CAS X320 Quick Start Manual
CAS X320 Quick Start Manual

CAS X320 Quick Start Manual

Ip69k washdown indicator
Hide thumbs Also See for X320:

Advertisement

IP69K Washdown Indicator
Quick Start Manual
(800) 223-4227
www.cas-usa.com
U21022050

Advertisement

Table of Contents
loading

Summary of Contents for CAS X320

  • Page 1 IP69K Washdown Indicator Quick Start Manual (800) 223-4227 www.cas-usa.com U21022050...
  • Page 2: Table Of Contents

    Quick Start Manual Rev 1.33 Table of Contents 1. INTRODUCTION.............2 1.1. Approvals..............2 1.2. Manuals ..............2 2. SHIPPING CONTENTS ..........2 3. SPECIFICATIONS ............3 4. WARNINGS ..............4 4.1. General ..............4 4.2. Configuration Issues ..........4 5. INSTALLATION ..............5 5.1. Electrical Safety ............5 5.2. opto-LINK..............6 6.
  • Page 3: Introduction

    Quick Start Manual Rev 1.33 1. Introduction This manual contains information on the installation, calibration and setup of the instrument. 1.1. Approvals • C-tick approved and CE approved. 1.1.1. Trade versions • NSC approval (4000 divisions at 0.8µV/division). • NMI approval (4000 divisions at 0.8µV/division). •...
  • Page 4: Specifications

    Quick Start Manual Rev 1.33 3. Specifications Performance Resolution Up to 30,000 divisions, minimum of 0.25µV/division (K307 up to 60,000 divisions) Zero Cancellation 2.0mV/V Span Adjustment 0.1mV/V to 3.0mV/V full scale Stability/Drift Zero: < 0.1µV/°C (+ 8ppm of deadload max) Span <...
  • Page 5: Warnings

    Quick Start Manual Rev 1.33 4. Warnings 4.1. General • Indicator not to be subject to shock, excessive vibration or extremes of temperature (before or after installation). • Inputs are protected against electrical interference, but excessive levels of electro-magnetic radiation and RFI may affect the accuracy and stability.
  • Page 6: Installation

    Quick Start Manual Rev 1.33 5. Installation The following steps are required to set up the indicator. • Inspect indicator to ensure good condition. • Use connection diagrams to wire up load cell, power and auxiliary cables as required. • Connect Power to indicator and press <POWER> key to turn the instrument On.
  • Page 7: Opto-Link

    Quick Start Manual Rev 1.33 5.2. opto-LINK The optional opto-LINK cable can be used to transfer setup and calibration information from a PC (eg. to be stored for later use and/or transferred to other instruments). It can also be used to download software upgrades to the instrument from a PC.
  • Page 8: Connections

    Quick Start Manual Rev 1.33 6. Connections 6.1. Cable Connections • All cable connections are made to the rear of the instrument using pluggable screw terminals. 6.2. DC Power (DC PWR + , DC PWR –) • The DC supply need not be regulated, provided that it is free of excessive electrical noise and sudden transients.
  • Page 9 Quick Start Manual Rev 1.33 6.3.1. 4-Wire Connection The minimum connectivity requirements are the connection of four wires (ie. Excitation + and – along with Signal + and –). The BUILD:CABLE option must be set to 4 to allow for 4-wire connection.
  • Page 10: Auxiliary Connections

    Quick Start Manual Rev 1.33 6.4. Auxiliary Connections This section provides diagrams to illustrate the terminal connections. 6.4.1. RS-232 Serial to PC Network: One Instrument to PC (DB9) (RXD, TXD, GND) Network: One Instrument to PC (DB25) (RXD, TXD, GND) Page 9 003X-608-133...
  • Page 11 Quick Start Manual Rev 1.33 Ring Networks: Multiple Instruments to PC (RXD,TXD,GND) Instruments can be configured in a Ring Network. The Short Ring Network layout should only be used up to a total cable run length of about 150 m (500 ft) at 9600 baud in a clean EMC environment.
  • Page 12 Quick Start Manual Rev 1.33 Long Ring Network The Long Ring Network layout can be used where each leg of the cable run can be up to about 150 m (500 ft) at 9600 baud. If comms errors persist, lower the baud rate to 4800 or 2400. This section provides diagrams to illustrate the terminal connections.
  • Page 13 Quick Start Manual Rev 1.33 6.4.2. Printer Connections (RXD/TXD, GND and DTR) RS-232 – Instrument to Printer (DB25) 6.4.3. Remote Display (TXD, GND) • Connect TXD to RXD and GND to GND on the remote display. Page 12 003X-608-133...
  • Page 14 Quick Start Manual Rev 1.33 6.4.4. Remote Input • The indicator requires a voltage free contact between TXD and RXD to enable the remote input (ie. SPEC:REM.FN). Note: The remote input will not function when in setup or when using the opto-LINK. WARNING The remote input is a voltage free contact (eg.
  • Page 15 Quick Start Manual Rev 1.33 Outputs to Drive Relay Outputs to Drive PLC Page 14 003X-608-133...
  • Page 16: Connecting Shields

