  • Page 1 EDMUNDS GAGES Accu-Touch Amplifier Instruction Manual...
  • Page 2 Attach any LVDT cables from tooling to the rear of amplifier noting proper location of connectors. Also attach any air lines from the tooling to the front of the Accu-touch docking unit, if applicable, noting their proper connection.
  • Page 3 .5% of full scale (LVDT) and .1% full scale (AIR). Linearity will be .25% full scale. The Accu-Touch amplifier is designed to be used with Edmunds LVDT gage probes and also two master system air tooling when used with the optional “docking unit”. No inference to compatibility with other types or brands of LVDT probes is implied or guaranteed.
  • Page 4: Table Of Contents

    Contents Hardware What you need Setup - section 1 Features - section 2 Formulas – section 3 Database setup – section 4 Setup Database Export/Import Setups Manual calibration – section 5 Setting pneumatic mag – section 6 Software calibration – section 7 Auto air-shutoff setup –...
  • Page 5: Hardware

    Hardware Front view Amplifier Touch screen Front USB port - data offload only Zero adjusting screw Mag adjusting screw Air hose, tooling connection Docking unit air gaging – optional Rear view Power on/off switch RS232C serial port LVDT ports rear USB port - data offload or optional data send/TIR reset footswitch LVDT jumper cable,...
  • Page 6: What You Need

    3.) Certifications by a traceable calibration lab are recommended to ensure the setup and results of calibration are as accurate as possible. 4.) A zero or nominal master can be used but not recommended by Edmunds Gages for electronic gaging only. Two masters are required for air gaging.
  • Page 7 Setup – section 1 1.1 – Main screen - shown with 4 check results. 1.2 - For a new setup, touch “Setup”, “Setup Access” menu appears. 1.2.1 – Touch “Measurement Setup”, setup now appears - if unlocked or if password protection has not been enabled. 1.2.2- The (11) selectable items related to for your use are available to be selected or entered by touching each software key.
  • Page 8 Features – section 2 Sensor type – touch to select “electric” for electronic gaging or air-auto for air gaging. Units – touch “inch” or “mm” based on application. Resolution – touch as many times as necessary to select resolution required for application. Two to six decimal place resolutions available.
  • Page 9 Features - continued Nominal – touch and enter the nominal size of your application using the software keyboard as a “whole” number if using absolute method or zero’s if using deviation method. Touch “Done” when complete. Upper limit/Lower limit – touch and enter the upper and lower limits of the part tolerance using the software keyboard.
  • Page 10 Measurement (Formula) – section 3 3.1 – Formulas – if the standard measurement choices shown under “Measurement” do not apply or if a special combination of inputs or formulas are required, the formula generator option is available. The formula generator is available on base version E7204A.
  • Page 11 Measurement (Formula) – continued 3.2.1 – Formula Generator - Touch the Setup software key and Setup Access menu appears. Enter the password if required, if not touch the System Setup key to display System Setup. 3.2.2 – If the optional Formula Generator (base version E7204A) has been purchased, the Formula Generator key will be present.
  • Page 12 Measurement (Formula) – continued 3.2.4 – If one of the existing formulas is not appropriate, touch the New Formula software key at the lower left of the screen and formula generator screen allowing the creation of the formula appears. See (Appendix A – Formula Generator) for complete description. 3.2.5 - Gaging formulas –...
  • Page 13 Measurement (Formula) – continued It may be necessary to apply a multiplier to the probes outputs depending on the application. The multiplier is determined by the number of probes used to perform the measurement or to correct any chordal error if they are not equally spaced.
  • Page 14 Measurement (Formula) – continued 3.2.6.1 – Select the “Type” key to select the “Multi-Check” format instead of the default “Input” format. Enter the formula “C1-C2” by selecting C1 under “Check” selection box. Select minus “-“ under the “Operators” selection box. Select C2 under the “Check”...
  • Page 15: Database Setup - Section

    Database setup – section 4 4.1 - Copies of the entire setups can be stored onboard of the Accutouch amplifier for future use. With optional version setups can be Exported/Imported to USB drive. 4.1.1 – storing new setups - touch “Setup”, “Setup Access” menu appears. Touch “System Setup”...
  • Page 16 Setup Name key. A software keyboard will appear allowing up to ten alpha numeric characters to be selected. Once finished press the Done key. 4.1.3 – Saving a Setup - the setup can be saved to the Accu-touch memory by touching the “Save to DB” key. Once saved the name will appear in the list.
  • Page 17: Export/Import Setups

    USB drive. The “AccuTouchSetups” directory will be created if does not already exists. 4.2.2 – Import a Setup - Insert Edmunds qualified USB drive into either USB ports. The USB drive must contain valid setup file(s) located in “AccuTouchSetups” directory on the USB drive. Touch the “Import from USB”...
  • Page 18 Select the setup to import by touching the setup name shown or scrolling down or up to the setup name required. The system will respond with a “YES” “NO” prompt to confirm. . If the setup already exists in the internal database a warning before overwriting the existing setup will be displayed.
  • Page 19: Manual Calibration - Section

