the door, the handle, or the lower housing. Lift the cabinet at the four lower corners with the aid of 4 people You can order transport packing for moving or shipping purposes from your local Datacolor service rep- resentative.
• If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F. • After your local Datacolor service representative has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several times a year. Avoid any conductive dust in the ambiance according to the cabinet layout complying with pollution de- gree 2 (IEC 61010-1). Datacolor CONDITIONER™ (MB2) 06/2017 page 21/126...
Please fix both spacers with the supplied screws at the cabinet rear. This serves to ensure the prescribed minimum distance to the rear wall of 100 mm / 3.94 in. Figure 7: Spacer for wall distance Figure 8: Cabinet rear with mounted spacers Datacolor CONDITIONER™ (MB2) 06/2017 page 22/126...
Any problems and malfunctions that might arise following use of water of deviating quality is excluded from liability by the manufacturer. The warranty becomes void in the event of use of water of deviating quality. Datacolor CONDITIONER™ (MB2) 06/2017 page 23/126...
6m water hose (PVC black, 12 mm (½’’) inner diame- ter), which should be cut into 2 accordingly, for the feed hose and drain. Pressure-resistant up to 10bar, heat resistant up to 95 °C / 203 °F. Datacolor CONDITIONER™ (MB2) 06/2017 page 24/126...
Page 25
Continue by testing the function and tightness of the mounted cabinet Check: Quickly open the water tap while there is no cabinet connected – the valve should cut off the wa- ter flux without any delay. Datacolor CONDITIONER™ (MB2) 06/2017 page 25/126...
The cans are placed in holding devices. You can affix them directly at the rear of the cabinet or place them next to the cabinet. Figure 10: Cabinet rear view with installed external water cans (option) Datacolor CONDITIONER™ (MB2) 06/2017 page 26/126...
Cable connections Connect the plug of the cable to the socket (3) at the rear of the cabinet. The socket (3) is marked with a sticker: Socket (3) Figure 12: Connections at the cabinet rear Datacolor CONDITIONER™ (MB2) 06/2017 page 27/126...
Empty the wastewater can in a timely manner before it is full. Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the wastewater can. Datacolor CONDITIONER™ (MB2) 06/2017 page 28/126...
Hose connection of the freshwater can (11) Figure 14: Freshwater can (option) Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the freshwater can. Datacolor CONDITIONER™ (MB2) 06/2017 page 29/126...
Damage to the equipment. Check the power supply voltage before connection and start-up. Compare the power supply voltage with the data indicated on the type plate. See also electrical data (chap. 23.4). Datacolor CONDITIONER™ (MB2) 06/2017 page 30/126...
0.4 m / 1.3 ft. CAUTION Danger of overheating. Damage to the voltage changer. ∅ Do NOT install the voltage changer in unventilated recesses. Ensure sufficient ventilation for dispersal of the heat. Datacolor CONDITIONER™ (MB2) 06/2017 page 31/126...
Page 32
Lateral wall clearance of the cabinet to set up the voltage changer (minimum) Electrical connection data of the voltage changer Input side 26,9 Output side (to the cabinet) 13,0 Power frequency 50 / 60 Datacolor CONDITIONER™ (MB2) 06/2017 page 32/126...
(cabinet with lights), fan speed and special controller functions for each program section. You can enter values or programs directly at the controller or use the APT- COM™ 3 DataControlSystem software (option) specially developed by BINDER. Datacolor CONDITIONER™ (MB2) 06/2017 page 33/126...
Page 34
Figure 17: Normal display of the MB2 program controller (sample values), cabinet without lights Operating mode Temperature values Humidity values Light intensity values Actual values Set-point values Functional icons Figure 18: Normal display of the MB2 program controller (sample values), cabinet with lights Datacolor CONDITIONER™ (MB2) 06/2017 page 34/126...
Back to Normal display Information Program start Setpoint entry Event list Display of active alarms Access to main menu Figure 19: Operating functions of the MB2 controller in normal display (example values for cabinet with lights) Datacolor CONDITIONER™ (MB2) 06/2017 page 35/126...
