  • Page 2 Explanation of symbols: Do not re-use Use-by date Batch code In vitro diagnostic use in animals. Temperature limit Humidity limitation Caution (refer to accompanying documents). Please refer to safety-related notes in the manual accompanying this instrument. Consult instructions for use. Keep away from sunlight.
  • Page 3: Table Of Contents

    1. Contents Page Section Topic 1. Contents 2. Introduction 3. Appearance and Key Functions of the meter 4. Safety Information 5. Quick Testing Instructions Preparing for Testing 6. Installing the Battery 7. Settings (Date/Time/Alarm/Measuring Unit/ 8-11 Memory Deletion ) 8. Coding the Meter 9.
  • Page 4: Introduction

    EverPaw(model: 5228P) Blood Glucose Monitoring System. Please read this User Guide and the package insert that comes with the EverPaw(model: 5228P) Glucose Test Strips before testing. The system is available either as a meter alone or as a kit. If you have a meter only, you can purchase the system supplies from your system provider.
  • Page 5 The starter kit of EverPaw(model: 5228P) Blood Glucose Monitoring System includes the following items: 1.EverPaw(model: 5228P) Glucose Meter with 2 AAA batteries 2.User Guide 3.Carrying Case 4.EverPaw(model: 5228P) Glucose Test Strips 5.EverPawControl Solution 6.Sterile Lancets 7.Lancing Device 8.Package Inserts for Test Strips and Lancing Device "A single level control solution is provided in the "Starter Kit"...
  • Page 6: Appearance And Key Functions Of The Meter

    3. Appearance and Key functions of the meter 1.Test strip slot - When the strip is inserted into the slot, the meter will automatically turn on. 2. LCD Display - Guide you through the test using symbols and simple messages. 3.
  • Page 7: Safety Information

    4. Safety Information Please use this device only for the intended use described in this user guide. Before using your EverPaw(model: 5228P) Blood Glucose Monitor please ensure you read the enclosed User Guide thoroughly before carrying out a blood glucose test. Then carry out a practice test and calibration test to familiarise yourself with how the system works.
  • Page 8: Quick Testing Instructions

    5. Quick Testing Instructions (For detailed instructions start on page 13 for Control Solutions and page for Blood Testing.) 1.Check the code on the test strip vial. If necessary, change the code on the meter by pressing C or M after you enter a test strip.
  • Page 9: Installing The Battery

    6. Installing the battery 1.Turn off the meter by pressing “M” button and hold for 2 seconds before replacing the batteries. Slide down the battery cover on the back of the meter by pushing the cover in the direction of the arrow and put the cover aside.
  • Page 10: Settings

    7. Settings (Date / Time / Alarm / Memory Deletion) The LCD screen will show as below: (1)Measurement Units (2)Memory (3)Setting (4)Blood drop icon (5)Battery Symbol (12)Dog Blood Testing (6)Day Average (13)Cat Blood Testing (7)Alarm Reminder (14)Control Solution (8)Month (11)Minute (9)Day (10)Hour 1.
  • Page 11 Set the Year - With the meter off, by pressing the M key and hold for 2 seconds, you will enter the setting modes, press C key to adjust the year until the desired year is displayed    and then press the M key to confirm the year setting.
  • Page 12 Set the Hour - Press C key to adjust the hour until the desired hour appears on the screen and then press the M key to confirm the setting. You will then see the minute segment is flashing on the ...
  • Page 13: Memory Deletion )

    Delete Memory - Press C key to toggle between a flashing “dEL” for memory deletion or a non-flashing “dEL”. If you want to delete all memory, press M key while the “dEL” is flashing. If you do not want to delete all memory, press M key when the “dEL”...
  • Page 14: Coding The Meter

    8. Coding the meter (This step is only for coding version.) IMPORTANT!!! Always confirm/set the Code when you open a new vial of test strips, verify the code number on the meter matches the code number printed on the strip vial. 1.
  • Page 15: Running A Control Solution Test

