Summary of Contents for Saebo SaeboStim Micro

  • Page 1 SaeboStim Micro Saebo INSTRUCTION MANUAL...
  • Page 3 This manual contains important information for both the person who will wear the SaeboStim Micro and the clinician who may provide and setup the device. Please be sure to review all information carefully. If you have questions or require further information, please contact: Saebo, Inc.
  • Page 4: Table Of Contents

    Index Features....................... 5 Indications for Use ..................6 Contraindictions ..................7 Warnings ..................... 8 Precautions ....................10 Adverse Effects ..................11 System Components .................. 12 Getting Started ..................13 Powering the Device ................... 13 Lead Wires ....................14 Reconnecting the Lead Wires..............14 Electro-Mesh Garment Electrode ..............
  • Page 5: Features

    Features • A one channel transcutaneous electrical nerve stimulator and electrical neuromuscular stimulator that delivers a pulsed DC current with a monophasic waveform to the surface area of the Garment Electrodes to provide electrical stimulation where there is an indication for use. • Microprocessor controlled, allowing easy alteration of the treatment parameters and precise control of each setting.
  • Page 6: Indications For Use

    Indications for Use TENS (Transcutaneous Electrical Nerve Stimulation) • Symptomatic relief and management of chronic intractable pain. • Adjunctive treatment for post-surgical and post-trauma acute pain. NMES (Neuromuscular Electrical Stimulation) • Relaxation of muscle spasm. • Prevention or retardation of disuse atrophy. • Increasing of local blood circulation.
  • Page 7: Contraindictions

    Contraindications TENS • Do not stimulate over the carotid sinus nerves, laryngeal or pharyngeal muscles (anterior throat area); severe spasm may occur causing contractions that may be strong enough to close the airway or cause difficulty in breathing. • Do not use TENS device on undiagnosed pain symptoms until the etiology has been established.
  • Page 8: Warnings

    Warnings TENS • Safety has not been established for the use of electrical stimulation devices during pregnancy. • TENS is not effective for pain of the central origin. This includes headaches. • TENS devices should be used only under the continued supervision of a physician or qualified professional.
  • Page 9 NMES • Safety has not been established for the use of electrical stimulation devices during pregnancy. • Long-term effects of chronic electrical stimulation are unknown. • Precautions should be taken in the case of persons with suspected or diagnosed epilepsy. • Precautions should be taken in the case of persons with suspected heart problems.
  • Page 10: Precautions

    Precautions TENS • Isolated cases of skin irritation may occur at the site of electrode placement following long-term application. • Effectiveness is highly dependent upon patient selection by a person qualified in the management of pain. NMES • Precautions should be taken in the presence of: Recent surgical procedures when muscle contraction may disrupt the healing •...
  • Page 11: Adverse Effects

    • Electrode placement and stimulation settings should be based on the guidance of the prescribing physician. • Powered muscle stimulators should be used only with the lead wires and electrodes recommended for use by the manufacturer. Adverse Effects Skin irritation and burns beneath the electrodes have been reported from use of electrical stimulators.
  • Page 12: System Components

    System Components SaeboStim Micro Device LCD Display Red Lead Wire Black Lead Wire Intensity Control Decrease TheraCream Intensity Control Power Button AAA Battery Increase Glove Arm Sleeve Velcro Arm Strap...
  • Page 13: Getting Started

    Getting Started Remove the SaeboStim Micro and all of the components from the packaging. Verify that all parts listed under the System Components section are present. Powering the Device The SaeboStim Micro is powered by one (1) AAA battery. Remove the battery compartment cover from the back of the device.
  • Page 14: Lead Wires

    If you accidentally disengage the Lead Wires from the unit, reconnecting is very simple. To reconnect the Lead Wire snaps to the SaeboStim Micro unit, insert the snap stud end into the inside position of the slot and slide toward the outside position to lock in place.
  • Page 15: Electro-Mesh Garment Electrode

    Electro-Mesh Garment Electrode Warning: The Electro-Mesh Garment Electrodes should only be used with electrical sources designed and approved for medical electrical stimulations. The Garment Electrodes are for single patient use. The Electro-Mesh Glove and Sleeve are highly conductive and made of silver treated nylon fibers blended with Dacron®...
  • Page 16: Applying The Device

    Applying the Device: Important: 1. Remove all jewelry and clean skin thoroughly. Electrodes do not work well if any lotion, oil, or dirt is on the skin. 2. Prior to putting the device on, it is important that you apply the conductive Thera- Cream over the treatment area (arm and hand).
  • Page 17 4. Apply the Arm Sleeve onto the arm and position over the elbow region with the Connector positioned on top. Make sure the Sleeve fits snugly over the entire treatment area. See Figure 10. Fig 10...
  • Page 18 5. Apply the Wrist Strap and secure just below the Arm Sleeve. Please be sure the Strap is not too tight, but is secured. The Strap will be positioned between the Arm Sleeve and the Glove. See Figure 11. Fig 11 6.
  • Page 19: Turning The Device On

    Turning the Device On The SaeboStim Micro is powered on by pressing and releasing the Power button for approximately 2 seconds until the display lights up showing P1 on the screen. See Figure 14. Only the “P” will be flashing. The Unit should be connected to the Garment Electrodes and the Garments should be worn at this point.
  • Page 20: Program Selection

    Research indicates that increased stimulation to the affected limb can lead to improved motor recovery. The SaeboStim Micro offers 2 novel programs (P1 and P2) that are widely used in scientific studies. The SaeboStim Micro and Electro-Mesh Glove and Arm Sleeve are very safe to use and can be used as often as possible –...
  • Page 21: Treatment Intensity

    Treatment Intensity Two buttons on the device control intensity: Increase ( ) and ▲ Decrease ( ▼ ). See Figure 16. The intensity defaults to 5 when the device is powered on or when the batteries have been removed and replaced.
  • Page 22: Parameters

    Parameters Waveform The SaeboStim Micro delivers a high voltage twin peak monophasic waveform. The maximum voltage coincides with the numeric intensity value on the LCD display (5-100 volts in increments of 5). Rate Rate selects the number of times the waveform repeats every second. Note that P1 is programmed at 50 (pulses per second) pps for 30 minutes for a total treatment of 30 minutes.
  • Page 23: Powering The Device Off

    0, release it and press it again and the device will turn off. Battery Care Fig 17 The SaeboStim Micro uses one AAA battery. To replace the battery, open the battery compartment on the back of the device and insert a new battery. Low Battery When the battery is low, a battery symbol will flash on the far right of the display.
  • Page 24: Care And Cleaning

    • Do not immerse the device in liquid. • Avoid spilling liquids on the SaeboStim Micro. • The surface of the SaeboStim Micro may be wiped with a soft cloth or sponge dampened with a mild soap solution. Avoid caustic cleansers.
  • Page 25: Trouble-Shooting

    Trouble-Shooting Device will not turn on: • Check for proper battery installation. • Replace battery. No Stimulation: • Check Lead Wires for proper connection onto the device. • Check for proper patient electrode application. • Apply Conductive Spray or Cream to the area to be stimulated. • Check treatment time for expiration.
  • Page 26: Helpful Hints

    Helpful Hints • Patients with sensory loss should not over-stimulate. Consult your health professional for proper setup and protocol. • Always prepare the skin correctly by washing it and removing all dirt, oils, and dead skin before applying electrodes. • If the stimulation becomes uncomfortable, causing a stinging sensation, when using the Electro-Mesh garments during treatment, re-apply TheraCream™...
  • Page 28 Saebo www.saebo.com...