Press the Normal display icon to return from program display and chart recorder display back to Normal display. • Normal display (actual values / setpoint values) (Cabinet without lights) (Cabinet with lights) • Program display (example: time program) • Chart recorder display (Cabinet without lights) (Cabinet with lights) Datacolor CONDITIONER™ (MB2) 06/2017 page 36/126...
Reset alarm Acknowledge the alarm and mute the buzzer. Change between uppercase and lower case characters, digits Change keyboard and special characters Edit Edit settings of time and week programs Datacolor CONDITIONER™ (MB2) 06/2017 page 37/126...
Page 38
“Door open” Cabinet door is open “Humidity off” The humidification / dehumidification system is turned off Information icon in the touchscreen calibration menu Symbol Information Remaining time to touch the display when calibrating the touchscreen. Datacolor CONDITIONER™ (MB2) 06/2017 page 38/126...
An entered week program for temperature, humidity fan speed and light intensity (cabinet with lights is running. The controller offers 5 program memory places with 100 switching points each. The switching points can be distributed over all days of the week. Datacolor CONDITIONER™ (MB2) 06/2017 page 39/126...
/ cancel an already entered, respectively a running 9.2, 10.1 week program Unless noted otherwise, the figures show the functional range, which is available for the user with “Ad- min” authorization level. Datacolor CONDITIONER™ (MB2) 06/2017 page 40/126...
• Access to service data, controller reset to factory settings (chap. 5.5.3) • Available only for users with “Service” and “Admin” authorization level. Full functional range only for Datacolor service representatives (users with “Service” authorization level). “Programs” submenu • Access to the controller’s program functions (chap. 8, 9, 10) Datacolor CONDITIONER™...
5.5.3 “Service” submenu The “Service” submenu is available for users with “Service” or “Admin” authorization level. When logged- in with “Admin” authorization level the user will find information to tell your local Datacolor service repre- sentative in service case. Path: Main menu >...
The week program continues with the values corresponding to the current time. Power failure and power return are noted in the event list (chap. 15.3). If during power failure an alarm has occurred (tolerance range, safety controller), confirm the alarm. See chap. 11.3. Datacolor CONDITIONER™ (MB2) 06/2017 page 43/126...
The controller will function in the operating mode, which was active before the last shut-down. It controls temperature and humidity in fixed value operating mode to the last entered values and in the program mode to the set points achieved beforehand. Datacolor CONDITIONER™ (MB2) 06/2017 page 44/126...
“Setpoints” menu. You can turn humidity control (humidification and dehumidification) on or off. Select “Control on/off”. If the “Humidity “checkbox is marked, humidity control is active. Mark / unmark the checkbox to change the setting. Datacolor CONDITIONER™ (MB2) 06/2017 page 45/126...
When operating without humidity by setting “Control off” (chap. 6.3), the humidity tolerance range function is automatically deactivated. When operating without humidity by activated operation line “Humidity off” (chap. 7.3), set the humidity tolerance range to “0” in order to avoid tolerance range alarms (chap. 11.4). Datacolor CONDITIONER™ (MB2) 06/2017 page 46/126...
Confirm icon and repeat the entry with a correct value. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Datacolor CONDITIONER™ (MB2) 06/2017 page 47/126...
When the configured moment for program start is reached, the program delay time ends and the program starts running. The controller equilibrates to the values which had been active during program start Datacolor CONDITIONER™ (MB2) 06/2017 page 49/126...
Press the Confirm icon. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. Datacolor CONDITIONER™ (MB2) 06/2017 page 50/126...
The grey bar shows how much time of the whole time is elapsed. If program duration has been set to infinite, the grey bar is not displayed. Datacolor CONDITIONER™ (MB2) 06/2017 page 51/126...
“Idle mode” in the last program section. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. Datacolor CONDITIONER™ (MB2) 06/2017 page 52/126...
3) or create a new program (chap. 9.4). There are the following options: The program view opens. Select a program section to open the section editor (chap. 9.6) Press the Edit icon to open the program editor Datacolor CONDITIONER™ (MB2) 06/2017 page 53/126...
In the program view press the Edit icon to open the program editor In the program editor select “Delete program” and press the Confirm icon. The program is deleted. The controller returns to the program view. Datacolor CONDITIONER™ (MB2) 06/2017 page 54/126...
