Important Information About the SenseWear
Patient Environment
1.5m
Computer and Wireless Communicator
Guidance and Manufacturer's Declaration - Emissions
®
The BT-2.4-WS (SenseWear
biotransceiver) is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the BT-2.4-WS should ensure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonics
Class A
IEC 6100-3-2
Flicker
Complies
IEC 6100-3-3
5
®
biotransceiver
2.5m
1.5m
1.5m
Diagram not to scale.
Electromagnetic Environment - Guidance
The BT-2.4-WS uses RF energy only for its inernal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The BT-2.4-WS is suitable for use in all establishments,
including domestic, and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Important Information About the SenseWear
Guidance and Manufacturer's Declaration - Immunity
The BT-2.4-WS (SenseWear
®
biotransceiver) is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the BT-2.4-WS should ensure that it is used in such an environment.
IEC 60601
Immunity Test
Test Level
ESD
±6kV Contact
±6kV Contact
IEC 61000-4-2
±8kV Air
±8kV Air
EFT
±2kV Mains
N/A
IEC 61000-4-4
±1kV I/Os
Surge
±1kV Differential
N/A
IEC 61000-4-5
±2kV Common
Voltage Dips/
>95% Dip for
N/A
Dropout
0.5 Cycles
IEC 61000-4-11
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5
Seconds
Power
3A/m
3A/m
Frequency 50/60Hz
Magnetic Field
IEC 61000-4-8
®
biotransceiver
Compliance
Electromagnetic
Level
Environment - Guidance
Floors should be wood, concrete, or ceramic
tile. If fl oors are synthetic, the r/h should be at
least 30%.
N/A
N/A
N/A
Power frequency magnetic fi elds should be that
of a typical commercial or hospital environment.
6