4.6
Trouble Shooting
Refer to the following troubleshooting guide for assistance with
possible meter malfunctions:
TROUBLE
Fluid will not flow
through the meter
Reduced flow through
the meter
Meter readings
inaccurate
Meter not giving pulse
signal
CAUSE
Meter clogged.
Line to meter
blocked.
Meter partially
clogged.
Line to meter
partially blocked.
Fluid flowrate is
not within meter
flow range.
Meter drag due to
worn or damaged
parts.
Faulty pickup coil.
Meter internals not
turning due to
worn or damaged
parts.
Maintenance
REMEDY
Clear meter.
Clear line to meter.
Clear meter.
Clear line to meter.
See "Specifications" for
min and max flowrates.
Replace worn or
damaged parts.
Replace pickup coil.
Replace worn or
damaged parts.
17
HP-210