Table 5-9. Ohms Source end Resistor Checks
Display
4
TP9"
5
INPUT
HI to LO
6
INPUT HI to LO
7
1.9kl7 resistor.
8
Display
9
19kR resistor.
10
Display
11
190kC resistor.
12
Display
13
1900kD resistor.
14
Display
15
1OMD resistor
16
Display
..w'IY..Y"
-.,.."...-..
Check chopper
amplifier
operation
per Table 5-7, steps 1 thru 10.
Select 2kD range. Short Input.
.OOOO *I
digit.
-IV Approximately
Remove
short end connect
ammeter
from INPUT HI to LO. Source
current
should be approximately
ImA,
HI terminal
negative.
Remove ammeter
end measure open
circuit voltage.
Voltage
must be
-2 to -5v.
Apply
to INPUT.
1.9000
* 12 digits.
Apply
to INPUT.
Select 20kD range.
19.000
-f 12 digits.
Apply
to INPUT.
Select 200kR range.
190.00
f 12 digits.
Apply
to INPUT.
Select 2000kR
range.
1900.0
f 12 digits.
Apply
to INPUT.
Select 20MI2 range.
10.000
* 12 digits.
NOTE: Reference
loading
by the
ohms source does not affect D
readout
since AID
converter
is
ratiometric.
I
*See Figure 5-l.
Table 5.10. Current
Shunt Checks
Step
Item/Component
Required
Condition
F
.." ..-...-
Inverted
reference
voltage
plus calibrated
offset of
u107.
If incorrect,
check R158 end
switching
to INPUT
HI end
chopper
amplifier.
HI negative
with respect
to
LO. Display should be blink-
ing, indicating
overrange.
Calibrated
resistance.
Subtract
test lead resistance
from reading.
Checks
accuracy
of R158 and feed-
beck loop to U107.
Calibrated
resistance.
Source resistors:
RI56
R154
(6975kDl.
R157 (wiper).
Calibrated
resistance.
Source resistors:
R158, RI57
R156 (wiper),
R154 l89.75kC.
8.975M).
Calibrated
resistance.
Source resistors:
R158, R157.
R156. RI55 fwiperl.
R154, (all
except 6975MRl.
Calibrated
resistance.
Source resistors:
all R range
resistors.
NOTE: To con-
veniently
measure the resis-
tance of all D range resistors,
turn power off, set all
function
switches
out end
measure
resistance
INPUT
HI
to LO.
.___
Remarks
L
Apply
a known
% scale
current
end measure
voltage
across shunt.
( Cl;;ping
must occur et
5-11