Dccurrentmeasutement; Accurrentmeasurement - Keithley 177 Instruction Manual

Microvolt dmm and battey pack
Hide thumbs Also See for 177:
Table of Contents

Advertisement

A. Select 200 range.
D. Adjust front panel ZERO knob for a display indication
of
B. Disconnect
the test leads at the circuit to be measured
O.OOO,rA * ,001 flashing.
and short them.
If measurement
is to be made right at
2. After adjusting
ZERO, use the Model 177 to measure
DC
the INPUT terminals
(no test leads used) place a short,
current as follows:
such as a piece of copper
wire,
across the binding
posts.
CAUTIONS
C. Adjust
the front panel ZERO knob for a display indica-
Do not install
e larger capacity
fuse then
tion of 0.000
*O.OOlG
(flashing).
The nominal
ZERO
the one originally
supplied
L?Al. Current
adjustment
span is 200mD.
fuse FlOl (shown in Figure 5-4) protects
the
instrument
against over-current.
NOTE
The specified
accuracy
for the 2OD and 2OOG
A. Select the desired range from the six ranges available.
ranges assumes that ZERO was adiusted
on the
8. Connect
the circuit to be measured
to the INPUT ter-
20irange.
Re-adjust zero on the 2dg range each
time
that
a different
set of leads is used or
whenever
measurements
have been made on
minals.
C. Read the displayed
polarity,
digits and decimal point.
Select the next higher range if overrange
is indicated.
either the two lowest
DC voltage
ranges or any
DC current
range.
The zero setting
for those
measurements
will be different
than the zero set-
ting of ohms, Also, re-adjust
zero whenever
the
instrument
is used
outside
the
temperature
range of 18O to 28OC.
D. Reconnect
the test leads to the circuit to be measured.
E. Select
either 2OD range
or 2000
range.
Read
the
displayed
digits and decimal
point.
NOTE
An input overload to the Model 177 may cause a
temporary
zero shift due to thermal
gradients.
2.12 DC CURRENT MEASUREMENT
2.13 AC CURRENT MEASUREMENT
The Model 177 measures
AC current from 1OnAldigit
to 2A.
over
five
ranges.
(The' 20r~A and
20M
ranges
are not
operative.
See note
below.1
The 200r~A range
is direct-
reading
in microamperes.
The remaining
AC current
ranges
are direct-reading
in milliamperes.
The maximum
reading
is
19999. Overrange
is indicated
by a flashing 0000. Overload
is
protected
by fuse FlOl,
a 2A, 3AG normal-blow
fuse. The
instrument
measures
the true root mean square ITRMSj
of
waveforms
in the frequency
range of 45 to 10kHz. Use the
Model 177 to measure AC current
as follows:
1. Turn on power
and set the AC/DC
pushbunons
to in or
AC position.
Depress the A pushbunon.
CAUTION
The Model 177 measures
DC current
from lnA/digit
to 2A.
over six ranges
(the 20M
range
is not operative
and the
Do not install
a larger capacity
fuse then
instrument
will read zero if 20M pushbutton
is depressed).
the one supplied.
Current fuse FlM (shown
The maximum
reading is 19999. Overrange
is indicated
by a
in
Figure
5-41 protects
the
instrument
flashing 0000. Overload
is protected
by fuse FlOl, a 2A 3AG
against over current.
normal-blow
fuse.
Nominal
settling
time
for
DC current
2. Connect the circuit to be measured
to the INPUT terminals
ranges is one second to within
one digit of final reading ex-
and select the desired range from the five ranges available.
cept three seconds
on 20pA range.,
NOTE
1. Before making
DC current
measurements,
the front panel
ZERO must be adjusted
on the 20@
range. This adjust-
The 20/1A and 20M ranges are not operative
for
AC current
measurements.
If the 20uA
push-
ment
is necessary
to remove
any offset
that was intr-
button
is inadvertently
depressed,
an overrange
oduced by adjusting
ZERO for a low DC voltage
measure-
mentor
a low g measurement.
This adjustment
will also be
will be indicated,
signalling
the user to go to the
necessary
whenever
the instrument
is used outside of the
next higher range. If the 20M pushbutton
is de-
temperature
range of 18O to 28Y.
The specified
accuracy
pressed, the instrument
will read zero.
for the DC current
ranges assumes
that ZERO was ad-
3. Read the displayed
digits and decimal
point.
justed on the 2OpA range. Adjust
ZERO as follows:
A. Disconnect
all circuits
from the Model 177 INPUT ter-
NOTE
minals.
The INPUT
terminals
must have an open cir-
AC accuracy
is specified
for 2000 counts
and
cuit~
above. The method
of calibratina
the converter
_-...
8. Turn on power and set the AC/DC
pushbunon
to the
out or DC position.
C. Depress A and 20~ pushbuttons.
may yield an offset
up to 80 digits
with
the
Model 177 input shorted.
This does not affect in-
strument
accuracy.
There
will be some addi-
tional measurement
error for signals with a crest
2-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1788

Table of Contents