    9.Running a Control Solution Test IMPORTANT: Always make sure you press C key for Control Solution test, otherwise the control solution test result will be stored in the memory, and affect your AVG results. 1. Insert a test strip into the meter while the meter is off.
  • Page 16 Control Solution mode, the test result will not be stored in the memory.) 3. Use only EverPawControl Solution with EverPaw(model: 5228P) Glucose Meter. 4. EverPawControl Solutions are used to check that the meter and the test strips are working together as a system and that you are performing the test correctly.
  • Page 17: Running A Blood Glucose Test

    10.Running a Blood Glucose Test 1. Twist and pull off the cap of the lancing device. 2. Insert a new lancet firmly into the lancet holder. 3. Twist off the protective cover of the lancet. 4. Replace and twist the adjustable depth cap onto the body of the lancing device until you hear the click.
  • Page 18 6.Open the test strip vial. Take one new test strip out of the vial and recap the vial quickly and firmly. 7.Check the code on the test strip vial. If necessary, change the code on the meter by pressing C or M after you enter a test strip.
  • Page 19 9.Apply a small blood drop to the front edge of the test strip, and the blood should be pulled into the confirmation window before the meter begins to count down. Do not push your sample site against the test strip or try to apply a smeared blood sample.
  • Page 20: Recalling The Memory And Viewing The Average

    11. Recalling the Memor y and Viewing the Average A. Using the Meter Memory The test results for blood sample will be stored in the memory automatically. The meter can store up to 500 of the most recent test results. You can also view the average of test results for various periods of 7, 14, 28, 60 and 90 days.
  • Page 21 C. Viewing the Average (AVG) of Test Data in Memory 1. To display the average (AVG) test result, please turn meter off. Press C and M keys at the same time and hold for about 2 seconds, wait for AVG displayed on the screen, then release the two buttons.
  • Page 22: Expected Values

    EverPawmeter gives plasma equivalent results. 13.Transmitting Results Note: EverPaw (model: 5228P) Glucose Meter allows you to transfer the test results stored in its memory to your personal computer. However, you will need to order software and a Data Download Cable separately from your distributor.
  • Page 23: Maintenance Of Your System

    14.Maintenance of your system Please, treat this meter with proper care, and keep it in good condition. 1. Store your meter in its case, in a clean dry place at 35.6~86℉(2~30℃). 2. Always clean your meter after use. Wipe and clean surface of the meter with a soft cloth that has been slightly dampened with alcohol.
  • Page 24: Troubleshooting

    15.Troubleshooting : Following is a summary of all Error Messages. These messages help to identify certain problems,but do not appear in all cases when a problem has occurred. Improper use may cause an inaccurate result without producing an error message or a symbol.In the event of a problem,refer to the information in the table under Solution.
  • Page 25 Problem Possible Cause Solution The meter is operating Repeat the test after the in an ambient The meter displays meter and strip are placed in temperature below an E_1 error a warmer environment and 10 ℃ or 50℉ which is message.
  • Page 26: Customer Service

    16.Customer Ser vice If you need assistance with your EverPaw (model: 5228P) Glucose Monitoring System, please contact: Manufactured for and distributed by: EverPaw Brooklyn, NY 11219 Questions or Comments? Get in Touch: EverPaw.com For questions related to your pet's health condition, please call your veterinarian or healthcare professional.
  • Page 27: Specifications

    Auto power off after 2 minutes without action when a strip is inserted, or 60 seconds if there is no strip inserted. Traceability of glucose monitoring system: The accuracy of (model: 5228P) Blood Glucose Monitoring EverPaw System was assessed by comparing blood results obtained by...
  • Page 28: Ordering Supplies

    This warranty does not cover device failure due to owner misuse or negligence, or normal wear and tear. If you have a question about your EverPaw (model: 5228P) Glucose Meter or this warranty, please contact: EverPaw Brooklyn, NY 11219 Questions or Comments?