• Replace section: Replacing an existing section with the copied section. This menu point is visible only after a section has been copied. • Insert section: Adding the copied section. This menu point is visible only after a section has been cop- ied. • Delete section • Add new section Datacolor CONDITIONER™ (MB2) 06/2017 page 55/126...
The current section (example: section 1) is the copied section shall be inserted… copied. The controller returns to the section view. Datacolor CONDITIONER™ (MB2) 06/2017 page 56/126...
In the section view press the Edit icon to open the section editor In the section editor select “Delete section” and press the Confirm icon. The section is deleted. The controller returns to the section view. Datacolor CONDITIONER™ (MB2) 06/2017 page 57/126...
Select the field “Duration” indicating the time. “Duration” entry menu. Enter the desired section duration with the arrow keys and press the Confirm icon. Setting range: 0 up to 99 hours 59 min 59 sec. Datacolor CONDITIONER™ (MB2) 06/2017 page 58/126...
The entered setpoint of the section is reached as fast as possible and then held constant for the re- maining section duration. Selecting the setting “Ramp” or “Step” Section view (partial view). In the field “Course” select the desired setting “Ramp” or “Step”. Datacolor CONDITIONER™ (MB2) 06/2017 page 59/126...
“Functions on/off” entry menu. Select the field “Functions on/off”. Mark / unmark the checkbox of the desired function to activate / deactivate it and press the Confirm icon. The controller returns to the section view. Datacolor CONDITIONER™ (MB2) 06/2017 page 60/126...
This is indicated on the display (see below). When the actual temperature is situated again within the entered tolerance limits, the program automatically continues. Therefore, the duration of the program may be extended due to the programming of tolerances. Datacolor CONDITIONER™ (MB2) 06/2017 page 61/126...
“Start section for repetition” To have sections repeated infinite- ly, enter the number of repetitions as “-1”. The selected sections are repeated as many times as selected. Then the program continues. Datacolor CONDITIONER™ (MB2) 06/2017 page 62/126...
The controller changes to the Normal display. The controller changes to the program view. To save the programming it is absolutely required to press the Confirm icon. Otherwise all settings will be lost! There is no confirmation prompt! Datacolor CONDITIONER™ (MB2) 06/2017 page 63/126...
A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be cancelled. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. Datacolor CONDITIONER™ (MB2) 06/2017 page 64/126...
For the first section no weekday is specified. Therefore the section is first marked in red and cannot be saved. Note for cabinet with lights: The light setpoint is not visible in the week program view. Datacolor CONDITIONER™ (MB2) 06/2017 page 65/126...
• Copy program • Replace program: Replacing an new or an existing program with the copied program. This menu point is visible only after a section has been copied. • Delete program • Add new section Datacolor CONDITIONER™ (MB2) 06/2017 page 66/126...
In the program view press the Edit icon to open the program editor In the program editor select “Delete program” and press the Confirm icon. The program is deleted. The controller returns to the program view. Datacolor CONDITIONER™ (MB2) 06/2017 page 67/126...
• Replace section: Replacing an existing section with the copied section. This menu point is visible only after a section has been copied. • Insert section: Adding the copied section. This menu point is visible only after a section has been cop- ied. • Delete section • Add new section Datacolor CONDITIONER™ (MB2) 06/2017 page 68/126...
(example: section 2). The current section (example: section 1) is Press the Confirm icon copied. The controller returns to the program view. The controller returns to the section editor Datacolor CONDITIONER™ (MB2) 06/2017 page 69/126...
You can define the type of temperature and humidity transitions for the entire week program. Select the desired program and press the Edit icon to open the program editor. In the program editor select the “Change program name” function and press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 70/126...
Section view (example: cabinet with lights). 10.6.3 Start time Entry menu “Moment”. Section view (example: cabinet with lights). Select the field “Moment”. Select with the arrow keys the desired start moment of the section and press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 71/126...
• Operation line Idle mode” activates / deactivates the operating mode “Idle mode”. The other operation lines are without function. Select the desired program and section. You can set the operation lines in the “Functions on/off” field. For details please refer to chap. 9.7.3. Datacolor CONDITIONER™ (MB2) 06/2017 page 72/126...
