Zico ZI-9100 Instruction Manual

Digital insulation digital insulation tester

Advertisement

Quick Links

DIGITAL INSULATION TESTER
INSTRUCTION MANUAL
ZICO
1
ZI-9100
SAFETY INFORMATION
Read the following safety information carefully before
attempting to operate or service the meter.
To avoid damages to the instrument do not apply the
signals which exceed the maximum limits shown in the
technical specifications tables.
Do not use the meter or test leads if they look damaged.
Use extreme caution when working around bare
conductors or bus bars.
Accidental contact with the conductor could result in
electric shock.
Use the meter only as specified in this manual;
otherwise, the protection provided by the meter may be
impaired.
Read the operating instructions before use and follow
all safety Information.
Caution when working with voltages above 60V DC or
30V AC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing
acoustic continuity, disconnect circuit from main power
supply and all loads from the circuit.
2

Advertisement

loading

Summary of Contents for Zico ZI-9100

  • Page 1  To avoid damages to the instrument do not apply the ZICO ZI-9100 signals which exceed the maximum limits shown in the technical specifications tables.  Do not use the meter or test leads if they look damaged.
  • Page 2: Specifications

    Safety symbols: Display: Large LCD with dual display Measurement Range: 200Ω, 200kΩ, 200MΩ/250V, Caution refer to this manual before using the meter. 200MΩ/500V, 2000MΩ/1000V, 750V/ACV, 1000V/DCV. Dangerous voltages. Sampling Rate: 2.5 times per second. Zero Adjustment: Automatic adjustment. Meter is protected throughout by double insulation or Over Range Indicator: Number 1 of highest digit is reinforced insulation.
  • Page 3 1-2 Electrical Specifications AC Voltage (40Hz~400Hz) Accuracies are specified in the way: Range Resolution Accuracy Input Overload ±(…% of reading +…digits) at 23ºC±5ºC,below 80% RH. Impedance Protection 750V +(1.2%+10) 10MΩ 750Vrms OHMS Range Resolution Accuracy Max. open Overload Meg OHMS Circuit Protection Range...
  • Page 4: Parts And Controls

    2. PARTS & CONTROLS 3. BATTERY REPLACEMENT 3-1 How to connect test leads. a). On MΩ Range: Connect the red test lead into the “VΩ” terminal and the black lead into the “COM” terminal. b) On 200Ω and ACV Range: Connect the red test lead into the “VΩ”...
  • Page 5: Ac/Dc Voltage Measurements

    that subdivision of large installations might be necessary likely to be damaged if tested at 1000V. Examples are because of the large number of parallel insulation power factor correction capacitors, low voltage mineral resistance. In such a case, an installation may be divided insulated cables, electronic light dimmers, electronic ballasts and starters for fluorescent lamps etc…...
  • Page 6 7. POWER TOOLS AND AMALL brush rigging by connecting one megohmmeter lead to each APPLIANCES of them individually, leaving the other connected to the grounded motor housing. The above also applies to DC This test would also apply to other similar equipment that has Gemerators.
  • Page 7 CABLES Disconnect the cable from the line. Also disconnect opposite end to avoid errors due to leakage from other equipment. Check each conductor to ground and /or lead sheath by connecting one megohmmeter lead to a ground and /or lead sheather and the other megohmmeter lead to each of the conductors in turn.