Summary of Contents for Measurement Systems MSI-6260CS
Page 1
MSI6260cs RF Digital Crane Scale Calibration Guide...
Page 2
3750CS, MSI-9850, or the MSI-9750A Handheld. Calibration procedures using these devices are covered in their respective User Guides. If above listed remote devices are not in use with the MSI-6260CS, the only way to calibrate it is through the Comm. Port using the Terminal Access Mode. MSI recommends having test weights of at least 50% capacity to properly calibrate the Crane Scale.
Page 3
CALIBRATION PROCEDURE Select Channel Calibration press 1. ***** Channel Setup ***** Channel Calibration Data backup Filter (software) Channel Scanlist functions Motion Compensation (Accelerometer option required) Motion Detection Tare Total Units Lock Zero Standard Mode: Industry 4 to 20mA Temperature Sensor Installed: No (ESC) Exit Start the Calibration procedure by pressing 1.
Page 4
This list is for a type ‘A’ scale with up to 10000 counts resolution. This is the standard default. MSI recommends all MSI-6260CS Series crane scales calibrated at 5000 divisions or less. Too many divisions will compromise stability and settling time. Use the largest division size suitable for your application.
Page 5
Next, the MSI-9000 CellScale prompts for the value of the first calibration test weight. This step defaults to capacity. If this is the actual calibration test load push Y, otherwise push N to input the actual cal weight. If Y is selected, skip to step 10. ***** Channel Calibration ***** Calibrating Ch # 1-1, named: …....
Page 6
Calibration is now complete. After 15 seconds, press R, to assure Main Menu appears in full as confirmation of data stored. Disconnect serial cable from scale and cycle power ---------- MSI 9000 - CellScale ------------------------ -(c) 2005 Measurement Systems International - ---------------------------------------------------------------- 13:48:06...
Page 7
Notes on terminal programs There are several terminal programs available for every platform. Below are some notes on terminal programs that are used regularly at MSI for communicating with CellScale products. TeraTermPro Get from downloads.com or MSI can e-mail “ttermp23.zip” After installing, start the program Start ->...
Need help?
Do you have a question about the MSI-6260CS and is the answer not in the manual?