Valhalla Scientific 4300C Operation Manual page 83

Programmable digital µ-ohmmeter
Hide thumbs Also See for 4300C:
Table of Contents

Advertisement

4300C User Manual
OHMS? QUERY
Responds with reading from the front panel display and causes an immediate update of
the conversion so repeated query requests gets most accurate data. Send range
command or AUTO prior to sending this to get the scaling factor.
The Ohms omega, milli and kilo characters are removed during remote formatting.
Syntax:
Example:
Response:
Power-on default = 0.000
Note: For Scientific notation/Engineering format, use RDNG?
*OPT? QUERY
Returns the option text strings factory installed on meter from nonvolatile.
Syntax:
Example:
VOLTAGE RANGE COMMAND
Selects a voltage range
Syntax:
Range number = 1 – 3
1 = 20mV
2 = 200mV
3 = 2V
OHMS?
OHMS?<crlf>
20.000<crlf>
(on 20m or 20k Ohm ranges)
<20.000<crlf>>
*OPT?
*OPT?<crlf>
Option(s) :
GPIB(IEEE488.2),BCD,PLC,HLC,TCM,AUTO,RS232,USB<crlf>
VRANGE <range number>
6-28