The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
Page 3
United State es. If meter d isplay shows s “mmol/L” d during settin g or test, ple ease contact iGlucose Dia abetes Custo omer Support Center. Use e of the wron ng unit of me easurement may cause in ncorrect trea atment.
Page 4
- Do not perform the blood glucose test at temperatures below 10°C (50°F) or above 40°C (104°F), nor below 10% or above 90% relative humidity. - iGlucose Blood Glucose Test Strips are designed for use with capillary whole blood samples. Do not use serum or plasma samples.
Page 5
- Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar patients. - Not for neonatal use.
Table of Contents PAGE About the Blood Glucose Monitoring System The iGlucose Blood Glucose System The iGlucose Blood Glucose Meter The iGlucose Test strip Meter Activation and Battery Charging Setting Up Your Meter - Setting the Date, Time Turning On/Off the meter...
GC55 (Lev vel 2 and Lev el 4 of Right est Control S Solution GC5 550 are availa able for the iGlucose Blo Gluc cose Monito ring System but must be e purchased s separately.) 8. iGlu cose Lancing g Device* 9.
The iGlu ucose Blood Glucose Me The iGlu ucose Blood Glucose Me ter Icons Spelled o out on Average resu display ndicates a co ontrol solutio on test result mg/dL Unit of test re esult Test result Warns when the batteries s are low or must be rec harged...
45 minutes bef ore measure ement. - For det tailed inform mation, pleas se refer to th he iGlucose B Blood Glucos e Test Strip, Package Ins ert. -The use e of other tes st strips may y cause strip error.
Insert the battery 1. Take the battery out of the plastic bag. 2. Open the back cover of the device by inserting your thumbnail into the small groove on the right-hand side near the bottom of the back cover and lifting up the cover. 3.
Removin ng the iGluco ose Blood Gl ucose Test S Strip Hold the iGl ucose Blood Glucose Tes st Strip as sh own. Rotate the iG Glucose Bloo od Glucose T Test Strip cou unterclockwi se and pull u up simultane eously.
Page 13
1. Hold t the adjustab ble depth cap p of the lanci ing device in one hand w while holding the base in other ha and. Slightly force downw ward. 2. Pull th he two parts s in opposite directions to o remove th e adjustable depth cap.
Close th he vial cap im mmediately. 4. Inser t the strip in nto the test s trip port of t the iGlucose Blood Gluco ose meter wi ith the view window facing g up. Once t he strip is in...
Page 15
9. Pull o off the depth h adjustable cap. Withou t touching th he used disp osable lance et, insert the lancet tip in to the prote ective lancet cover. 10. Hold d the release e button in o ne hand and d pull on the plunger with h the other h...
If you ar re experienci ing symptom ms that are n ot consisten nt with your b blood glucos se test result ts and you have e made sure e to follow al l instructions s in this man nual, contact your health care profess sional...
Center. When sh hould a qual lity control t test be perfo ormed? - To ens ure that you ur iGlucose B lood Glucose e meter and iGlucose Blo ood Glucose Test Strip ar working g properly. - To con...
Perform ming a Qualit ty Control Te 1. Take o one iGlucose e Blood Gluc ose Test Stri p from the v vial and close e the vial cap p immediate 2.
Page 20
6. Place a drop of co ontrol solutio on onto the t top of the ca 7. Gently y touch the s sample port of the strip w with the con ntrol solution n from the to op of the cap 8.
- Your B lood Glucose e Test Strip h has expired. - You lef ft the cap of the iGlucose e Blood Gluco ose Test Stri p vial or the control solu ution off for a a period of time.
If your test strip port is stained with blood, control solution or any liquid, please use disinfecting wipes listed on next page to clean it immediately. Do not immerse in water. Perform a quality control test to ensure the iGlucose Blood Glucose meter is working properly. Cleaning and Disinfecting Procedures Indirect transmission of Human Immunodeficiency Virus (HIV), Hepatitis B Virus (HBV) and Hepatitis C Virus (HCV) during the delivery of healthcare services has been increasingly reported.
Page 24
To clean n the iGlucos se lancing de vice: 1. Thoro oughly wipe t the entire su urface of the iGlucose lan ncing device with disinfec cting wipes l isted above to o clean any p possible dirt, dust, blood and other b body fluids.
NOTE - Your iG Glucose Lanc cing Device h as been test ted to ensure e that there is no change e in the perfo ormance of the de evice after 5 550 cleaning cycles and 5 550 disinfecti ing cycles. Th he testing sim mulates 2 cle eaning...
Page 26
2. When the temperature is below 50 °F (10 °C) or over 104 °F (40 °C), the meter will not function and the " " symbol will blink. 3. If the iGlucose Blood Glucose meter and Test Strips are exposed to a substantial change in temperature, please wait 45 minutes before measurement.
Page 27
Samplin ng Error Please d do not apply the blood to o the sample port of the t test strip bef fore the met ter displays: “apply a drop of blood”. Othe erwise, the m meter will dis splay an erro or message.
the test strip. Please re-insert the unused test strip correctly. The screen MUST show the image of the test strip accompanied by the blood drop symbol flashing (this symbol take 3 seconds to show after correctly inserting the test strip) in order for a sample to be applied. Please consult your User Manual and/or contact Customer Service for support on how to correctly insert a Blood Glucose Test strip.
Test Sample Fresh capillary whole blood Test Result Plasma / Serum glucose Sample Size Less than 0.5 μL Measuring Time Less than 5 seconds Measuring Range 20mg/dL~600mg/dL Hematocrit Range 20% ~ 60% Working temp 10℃ ~ 40℃, Size - 2.4" QVGA Hi contrast TFT Colors 260K Resolution - 320 * 240 pixels...
This warranty does not apply to the performance of an iGlucose Blood Glucose meter that has been altered, misused, tampered with or abused in any way. This warranty applies only to the original purchaser of the iGlucose Blood Glucose Monitoring System.
Manufacturer: SteriLance Medical (SuZhou) Inc. No.68 LiTangHe RD, XiangCheng, SuZhou, JiangSu 215133, P.R. China General Electric Company and is under license by Bionime Corporation, No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan (R.O.C) Log Book Name: Address: Home Phone: Work Phone:...
(Please present this card for product replacement) Limited warranty: 1. Bionime warrants to the original purchaser only, that the iGlucose Blood Glucose meter shall be free of any defects in materials or workmanship and, provided it is not modified, altered or misused, will perform in accordance with specifications for a period of five (5) years from the original date of purchase.
Emergen ncy Card iGlucose e Blood Gluco ose Monitor ring System • User N Name: • Emerg gency Contac ct Phone No. • Blood Type: • Doctor r/Hospital: I am a p erson with d diabetes. If yo ou find me in...