Adjusting IF Gain for Bypass Path

Complete the following procedure to measure and adjust the IF gain for bypass path
performance of the NI 5694 IF conditioning module.
1.
Zero and calibrate the power sensors using the built-in functions in the power meter.
2.
Ensure that the hardware is properly configured as described in
3.
Run the self-calibration procedure for the NI 5622 digitizer.
4.
Initialize an external calibration session for the NI 5694.
5.
Initialize an IF attenuation calibration step.
6.
Configure the NI 5694 as follows:
Signal conditioning: Bypassed
IF conditioning downconversion: Disabled
Step gain: Disabled
7.
Configure RF source 1 as follows:
Single frequency mode
Frequency: 107.5 MHz
Power level: 0 dBm
8.
Configure the power meter as follows:
Channel 1: Power sensor A
Trigger with settling delay
Sensor setting: 0.1%
9.
Configure the NI 5622 digitizer as follows:
Sampling rate: 585.9375 kS/s
Number of samples to acquire: 40,960
Reference source: PXI_Clk
Reference Clock rate: 100 MHz
DDC enabled: TRUE
Data processing mode: Complex
Frequency translation enables: TRUE
DDC center frequency: 107.5 MHz
Dither enabled: FALSE
Bandpass filter enabled: FALSE
10. Read the power meter channel A power.
11. Calculate the NI 5694 IF input power using the following equation:
IF Input Power = Channel A Power + Splitter Tracking at RF Source Frequency