Unit Operation
Applying the Arm Cuff
1. Firmly insert air plug into opening located on left side of monitor unit.
2. With sticky nylon section facing outward, insert end of cuff underneath metal ring
of cuff
.
3. Fasten cuff about
1-2cm (0.4-0.8'')
bare arm and keep level with heart while testing
Note: Do not insert air plug into opening located on right side of monitor unit.
This opening is designed for an optional power supply only.
Unit Operation

3. Testing

After cuff inflation, air will slowly subside as indicated by the corresponding cuff
pressure value. A flashing "
heart beat detection.
Note: Keep relaxed during testing. Avoid speaking or moving body parts.
4. Result Display
Three short beeps sound when testing is complete. The screen will display
measurements for systolic and diastolic blood pressure.
Indicator
Note: Refer to Page
18~19
for detail WHO Blood Pressure Classification Information
above the elbow joint. For best results apply cuff to
.
" will appear simultaneously on screen signaling
mmHg
M
D
11
Unit Operation
Testing
1. Power On
Press and hold
" START/STOP "
appear for one second as unit performs a quick diagnosis. A long tone indicates
device is ready for testing.
Note: Unit will not function if residual air from previous testing is present in cuff.
The LCD will flash "
2. Pressurization
Initial pressure is first pumped to 190 mmHg. If the current user's systolic blood pressure
is over 190mmHg, the unit will automatically re-inflate to the proper shelf.
13
Unit Operation
Irregular Heartbeat Indicator
If the monitor detects an irregular heart rhythm two or more times during the measuring
process, the Irregular Heartbeat Symbol "
measurement results. Irregular heartbeat rhythm is defined as rhythm that is either 25%
slower or faster than the average rhythm detected while measuring systolic blood pressure
and diastolic blood pressure. Consult your physician if the Irregular Heartbeat Symbol
"
" frequently appears with your test results.
5.Storing Test Results
It will automatically store by date .If the number of tests surpasses the allotted 120 memories ,
the most recent tests will appear first , thus eliminating the oldest readings.
Power Off
The " START/ STOP" button can be pressed to turn off the unit in any mode. The unit can
turn off the power itself about 3 minutes if no operation in any mode.
Safety Precaution: If pressure in arm cuff becomes too extreme while testing,
press the " START/STOP " button to turn power off. The cuff pressure will rapidly
dissipate once the unit is off.

Memory Check

With power off, press and hold MEM button to turn the unit on. The LCD will display the
average test results from the last 3 readings.Press MEM button again,the LCD will display
the last measurement memory as NO: 01 reading . Older test result in memory can be viewed
by pressing the MEM button.
"
"
Note: If the memory is less than 3 groups , press the "MEN" button to display the last group.
button until a beep sounds.
The LCD screen will
" until pressure is stabilized.
mmHg
M
D
"appears on screen along with
"
"
"
"
N o
M
D
12