    Manual calibration – section 5 5.1 - Manual adjustment – performed first time gage is used with amplifier. 5.1.1.1- Electronic gaging - “single master”, touch “Calibrate” key and “Measure 1” will appear. 5.1.1.2 -Touch “Adv Cal Input Setup” key and the manual calibration menu appears.
  • Page 20 Manual calibration - continued 5.1.2.1- Electronic gaging “two master”, touch “Calibrate” key and “Measure 1” will appear (as previously shown). 5.1.2.2 - Touch “Adv Cal Input Setup” key and the manual calibration menu appears. Touch “select displayed results” if necessary to display “Uncalibrated”...
  • Page 21: Setting Pneumatic Mag - Section

    Setting pneumatic mag – section 6 - Adjusting pneumatic mag & zero example: Master 1 = .001; Master 2 = -.001 Note: when first purchased new with tooling, these settings will be pre- adjusted for the tooling . 6.1.1 -If first time setup, turn both mag (M) & zero (Z) screws carefully clockwise until they stop and unscrew both counterclockwise 2 ½...
  • Page 22: Software Calibration

    Setting pneumatic mag - continued Then turn the (Z) zero screw in the same direction as the (M) mag screw was just turned until the digital reading is close to or equal to the min master size. -.001000 zero MIN size = good 6.3.3 - set mag (spread too long) - If the digital reading for the min master size is smaller than it should be, the spread is “long”, mag adjustment is required.
  • Page 23 Software calibration – section 7 7.1- - If this is the first time the gaging is being used with this readout, the following software calibration procedure can be utilized after manual calibration has been verified. If this is not the first time, the manual calibration procedure is not necessary.
  • Page 24: Auto Air-Shutoff Setup

    Auto-air shutoff setup – section 8 8.1 – This feature can be used with air gaging when the amplifier is Model #E7204 with the capability to control the Edmunds auto air shut-off package. This combination of hardware and software allows the supply air pressure to be automatically cut down by approximately 85% of its normal volume when not in use.
  • Page 25 Auto-air shutoff setup - continued 8.1.2 – Software – the auto-air feature is enabled or disabled under the System Setup menu. Touch the Setup software key on the main menu and a pop-up menu, Setup Access will display. 8.1.2.1 - If password protected, enter password first and then Touch System Setup.
  • Page 26 Auto-air shutoff setup - continued 8.1.2.6 – Turn Air ON/OFF – touch this key to manually over-ride the settings and turn the air on or off. 8.1.3 – Manual adjustment – the “silver” colored knurled adjusting need valve on the air shut-off valve will need to be adjusted each time a different air gage is used with this readout.
  • Page 27: Measuring – Section

    Measuring – section 9 9.1 – Once successfully calibrated, the amplifier then can be used to measure parts. 9.1.1 – The display can be selected to show a single measurement or up to four measurements, if setup, at one time. If multiple measurements are setup, touch the screen or the software keys at the bottom of the screen to select.
  • Page 28: Data Offload Setup - Section

    10.1 – At anytime, the readings being displayed can be offloaded to either USB ports, front or back, if an Edmunds qualified USB drive is installed or to an external device thru the 9 pin serial port at the rear of the unit. No onboard data storage capability is available.
  • Page 29 Data offload setup - continued 10.2.4 - The data on the USB drive can be viewed by going to the System Setup menu as previously described. Once in the System Setup menu, touch the RESULTS VIEWER key and all data on the USB drive will show. The data can be viewed in the column format or the chart format.
  • Page 30 Data offload - continued 10.3 – Serial setup – touch the setup software key and a pop-up menu, Setup Access, will display. If password protected, enter password first and then touch System Setup. 10.3.1 - At the top of the System Setup menu there is a Method key to select the type of output.
  • Page 31 Data offload - continued 10.3.4 - The RS232C serial port provides a path to communicate to external devices such as a Data Collector, Personal Computer (PC), or serial printer. The RS232C male connector is a standard 9-pin “D” style and is located on the back of the Accutouch amplifier.
  • Page 32 Reports – section 11 11 – A copy of the complete setup can be offloaded to a USB drive for record purposes. 11.1 – Touch the setup software key and a pop-up menu, Setup Access, will display. If password protected, enter password first and then touch System Setup.
  • Page 33 Reports – continued 11.4 – If the report is to be offloaded, ensure the USB drive is installed and touch the YES key. If the drive is recognized and there is not an existing copy, the message will change to “Report Completed Successfully”. Touch the OK key and the Done key twice to exit out.
  • Page 34: Setup Report

