PREFACE Thank you for choosing CHARDER MEDICAL product. All features of this product were designed to state of the art and are optimized for simple and straightforward use. If you have any queries or experience any problems not addressed in the operating instructions, please contact your CHARDER MEDICAL service partner, or visit us on the Internet at www.chardermedical.com...
The regularity of these checks is dependent on the level of use and the state of the scale. If any inaccuracies occur, please contact your local dealer or CHARDER MEDICAL service partner.
, overuse, chemical, electrochemical or electrical interference or humidity, unless this is attributable to negligence on the part of CHARDER MEDICAL. If operating, climatic or any other influences lead to a major change in conditions or material quality, the treaty for perfect unit functioning shall be rendered null and void.
RECYCLING This product is not to be treated as regular household waste, but should be handed into an electrical/electronic equipment recycling centre. You can obtain further details from your local council, your municipal waste disposal company or the firm which you purchased the product. READING BEFORE USE Before reading detailed instructions on how to use all the weighing functions that are built into your scale, please read the following...
EXPLANATION OF THE GRAPHIC SYMBOLS SN-T13000001 Designation of the serial number of every device, applied at the device. (Number as an example) “Please note the accompanying documents“ or “Observe operating instructions” Identification of manufacturer of medical product including address Charder Electronic Co., Ltd. No.103, Guozhong Rd., Dali Dist., Taichung City 412, Taiwan (R.O.C.) Type B applied part...
Page 8
Carefully read this operation manual before setup and commissioning, even if you are already familiar with Charder scales. Transport and storage temperature limit indicating the upper and the +60°C lower limit (Transport and storage temperature -20°C on packaging)
Guidance and manufacturer’s declaration-electromagnetic emissions The MEDICAL SCALE MS4400 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS4400 should assure that it is used in such an environment. Electromagnetic...
Page 10
Guidance and manufacturer’s declaration-electromagnetic immunity The MEDICAL SCALE MS4400 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS4400 should assure that it is used in such an environment. IEC 60601 test...
Page 11
UT(>95% dip in mains interruptions, it is UT(>95% dip UT) for 5 s recommended that the in UT) for 5 s MEDICAL SCALE MS4400 be powered from an uninterruptible power supply or a battery. Power 3 A/m 3 A/m The MEDICAL SCALE...
Page 12
Guidance and manufacturer’s declaration-electromagnetic immunity The MEDICAL SCALE MS4400 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS4400 should assure that is used in such and environment. IEC 60601 test...
Page 13
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MEDICAL SCALE MS4400 is used exceeds the applicable RF compliance level above, the MEDICAL SCALE MS4400 should be observed to verify normal operation.
Page 14
Recommended separation distance between portable and mobile RF communications equipment and the MEDICAL SCALE The MEDICAL SCALE MS4400 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MEDICAL SCALE MS4400 can help prevent electromagnetic...
SPECIFICATION Capacity 10kg 15kg Division Accuracy for OIML ±15g ±30g APPROVAL Accuracy for ±20g ±40g Non-OIML APPROVAL Operating Temp. and 5℃ / 35℃ Humidity 15% - 85% RH Transport and Storage - 20° C + 60° C Temp. and Humidity 10% - 95% RH Unit kg unit is approval of OIML...
KEY FUNCTIONS KEY DESCRIPTIONS ON/OFF/ZERO 1. Press to turn the scale on. 2. Press this key and hold for 3 seconds to turn the scale off. 3. Press this key with the scale on to zero the scale. HOLD 1. Press this key to release the HOLD function. TARE 1.
CAUTIONS BEFORE WEIGHING VISUAL INSPECTING HOOK AND SLING ARE WITHOUT ANY BREAKAGE PERFECT PERFECT CLEANING AND REMOVE ANY SHARP AND HARD OBJECTIVE NEAR BY WEIGHING TABLE. cleaning table IT IS RECOMMENDED TO PLACE A CUSHION UNDER BABY SLING BEFORE WEIGHING TO MAKE BABY COMFORT AND SAFE.
CHARDER MEDICAL service partner for service. Zero counting value low Internal zero counting value is too low for operation. Please contact your local CHARDER MEDICAL service partner for service. Zero counting value error This warning show that the device detected the zero counting value fluctuates exceed ±10% as default...
TROUBLE SHOOTING Troubleshooting for defective modes: Original purchaser can enjoy the benefits under the effective Warranty against functional defects in material and workmanship subject to the terms and conditions listed in our Warranty Program & Return Policy. Our warranty service program includes the following: 1.
Page 31
Unstable platform feet adjustments according to bubble level indication Incorrect position or other external objects within weighing area The weighing-scale is not put in a solid & firm ground area, such as carpet floor or lawn. 3. Connection failure for data transmission to PC or printer Wrong connection wires or faulty wires for transmission between the digital indicator &...
Page 32
Key buttons failure - Front key panel damaged or disconnected If scale does not running Auto-Zero properly, please call for service from your Charder Authorized Dealer for service maintenance. 3. Buzzer malfunction Wrong welding of PVC wire Key buttons & control panel damaged or disconnected.
Page 36
Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonized European standards, following the provisions of the below stated directives: 93/42/EEC as amended by 2007/47/EC Medical Device Directive 2014/31/EU Non-automatic Weighing Instruments Directive Please see separate document showing on sticker of device for above CE marking.