(via operation line and/or by setting “Control on/off”) “Humidity off” immediately or Temperature setpoint below 0 °C or above 95 °C Cabinet door open “Door open” immediately Notifications are not shown in the event list. Datacolor CONDITIONER™ (MB2) 06/2017 page 73/126...
Maintenance of the humidity system is required. Contact “Humidity module after predefined time your local Datacolor service representative. service” (approx. 1 year) Messages concerning the humidity system are shown in the event list. Datacolor CONDITIONER™ (MB2) 06/2017...
• Acknowledging after the alarm condition has ended: The buzzer and the visual alarm indication are reset together. The alarm is then no longer in the list of active alarms. • The zero-voltage relay alarm output resets together with the alarm. Datacolor CONDITIONER™ (MB2) 06/2017 page 75/126...
If the condition persists, an alarm is triggered after the configured interval (“range alarm delay”). It is visu- ally indicated in Normal display. If the alarm buzzer is activated (chap. 11.5) there is an audible alert. The alarm is shown in the list of active alarms (chap. 11.3). Datacolor CONDITIONER™ (MB2) 06/2017 page 76/126...
If this limit temperature is reached, the over temperature protective device permanently turns off the cabi- net. The user cannot restart it. The protective cut-off device is located internally. Only a service specialist can replace it. Therefore, please contact an authorized service provider or your local Datacolor service representative.
• In the field “Mode” select the desired setting “Limit” or “Offset”. • Select the corresponding field “Limit” or “Offset” according to the selected mode and enter the desired safety controller setpoint. Confirm entry with Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 78/126...
12.2.4 Function check Check the safety controller at appropriate intervals for its functionality. It is recommended that the author- ized operating personnel should perform such a check, e.g., before starting a longer work procedure. Datacolor CONDITIONER™ (MB2) 06/2017 page 79/126...
• All passwords can be changed in the “log out” submenu (chap. 13.3). “Service” authorization level • Authorization level only for Datacolor service • Extensive authorization for controller operation and configuration, access to service data • The passwords for “Service”, “Admin” and “User” authorization levels can be changed in the “log out”...
Page 81
No lock icon is shown in the display header. User login is neither required nor possible. To activate the password protection and user login, perform new password assignment (chap. 13.5.3). Example: cabinet without lights Datacolor CONDITIONER™ (MB2) 06/2017 page 81/126...
Page 82
Display when at least some of the authorization levels are password protected and a user has logged in. The user’s authorization (by password entry) is shown. Example: user with “Admin” authorization. Datacolor CONDITIONER™ (MB2) 06/2017 page 82/126...
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Controller with deactivated password Datacolor CONDITIONER™ (MB2) 06/2017 page 83/126...
13.4 User change If the password function has been deactivated (chap.13.5.2) this function is not available. Path: Main menu > User > User change Controller with logged-in user Datacolor CONDITIONER™ (MB2) 06/2017 page 84/126...
A logged-in user can change the passwords of his current level and of the next lower level(s). Example: A user with “Admin” authorization can change the passwords for the “Admin” and “User” au- thorization levels. Path: Main menu > User > Password User logged-in with “Service” or “Admin” authorization Datacolor CONDITIONER™ (MB2) 06/2017 page 85/126...
Page 86
Example: access the digit entry window To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation (sample picture). For each character of the password, the re- quired keyboard appears automatically. Then press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 86/126...
Select the authorization level for which the Do NOT enter anything in the “Password” password shall be deleted. screen. Press the Confirm icon. The password is deleted. Do NOT enter anything in the “Confirm password” screen. Press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 87/126...
(Example: “Admin” authorization) To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation. For each character of the password, the required keyboard appears automatically. Then press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 88/126...
The available functions are indicated by marked checkboxes. Example: Extended configurations available. Under “Expiration date” the date of expiry of the code is displayed. “Activation code” menu with entered code (sample view). Press OK to take over the entry Datacolor CONDITIONER™ (MB2) 06/2017 page 89/126...
Confirm icon. 14.2 Selecting the temperature unit Following start-up of the cabinet: Or later: Path: Main menu > Settings > Chamber Select the desired temperature unit and press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 90/126...
