After having installed the autoclave, proceed with preparations and use.
8.1
Turn on the autoclave
Press the general switch (FIG.2–pos.
8.2

Loading pure water tank

Connect the water filling pipe (provided) with the connection for water filling pipe and
plastic funnel PICT.14. Then engage the connection in the charging hole (FIG.2–pos. 13).
Insert the other end of the pipe with filter in the demineralized or distilled water container.
Press the button
B-PUMP
start.
The pump loads the tank inside the autoclave. If maximum level is not reached within 180
seconds the pump automatically stops, therefore it is necessary to press the button again.
The pump stops automatically when maximum level is reached.
If you wish to use this method of filling or in the event of malfunctioning of the pump, the
filling of the tank can be carried out manually by the operator in the following way:
1. Turn on autoclave.
2. Remove the connection cover (FIG.2–pos.
3. Insert in this opening the hose connected to the funnel (FIG.15)
4. Pour distilled water in the funnel, keeping it higher than the loading top.
5. Add until the led of maximum level turns on (FIG.2–pos.
6. Loading completed, twist on the top.
With connection to a deionizer, if the maximum water level has not been reached, the
functioning of the autoclave will be obstructed.