    Quick Start Manual Rev 1.33 6.5. Connecting Shields To obtain full EMC or for RFI immunity, cable shields MUST be connected to the earth lug on the rear of the instrument. 6.5.1. Cable Shield Connection and Earthing • Care should be taken when connecting shields to maximise EMC or RFI immunity and minimise earth loops and cross- talk (interference) between instruments.
  • Page 17 Quick Start Manual Rev 1.33 6.6.1. Sealing Attachment of Lead Seal through optional Captive Security Screws. Please note that security screws are available as an option and that they are less corrosion resistant than the standard Captive Screws. Back View showing cable locations. Page 16 003X-608-133...
  • Page 18: Instrument Setup

    Quick Start Manual Rev 1.33 Instrument Setup 7.1. Calibration Counter The built-in calibration counter(s) monitor the number of times trade critical settings are altered. Refer to Trade Critical Settings below for more information and to the OPTION:USE setting page19. The table below describes when the counter(s) will increment. Industrial, NSC or OIML: The Calibration Counter increments when trade critical settings are changed.
  • Page 19: Access Full Setup

    Quick Start Manual Rev 1.33 7.3. Access Full Setup Full Setup provides access to configure and calibrate the instrument. All items in all menus will be enabled in Full Setup. WARNING: Care should be taken to avoid inadvertently altering the Build or Calibration settings. •...
  • Page 20 Quick Start Manual Rev 1.31 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults BUILD Decimal Point Position 000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000 Accept DP⊗ Maximum Capacity 000100 to 999999 Default = 003000 Accept CAP⊗ <SEL> changes position, <EDT> changes digit. Resolution (Count-By) 1, 2, 5, 10, 20, 50, 100 Accept...
  • Page 21 Quick Start Manual Rev 1.33 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults OPTION Zero 'Dead' Band <SEL> changes position, <EDT> changes digit. Accept Z.BAND⊗ (ctd) Default = 0 Enable Rear Pushbutton OFF, ON Accept R.ENTRY⊗ for Full Setup Entry Zero Calibration Routine <SEL>...
  • Page 22 Quick Start Manual Rev 1.31 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults Edit Linearisation Points <SEL> to step through list of points. ED.LIN⊗ <OK> to change selected point or <ITM> to exit. (ctd) Set between zero and full (K304 only) After <OK>...
  • Page 23 Quick Start Manual Rev 1.33 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults Direct Span Calibration <OK> key to start. Current weight displays. DIR.SPN⊗ <OK> to enter direct span setting (in mV/V). (ctd) (Current weight displays) <SEL> changes position, <EDT> changes digit. <OK>...
  • Page 24 Quick Start Manual Rev 1.31 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults SPEC SAFE.PC Safe Setup Passcode Accept (000000 no passcode). Set up to 6 digit passcode. <SEL> changes position, <EDT> changes digit. Activated only when FULL.PC is also set. FULL.PC Full Setup Passcode (000000 no passcode).
  • Page 25 Quick Start Manual Rev 1.33 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults SPEC REM.CHR Remote Input Transmit Idle Accept 000 to 255 Default = 005 Character <SEL> changes position, <EDT> changes digit. (ctd) BAT.VLT Battery Voltage Accept PWR, 9.6, 12, 24 W.D.LOCK...
  • Page 26 Quick Start Manual Rev 1.31 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults CHECK.W ENABLE Enable Check Weighing OFF, ON Accept Weight Source GROSS, DISP, NET (Displayed) Accept OVER Target for Over Weight -99999 to 999999 Default = 000000 Accept UNDER Target for Under Weight...
  • Page 27 Quick Start Manual Rev 1.33 GROUP ITEM SELECT EDIT (GRP) (SEL) (EDT) (ITM) Underline = Defaults TEST SCALE Scale Base Test Display Display reads in millivolts-per-volt (factory calibrated to 0.1%) Verify the correct load cell capacity and/or load cell wiring is used.
  • Page 28: Error Messages

    Quick Start Manual Rev 1.33 8. Error Messages 8.1. Weighing Errors • Check Setup = This item can be checked on site by service personnel Error Description Resolution (U - - - - -) The weight is below the Increase the weight or minimum allowable weight decrease the minimum reading.
  • Page 29: Setup And Calibration Errors

    Quick Start Manual Rev 1.33 8.2. Setup and Calibration Errors Error Description Resolution The instrument may be in Access Full Setup to edit (ENTRY) Safe Setup and an item that the item. (DENIED) needs Full Setup has been selected for editing. When accessing setup, Turn the instrument off.
  • Page 30: Diagnostic Errors

    Quick Start Manual Rev 1.33 8.3. Diagnostic Errors • Check: This item can be checked on site by service personnel. • Return for Service: The instrument must be returned to the manufacturer for factory service. Error Description Resolution (E0001) The power supply voltage is too Check supply low.
  • Page 31 Quick Start Manual Rev 1.33 Notes: Page 30 003X-608-133...