Fluke 5205A Instruction Manual page 57

Precision power amplifier
Hide thumbs Also See for 5205A:
Table of Contents

Advertisement

5205A
.
/
j
5205
STATUS
14
tOOOV
CONTROL
2
S206
7R(I>
15
TO
P41
S200
•SVdc
:
2
6200-^Vi}«
RgTORN
IS
ioociv
oi>eHA+£
SWOOMflO
17
SCNS8
OUTPUT
S
TO
TOS200
SENSE
OUTVUT
COMMON
s
TO
S2C0ASIONAL
INPUT
24
S200ASISNAL
INPUT
COMMON
23
^
81
K
RFO
!
'
/-"N
SLX
1
YFt.
CRftt
1
!
RIU
'
'
!
BLU
WMT
1
1
t
i
WKT
...
T
^
WKT
BIK
1
/•
N
rEt.
!
Tve,
Figure 4-9.
CALIBRATOR
MODE
INTERCONNECTIONS
4-131. Logic Trip
Adjustment
NOTE
This setting
monitors
the
logic
supply
voltage
and
is
adjusted
at
the factory.
It
should
not
require
normal
calibration,
however, adjustment
could be
required
if drift
within the
circuit
is
sufficient to
trip
the
instrument
into
FAULT.
4-1
30
If
required, adjust the
logic trip circuit
using the
following
procedure;
a.
Verify that
there are
no jumper
pcbs
or
jumpers
installed in
the
5205
A
and both power
cords
are
connected
to
the
line
source.
b.
Connect
the
DVM
Hi
input
to
TP
409
(+5)
and
the
LO
input
to
TP 407
(GND).
Prepare the
DVM
to read
+5
volts
dc.
Both
test
points
are
located
on
the
logic
pcb.
c.
Connect
a
jumper between
TP 418
and
TP
417.
(Located
near the
FAULT
and
OVERLOAD
in-
dicators
on
the
Logic
pcb.)
d.
Set the
POWER
switch
to
ON
and
wait
30
seconds
for
the
warm-up
period
to expire.
e.
The
DVM
should read
between
4.75
and
5.25V
dc.
f.
Rotate
R47
fully
clockwise
and
then very slowly
adjust
until
the
fault
indicator extinguishes.
g.
Check
at
high
&
low
line
voltage
limits
(nominal
±10%)
that the
Logic
circuits
+5
Volts
is
between
+4.75 and +5.25
Volts dc and the
instrument does
not
trip into
FAULT.
h.
Set the
POWER
switch
to
OFF
and
disconnect
the
test
equipment and jumper.
4-133.
TROUBLESHOOTING
4-134.
Troubleshooting
for
the
5205A
consists
of
the
tab-
ular
flow
chart
in
Table
4-3.
When
a step
on
the
flow
chart
is
completed
check
for
a
decision
transfer.
If
no
decision
is
required
perform
the
next
step
of
the table
in
sequence.
WARNING!
Lethal
vollages
up
to
2.5K
Volts
are
present
within
the instrument during troubleshooting. In addition,
the voltage
may
be
present
for
up
to
30
seconds
after
power
is
removed from
the instrument.
Before
coming
into physical
contact with
any
component
use a voltmeter
and
5kV
probe
to insure
there
is
no
voltage present
between tp502 (+2.5kV)
and
tpS03(LO), between tpS04 (— 2.5kV) and tp506
LO), and between tp505 (-2.5kV) and tp506 (LO).
All
test
points
are
on
the
Power Supply
pcb.
Extreme
caution should
be
exercised
at
all
times
when
working
with
this
instrument.
4/76
4-23

Advertisement

Table of Contents
loading

Table of Contents