CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to use.
Calibration of your 2101KL scale can be performed using calibrated kilogram (kg) or
pound (lb) weight standards. Minimum calibration is 200 lb / 100 kg. Maximum
calibration weight is 1000 lb / 454 kg. The required calibration standard will be the Unit
of Measure (UOM) for which the scale is configured.
Note: Only weights certified and traceable to national standards are to be used for
calibration.
Warning: Always ensure that the scale is returned to the correct UOM after scale
calibration.
Help
1. Press
to access the Help menu options.
2. While in the Help menu, press
3. The display will prompt for entry of the password. Press
the following keys in this sequence to proceed with the
calibration procedure.
4. Follow the screen prompt and ensure the scale is on a
level surface with no weight on the platform.
5. The display will show "Zeroing Scale" as it calculates the
zero value. Do not touch the scale during the zeroing
process.
6. When the zeroing process is complete the display will
prompt for entry of the calibration weight. Minimum
calibration is 200 lb / 100 kg. Maximum calibration
weight is 1000 lb / 454 kg. Calibration weight can be
adjusted in 10 lb increments. Press
calibration weight is set.
to select "Calibrate".
when the
Enter
20