Or later: Path: Main menu > Settings > Date and time “Date and time” submenu. “Date / time” entry menu. Select the field “Date / time”. Enter date and time and press the Confirm icon. Datacolor CONDITIONER™ (MB2) 06/2017 page 91/126...
Close icon to exit the menu without taking over the entries. 14.4 Display configuration 14.4.1 Adapting the display parameters This function serves to configure parameters like display brightness and operating times. Path: Main menu > Settings > Display > Display “Display” submenu. Datacolor CONDITIONER™ (MB2) 06/2017 page 92/126...
The clock icon shows how much time there is left to touch the currently activated box. If the box is not touched withing this period, calibration is aborted and the display changes to Normal display. After completing the calibration, i.e., touching all four boxes, the display changes to Normal display. Datacolor CONDITIONER™ (MB2) 06/2017 page 93/126...
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Datacolor CONDITIONER™ (MB2) 06/2017 page 94/126...
Setting range: 0 min to 65535 min. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Datacolor CONDITIONER™ (MB2) 06/2017 page 95/126...
• Select the field “Email sender” and enter the desired Email sender. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Datacolor CONDITIONER™ (MB2) 06/2017 page 96/126...
Depending on the user’s authorization level, different functions (highlighted in black) are available for the logged-in user. Available functions with “User” authorization level Available functions with “Admin” authorization level Available functions with “Service” authorization level Datacolor CONDITIONER™ (MB2) 06/2017 page 97/126...
15.2 Current operating parameters Press the Information icon to access the “Info” menu from Normal display. “Info” menu. Select the desired information. • Select “Program operation” to see infor- mation on a currently running program. Datacolor CONDITIONER™ (MB2) 06/2017 page 98/126...
Press the Update icon to update the event list. Attention: Following a modification of the language setting (chap. 14.1) or the storage inter- val of the chart recorder (chap. 16.2) the Event list is cleared. Datacolor CONDITIONER™ (MB2) 06/2017 page 99/126...
There are two counters: one for the cabinet operating hours and the other for the operating hours of the illumination. Both counters can be reset independently. The rest menu is available only for users with “Service” au- thorization. To have the values reset, please contact your local Datacolor service representative. 15.5 Technical information on the cabinet Path: Main menu >...
Display with cabinet with lights Display with cabinet without lights 16.1.2 Switch between legend pages Switch legend Press the Switch legend icon to switch between the legend pages page, cabinet without lights page page, cabinet without lights Datacolor CONDITIONER™ (MB2) 06/2017 page 101/126...
Move the central red line by tapping and holding to the desired position. The legend at the right side shows the values of the current line position. History display with cabinet with lights History display with cabinet without lights Then further icons appear: Datacolor CONDITIONER™ (MB2) 06/2017 page 102/126...
Page 103
“Curve selection” submenu (cabinet without lights) History display: Search the required instant Search Press the Search icon to access the “Search” submenu. Select the required instant by entering its date and time and press the Confirm icon “Search” submenu. Datacolor CONDITIONER™ (MB2) 06/2017 page 103/126...
Page 104
Press the Show scroll buttons icon to access the “Page selection” submenu. Scroll buttons are shown on the left and on the right. Use them to move along the timeline. “Page selection” submenu (cabinet without lights) “Page selection” submenu (cabinet with lights) Datacolor CONDITIONER™ (MB2) 06/2017 page 104/126...
Change the storage rate or scaling ONLY if the previously registered data is no longer needed. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Datacolor CONDITIONER™ (MB2) 06/2017 page 105/126...
Discontinuous range (no continuous operation, up to 24 h) Range C: Optimized operating range from 15 °C to 31 °C and 38 % r.h. to 70% r.h. with stable opera- tion of the humidity system. Datacolor CONDITIONER™ (MB2) 06/2017 page 106/126...
Page 107
• Set the humidity to 0 % r.H. The humidity system must be activated. Operate the cabinet for at least 2 hours. • Only then, shut down the cabinet at the main power switch (1) and close the water supply tap. Datacolor CONDITIONER™ (MB2) 06/2017 page 107/126...
