NOTE: This User’s Manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. To ensure you have the latest documentation, please refer to www.tegam.com. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Figure 4: Main Board Parts Layout ..... 4-15 5. SERVICE INFORMATION Preparation for Repair or Calibration Service ..... 5-1 Expedite Repair and Calibration Form ...... 5-2 Warranty .............. 5-3 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Therefore in 3-wire mode, the test leads and their connections must be of identical resistance 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Note that the R1L-D1 is a 5 ½ digit instrument in order to provide 10 mΩ resolution when reading 1060 Ω RTD's. On lower ranges, this provides finer resolution, down to 1 µΩ...
Page 7
Note that the display may be used as an indication that power is on. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Page 8
Four binding posts +I, +E, -E, and -I. Proper connections to the resistor under test are described in Section 2.2 below. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
The R1L-D1 should be stored in a relatively dust free environment. Temperature: -40 to +71 °C. Relative humidity: 0 to 100%, non-condensing. See Section 4.7 below for shipping requirements. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
2.2 Preparation for Use • Release the two latches which secure the lid of the Model R1L-D1, and open the lid. Since the lid is well gasketed, it may be difficult to open if there is an appreciable difference between internal and external air pressure.
Page 12
After installing the cells, reinstall the battery cover with its four screws, and reinstall the instrument in its case and secure the ten mounting screws. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Page 13
, use +I and –I, +E and –E are not connected. WARNING DO NOT TOUCH THE BINDING POSTS WHEN THEY ARE CONNECTED EXTERNAL CIRCUITS. LETHAL VOLTAGES MAY BE PRESENT AT THESE POSTS. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
199,999 counts, it will automatically step to the next higher range. If it sees readings less than 20,000 counts, it automatically steps down a range, so that 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Page 15
CAUTION DO NOT USE EXCESSIVE FORCE TO OPERATE ANY OF THE PUSHBUTTONS. POWER OFF DOES NOT OCCUR UNTIL BUTTON IS RELEASED. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
I x R drop directly at the unknown resistor. This is the most accurate method of measuring small values of resistance. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Auto-range or manual range selection. 3. Set the resistor box to 1060 Ω and operate the R1L-D1. Check that the R1L-D1 reads between the resistor box 1060 Ω setting’s actual value ± 0.54 Ω in Auto-Range (default condition when power is turned on).
Page 19
MAINTENANCE 5. Set the resistor box to 10 Ω and operate the R1L-D1. Check that the R1L-D1 reads between the resistor box 10 Ω setting’s actual value ± 0.0051 Ω in Auto-range. Then step through the ranges to the 20 Ω manual range and check that the R1L-D1 reads between the resistor box 10 Ω...
OF THIS PROCEDURE. 1. If the R1L-D1 rechargeable battery cells have not been previously charged, or are low on charge, connect the R1L-D1 AC line cord and allow a charge time of at least one-half hour. The R1L-D1 "Battery Charging LED (as "...
Page 21
HyperTerminal and setup the RS-232 parameters with following setting. Rest the R1L-D1 front panel in its proper orientation in its yellow case. Step 1: Enter a name such as ‘R1L-D1’. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Page 22
Step 3: Enter port settings shown on the right. 9. Connect the 1 Ω standard and turn ON the R1L-D1. Confirm the R1L-D1 automatically goes to the 2 Ω range, and indicates 4-wire, an asterisk for auto-range, and the resistance. Resistance values should also appear on the PC screen.
Page 23
10 Ω range, 100 Ω range, and 1 kΩ range. 16. Turn the R1L-D1 OFF, wait 10 seconds, and turn it back ON. Connect the 1 Ω standard and verify a reading of within ± 0.00015 Ω of the calibration value. Connect the 1 kΩ...
± 0.004 Ω of the calibration value. 18. Disconnect the +E side of the test leads from the R1L-D1 (leaving connections between the +I and -I terminals). Depress the R1L-D1 "Mode" button to place the R1L-D1 in 2-wire mode.
Page 25
D converter, are plugged into sockets and may be removed easily, with no need for unsoldering. Other components, such as resistors, capacitors, etc., must be unsoldered for removal. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
4.8 Overhaul Instructions The Model R1L-D1 is an all solid-state unit and requires no periodic overhaul, other...
Page 28
The battery may be discharged and recharged at least 100 completed cycles before requiring replacement. Note that, it recommended that NiCad batteries completely discharged before recharging or their effective capacity is reduced. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com 4-12...
Once this information is prepared and sent with the instrument to our service department, we will do our part in making sure that you receive the best possible customer service and turnaround time possible. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Repaired products are warranted for the remaining balance of the original warranty, or 90 days, whichever period is longer. 10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com...
Page 35
TEGAM specifically disclaims any implied warranties of merchantability or fitness for a particular purpose. In no event will TEGAM be liable for special or consequential damages. Purchaser’s sole and exclusive remedy in the event any item fails to comply with the foregoing express warranty of TEGAM shall be to return the item to TEGAM;...