    Reports – continued example: ------------------- SETUP REPORT -------------------- ---- Date-Time Printed ---- Date: 3/14/2011 Time: 1:42 PM ---- AccuTouch Unit ---- Serial Number: 0123456 Release Date: 10/06/10 Software Version: 1.5.0.4 ---- System Setup ---- Setup Name: DEFAULT Offload Type: USB Offload Preference: USB ---- Check Setup ---- Number of Checks: 2...
  • Page 35: Password Protection

    12.1.2.1 – Touch the Change Password key and a software keyboard appears, Enter New Password. Enter a password constructed of up to any (4) alpha numeric characters. A master password can be obtained from Edmunds Gages to override this password should it be lost or forgotten *Note- this master password can never be changed so only authorized personnel should have access to it.
  • Page 36 Utilities – continued 12.1.2.2 – Once the selected password has been entered, touch the DONE key at the lower right hand location on the screen and a popup message will ask if the new password should be enabled. Touch YES to proceed or NO to cancel selection.
  • Page 37 Utilities – continued 12.1.3.2 - If NO, the Unlocked key will remain green and all access is allowed. 12.1.4 – Removing password protection – with the Unlocked key showing, touch the System Setup key and then the Change Password key. With the password showing as in section 11.2.1, touch the “Space”...
  • Page 38 Utilities – “ABOUT” 12.2 – “ABOUT” – is the feature that lists the serial number, software version and release date of the operating system loaded on the Accutouch amplifier. 12.2.1 – Touch the Setup software key and Setup Access menu appears. 12.2.2 - Touch the System Setup key, System Setup menu displays, with or without a password enabled.
  • Page 39 Utilities – “setting date & time” 12.3 – Date & Time – can be set or adjusted in the System Setup menu. 12.3.1 – Touch the Setup software key and Setup Access menu appears. 12.3.2 - Enter the password, if required, after touching the Locked key and then the DONE key.
  • Page 40: Screen Colors

    Utilities – “Screen Colors” 12.4 – Screen colors – the background color can be set to either the default blue color or an optional gray in the System Setup menu. 12.4.1 – Touch the Setup software key and Setup Access menu appears. 12.4.2 - Enter the password, if required, after touching the Locked key and then the DONE key.
  • Page 41 Utilities – backing up 12.5 – The operating system can be backed in the amplifier as a safeguard in case the active version ever becomes corrupted and inoperable. If an issue does arise, the system can be restored by loading a stored copy from memory. 12.5.1 –...
  • Page 42 Utilities – continued 12.5.3 – Restoring backup – Touch this key to start the restore process. A Restore System warning message will appear providing an opportunity to stop this overwrite of the running copy. Touch the No key to stop the process and return to the Backup/Restore menu.
  • Page 43: Notes

    Notes 1.) If “Locked” key appears, touch “Locked” and enter the password. Touch “done” when complete. Obtain password from authorized individual to unlock. 2.) Advanced air settings are available if the air tooling is non-standard or worn. Use only if calibration fails and manual adjustment is correct. 3.) Resolution selected will effect total range available to use for measurement.
  • Page 44: Appendix A - Formula Generator

    Appendix A – Formula Generator Description: Description of formula – 14 character max. Type: Type can be set to either “Inputs” or “Multi-Check”. Inputs: If the formula is to be based off sensor (LVDT or Air probe) the type should be set to “Inputs”. When the type is set to “Inputs” selections “A”,”B”,”C”,”D”...
  • Page 45 Operators: Mathematical operators that can be used in formula. : addition : subtraction or negative : multiplication : division : raise to the power of, example 4^2 = 16 Functions: Mathematical functions that can be used in formula. All functions that return an angle value are in Radians.
  • Page 46 Display Results in Degrees: If the formula has been programmed to return an angle measurement, the results can be displayed in degrees (deg:min:sec) format. If this option is not selected the results will be display as decimal degree. Note that functions will return angle values in radians and will have to be converted to degrees in the formula by multiplying result by (180/Pi).
  • Page 47: Definitions

    Definitions Absolute is the part size displayed as a whole number. An A/E (Air to Electric) transducer converts changes in pneumatic pressure into an electrical signal. LIVE is a function that displays the reading as a constantly changing value in real time.
  • Page 48 Definitions - continued Polarity is the (+) or (-) indicator applied to the magnification of an input to correctly display that reading based upon the direction the input value changes. In +PEAK (or -PEAK) modes display the largest or smallest size reading seen during the movement of the part after it has been reset.
  • Page 49: Troubleshooting

    Trouble shooting Error Common causes Unit will not turn on Incorrect supply voltage No supply voltage Blown fuse in Amplifier rear power connector. Fault in Amplifier power switch Pneumatic mag & zero for air tooling cannot be Contamination in A/E docking unit Damaged air tooling or wear on nozzles.