If the humidity system is turned off while there are descending temperature curves, then operation of the refrigeration system may cause dehumidification of the charging material. For error indications concerning water supply and humidity system, see chap. 11.1.3 and 22. Datacolor CONDITIONER™ (MB2) 06/2017 page 108/126...
Manually defrost the cabinet (see description above). Then, shut down the cabinet at the main power switch (1) and close the tap of the wa- ter supply. Keep removed the access port plugs. Datacolor CONDITIONER™ (MB2) 06/2017 page 109/126...
When operating the chamber with illumination: Restricted temperature and humidity range 20 °C / 68 °F to 60 °C / 140 °F. At ≤ 20 °C do not set the humidity set-point below 30 % r.H. Otherwise, no stable humidity control is guaranteed. Datacolor CONDITIONER™ (MB2) 06/2017 page 110/126...
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several times a year. Support Questions? If you need help with a Datacolor product, please contact one of our top-rated technical support teams located around the world for your convenience. You can find contact information below for the Datacolor office in your area.
∅ Do NOT spill water or cleaning agents over the inner and outer surfaces. Before cleaning, turn off the cabinet at the main power switch and dis- connect the power plug. Completely dry the appliance before turning it on again. Datacolor CONDITIONER™ (MB2) 06/2017 page 112/126...
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the charging material. If there is doubt regarding the suitability of cleaning products, please contact your local Datacolor service representative. We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough cleaning.
Page 114
Skin and eye damage due to chemical burns. ∅ Do not ingest. Keep away from food and beverages. ∅ Do NOT empty into drains. Wear protective gloves and goggles. Avoid skin contact. Datacolor CONDITIONER™ (MB2) 06/2017 page 114/126...
Do not use decontamination agents that may cause a hazard due to reaction with components of the device or the charging material. If there is doubt regarding the suitability of cleaning products, please contact your local Datacolor service representative. You can use the following disinfectants: Inner chamber Standard commercial surface disinfectants free from acid or halides.
According to Annex I of Directive 2012/19/EU of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), Datacolor CONDITIONER™ cabinets are classified as “monitoring and control instruments” (category 9) only intended for professional use“. They must not be disposed of at public collecting points.
In Europe, recovery of the refrigerant R 134A (1300) is mandatory accord- ing to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements regard- ing qualification of staff, disposal, and documentation. Datacolor CONDITIONER™ (MB2) 06/2017 page 117/126...
Cabinet fuse has responded. Cabinet without function. again, contact your local Data- color service representative. Controller defective. Nominal temperature exceeded Contact your local Datacolor ser- by 10° due to cabinet failure. vice representative. Over temperature protective de- vice (class 1) responds. Heating Controller defective.
Page 119
Menu functions not available with Log in with the required higher Menu functions not available. current authorization level. authorization. Contact your local Datacolor ser- No access to controller Password incorrect. vice representative. Chart recorder function: meas- New setting of storage rate or Change the storage rate or scal- ured-value memory cleared;...
Only qualified service personnel authorized by Datacolor representative must perform repair. Repaired cabinets must comply with the manufacturer’s quality standards. Technical description 23.1 Factory calibration and adjustment This cabinet was calibrated and adjusted in the factory. Calibration and adjustment were performed using standardized test instructions, according to the QM DIN EN ISO 9001 system applied by the manufactur- er (certified since December 1996 by TÜV CERT).
60 Hz power frequency 200-240 200-240 Current type Power plug NEMA 6-20P 6-20P Nominal power 2.00 2.00 Nominal current Installation category acc. to IEC 61010-1 Pollution degree acc. to IEC 61010-1 Over-current release category B, 2 poles Datacolor CONDITIONER™ (MB2) 06/2017 page 121/126...
23.5 Equipment To operate the cabinet, use only original Datacolor accessories or accessories / components from third-party suppliers authorized by Datacolor. The user is responsible for any risk arising from using unauthorized accessories. Regular equipment...
Plug for silicon access port d30 6016-0035 Set of two “Bio Vital” fluorescent tubes, set for 1 light cassette 8012-1812 For information on components not listed here, please contact your local Datacolor service representa- tive. Datacolor CONDITIONER™ (MB2) 06/2017 page 123/126...