ADJUSTMENT INSTRUCTIONS
3. Ignite the xenon bulb and check the current.
provide an accurate reading.
level of 180-185 amperes. If the current is not within the desired range, extinguish the bulb. It will be
necessary to increase or decrease the DC output.
4. Fine adjustment of the DC current is made to the NUMBERED taps found on the upper three
terminal blocks (TB4, TB5, TB6). Fine taps are numbered 1-2-3-4, with "1" providing the lowest
output, increasing to "4," yielding the highest output.
current approximately four amperes. The three fine tap terminal blocks are interconnected by
means of a three-lead jumper wire assembly attached to like-numbered terminals.
• To increase the DC output, move the jumper wire assembly to tap the next (3) higher numbered termi-
nals, for example, move from terminals "2" to terminals "3." ALL TAPS MUST BE ON THE SAME NUMBERED
POSITION (1-1-1, 2-2-2, 3-3-3, or 4-4-4).
nected, see the following instructions for adjusting "coarse" taps.
• To decrease the DC output, move the jumper wire assembly to tap the next (3) lower numbered
terminals, for example, move from terminals "3" to terminals "2." ALL TAPS MUST BE ON THE SAME
NUMBERED POSITION (1-1-1, 2-2-2, 3-3-3, or 4-4-4). If the DC output is still too high when terminals "1" are
interconnected, see the following instructions for adjusting "coarse" taps.
5. Coarse adjustment of the DC current is made to the LETTERED taps found on the lower three terminal
blocks (TB1, TB2, TB3). Coarse taps are lettered W-X-Y-Z, with "W" providing the lowest output, in-
creasing to "Z" at the highest output. The coarse tap terminals connect to contactor terminals T1, T2,
and T3. The (3) contactor leads must connect to the same lettered step (W-W-W, etc.).
tap adjustment raises or lowers the current approximately twelve amperes.
• To increase the coarse DC output, move each of the contactor leads to tap the next higher lettered
terminals, for example, move from terminals "W" to terminals "X." ALL TAPS MUST BE ON THE SAME
LETTERED POSITION (W-W-W, X-X-X, Y-Y-Y, or Z-Z-Z). Place the fine tap jumper on 1-1-1. Ignite the lamp,
check the output, and increase the fine tap setting as required.
• To decrease the coarse DC output, move each of the contactor leads to tap the next lower lettered
terminals, for example, move from terminals "Y" to terminals "X." ALL TAPS MUST BE ON THE SAME
LETTERED POSITION (W-W-W, X-X-X, Y-Y-Y, or Z-Z-Z). Place the fine tap jumper on 1-1-1. Ignite the lamp,
check the output, and increase the fine tap setting as required.
NOTE: Balance power supply output; set power supplies at or near the same output setting, i.e.:
6. Check the voltmeter and read the arc voltage. The voltage should read between 50 and 54 volts
DC. If the actual measured voltage is not in this range, consult your bulb supplier immediately.
7. Inspect tap connections to verify that the terminal is clamping the copper conductor, not the
insulation. Make certain all terminal clamping screws are tight.
8. Whenever making a coarse adjustment, again check the output current and make certain the cur-
rent is within the desired range.
taps.
9. After prolonged operation, the light output of the xenon bulb will decrease. This is a normal factor of
bulb aging, and can be compensated by raising the DC output of the xenon power supplies. If the
bulb was first operated at "nominal" current, the power supply output can gradually be increased
to, but not in excess of, the maximum current specified by the bulb manufacturer (210 A.). Increase
the current as instructed above. Decrease the power supply output to its former "nominal" current
level (180 A.) upon the installation of a new replacement bulb.
8
(continued)
A new 10000W-XT bulb should be first operated at the nominal current
If the DC output is still too low when terminals "4" are intercon-
Power Supply "A" X-X-X, 4-4-4 (or as required)
Power Supply "B"
Y-Y-Y, 1-1-1 (or as required)
A fine tap re-adjustment is frequently required after changing coarse
Allow (30) seconds for the current to stabilize and
A "fine" tap adjustment raises or lowers the
A "coarse"