Valhalla Scientific 4300C Operation Manual page 74

Programmable digital µ-ohmmeter
Hide thumbs Also See for 4300C:
Table of Contents

Advertisement

:SYST:ERR?
QUERY Sends the System Error value to the remote host.
Cleared with *CLS<crlf>
Syntax:
Example:
Response:
Power-on default = 0
BCD COMMAND
Selects BCD mode on or off
Syntax:
Example:
Power-on default = OFF
Note: BCD mode on puts the logic level BCD on the rear panel connector and the ranges
on the relays outputs. See the BCD Interface section of this chapter for BCD connector
pin-outs and relay descriptions.
BCD QUERY
Responds with BCD mode on or off
Syntax:
Example:
Response:
Power-on default = OFF
Chapter 6 – Remote Operations (RS-232 Interface)
:SYST:ERR?
:SYST:ERR?<crlf>
1F<crlf>
BCD <ON or OFF>
BCD ON<crlf>
BCD?<crlf>
ON<crlf>
BCD?
BCD?<crlf>
ON<crlf>
<"ON" or "OFF">
6-19